Results of the 2020 fastMRI Challenge for Machine Learning MR Image
Reconstruction
Matthew Muckley, Bruno Riemenschneider, Alireza Radmanesh, Sunwoo
Kim, Geunu Jeong, Jingyu Ko, Yohan Jun, Hyungseob Shin, Dosik
Hwang,
Mahmoud Mostapha, et al.
To cite this version: Matthew Muckley, Bruno Riemenschneider,
Alireza Radmanesh, Sunwoo Kim, Geunu Jeong, et al.. Results of the
2020 fastMRI Challenge for Machine Learning MR Image
Reconstruction. IEEE Transactions on Medical Imaging, Institute of
Electrical and Electronics Engineers, In press,
10.1109/TMI.2021.3075856. hal-03066150v2
This article has been accepted for publication in a future issue of
this journal, but has not been fully edited. Content may change
prior to final publication. Citation information: DOI
10.1109/TMI.2021.3075856, IEEE Transactions on Medical
Imaging
IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. XX, NO. XX, XXXX 2020
1
Results of the 2020 fastMRI Challenge for Machine Learning MR Image
Reconstruction
Matthew J. Muckley∗,1, Member, IEEE, Bruno Riemenschneider∗,2,
Alireza Radmanesh2, Sunwoo Kim3, Member, IEEE, Geunu Jeong3, Jingyu
Ko3, Yohan Jun4, Hyungseob Shin4, Dosik Hwang4,
Mahmoud Mostapha5, Simon Arberet5, Dominik Nickel6, Zaccharie
Ramzi7,8, Student Member, IEEE, Philippe Ciuciu7, Senior Member,
IEEE, Jean-Luc Starck8, Jonas Teuwen9, Dimitrios
Karkalousos10,
Chaoping Zhang10, Anuroop Sriram11, Zhengnan Huang2, Nafissa
Yakubova1, Yvonne W. Lui2, and Florian Knoll2, Member, IEEE
Abstract— Accelerating MRI scans is one of the prin- cipal
outstanding problems in the MRI research commu- nity. Towards this
goal, we hosted the second fastMRI competition targeted towards
reconstructing MR images with subsampled k-space data. We provided
participants with data from 7,299 clinical brain scans
(de-identified via a HIPAA-compliant procedure by NYU Langone
Health), holding back the fully-sampled data from 894 of these
scans for challenge evaluation purposes. In contrast to the 2019
challenge, we focused our radiologist evaluations on pathological
assessment in brain images. We also debuted a new Transfer track
that required participants to submit models evaluated on MRI
scanners from outside the train- ing set. We received 19
submissions from eight different groups. Results showed one team
scoring best in both SSIM scores and qualitative radiologist
evaluations. We also performed analysis on alternative metrics to
mitigate the effects of background noise and collected feedback
from the participants to inform future challenges. Lastly, we
identify common failure modes across the submissions, highlighting
areas of need for future research in the MRI reconstruction
community.
Index Terms— Challenge, Public Data Set, MR Image Reconstruction,
Machine Learning, Parallel Imaging, Com- pressed Sensing, Fast
Imaging, Optimization
I. INTRODUCTION
Due to advances in algorithms, software platforms [1], [2], [3] and
compute hardware, over the last five years there has been a surge
of research of MR image reconstruction methods
∗Equal contribution. 1Facebook AI Research, New York, NY, USA 2NYU
School of Medicine, New York, NY, USA 3AIRS Medical, Seoul, South
Korea 4Yonsei University, Seoul, Korea 5Siemens Healthineers,
Princeton, NJ, USA 6Siemens Healthcare GmbH, Erlangen, Germany 7CEA
(NeuroSpin) & Inria Saclay (Parietal), Universite
Paris-Saclay,
F-91191 Gif-sur-Yvette, France 8Departement d’Astrophysique,
CEA-Saclay, 91191 Gif-sur-Yvette,
France 9Radboud University Medical Center, Nijmegen, Netherlands
10Amsterdam UMC, Amsterdam, Netherlands 11Facebook AI Research,
Menlo Park, CA, USA We acknowledge grant support from the NIH under
awards
R01EB024532, P41EB017183, and R21EB027241.
based on machine learning [4], [5], [6], [7], [8], [9], [10], [11],
[12], [13], [14]. Traditionally, research in MR image
reconstruction methods has been conducted on small data sets
collected by individual research groups with direct access to MR
scanner hardware and research agreements with the scanner vendors.
Data set collection is difficult and expensive, with many research
groups lacking the organizational infrastructure to collect data at
the scale necessary for machine learning research. Furthermore,
data sets collected by individual groups are often not shared
publicly for a variety of reasons. As a result, research groups
lacking large-scale data collection infrastructure face substantial
barriers to reproducing results and making comparisons to existing
methods in the literature.
Such challenges have been seen before. In the field of computer
vision, the basic principles of convolutional neural networks
(CNNs) were proposed as early as 1980 [15] and became
well-established for character recognition by 1998 [16]. Following
Nvidia’s release of CUDA in 2007, independent research groups began
to use GPUs to train larger and deeper networks [17], [18].
Nonetheless, universal acceptance of the utility of CNNs did not
occur until the debut of the large-scale ImageNet data set and
competition [19]. The introduction of ImageNet allowed direct
cross-group comparison using this well-recognized data set of a
size beyond what most groups could attain individually. In 2012 a
CNN-based model [20] out-performed all non-CNN models, spurring a
flurry of state- of-the-art results for image recognition [21],
[22], [23], [24].
Since 2018, the fastMRI project has attempted to advance
community-based scientific synergy in MRI by building on two
pillars. The first consists of the release of a large data set of
raw k-space and DICOM images [25], [26]. This data set is available
to almost any researcher, allowing them to download it, replicate
results, and make comparisons. The second pillar consists of
hosting public leaderboards [25] and open competitions, such as the
2019 fastMRI Reconstruction Challenge on knee data [27]. The
dimension of public compe- titions is not new to the MR community.
Other groups have facilitated challenges around RF pulse design
[28], diffusion tractography reconstruction [29], and ISMRM
initiatives for reconstruction [30], [31].
The 2020 fastMRI Challenge continues this tradition of open
This work is licensed under a Creative Commons Attribution 4.0
License. For more information, see
https://creativecommons.org/licenses/by/4.0/
This article has been accepted for publication in a future issue of
this journal, but has not been fully edited. Content may change
prior to final publication. Citation information: DOI
10.1109/TMI.2021.3075856, IEEE Transactions on Medical
Imaging
2 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. XX, NO. XX, XXXX
2020
competitions and follows the 2019 challenge with a few key
differences. First, our target anatomy has been changed to focus on
images of the brain rather than knee. Second, for 2020 we updated
the radiologist evaluation process, asking radiologists to rate
images based on depiction of pathology rather than overall image
quality, emphasizing clinical relevance in com- petition results.
Lastly, we address a core traditional problem in MR imaging: the
capacity of models to generalize across sites and vendors. We
introduce a new competition track: a “Transfer” track, where
participants were asked to run their models on data from vendors
not included in training. This contrasts with the 2019 challenge,
which only included data from a single vendor for both training and
evaluation.
II. METHODS
This challenge focuses on MRI scan acceleration, a topic of
interest to the MR imaging community for decades. MRI scanners
acquire collections of Fourier frequency “lines”, commonly referred
to as k-space data. Due to hardware constraints on how magnetic
fields can be manipulated, the rate at which these lines are
acquired is fixed, which results in relatively long scan times and
has negative implications with regard to image quality, patient
discomfort, and accessibility. The major way to decrease scan
acquisition time is to decrease the amount of data acquired.
Sampling theory [32], [33], [34], [35] states that a minimum number
of lines are required for image reconstruction. This minimum
requirement can be circumvented by incorporating other techniques
such as parallel imaging [36], [37], [38] and compressed sensing
[39]. More recently, machine learning methods have demonstrated
further accelerations over parallel imaging and compressed sensing
methods.
To promote the advancement of methods for accelerated MRI, we
organized a public challenge. We applied retrospective downsampling
to fully-sampled MRIs and provided the down- sampled data to
challenge participants. Challenge participants ran their models on
the downsampled data and submitted it to the competition website at
https://fastmri.org, where we quantitatively evaluated it using the
fully-sampled data as gold standards. We selected six cases for
each of the top three teams in each track of the challenge (three
tracks total) and presented the cases to a group of six
radiologists for qualitative evaluation. The challenge winner was
selected based on the best depiction of pathology compared to the
ground truth as judged by radiologists.
At a high level we describe the principles of our 2020 challenge as
follows. Using knowledge we gained through the 2019 challenge, we
identified a few key alterations for 2020. These include:
• A new imaging anatomy, the brain, the most commonly- imaged organ
using MRI.
• A focus on an evaluation of pathology depiction rather than
overall image quality impressions to strengthen the connection
between the challenge evaluation and clinical practice.
• An emphasis on generalization with the introduction of a new
“Transfer” track where participants were asked to run their models
on multi-vendor data.
• We removed the single-coil track and moved to a pure multi-coil
challenge to increase the clinical relevance of the submitted
models.
• Due to easier practical implementation and removal of the
single-coil track, we used pseudo-equispaced subsam- pling masks
(i.e., equispaced masks with a modification for achieving exact
4X/8X sampling rates) rather than random. This follows more closely
sampling patterns (and relaxation effects) that are used for
parallel imaging in vendor sequences, facilitating easier clinical
deployment. We maintained the fully-sampled center due to its
utility for autocalibrating parallel imaging methods [36], [38],
[40] and compressed sensing [39].
• In the 2019 challenge our baseline model was a U- Net [41];
however, winning models [42], [43], [44] of the 2019 challenge were
variational network/cascading models [27]. For the 2020 challenge,
we provided a much stronger baseline model based on an End-to-End
Variational Network [14].
We kept the following principles from the 2019 challenge:
• We again used a two-stage evaluation, where a quantitative metric
was used to select the top 3 submissions. These finalists were then
sent to radiologists to determine the winners. We used the
structural similarity index (SSIM) [45] as our quantitative image
quality index for ranking submissions prior to submission to
clinical radiologists [27].
• We wanted to maintain realism for a straightforward, 2D imaging
setting, and so all of the competition data was once again based on
fully-sampled 2D raw k-space data.
• For the ground truth reference, we had discussions on
alternatives to the root sum-of-squares (RSS) method used for
quantitative evaluation in 2019. Although there was some consensus
on the drawbacks of RSS [46], [47], there was no consensus on a
single best alternative. In the following sections we discuss the
impact of this choice further.
A. Challenge Tracks
In the 2019 challenge we included three submission tracks:
multicoil with four times acceleration (Multi-Coil 4X), multi- coil
with eight times acceleration (Multi-Coil 8X), and single- coil
with 4X acceleration (Single-Coil 4X). Among these tracks, the
single-coil track garnered the most engagement, but due to its
distance from clinical practice we decided to remove it from the
2020 challenge, replacing it with the Transfer track. For the
standard multicoil tracks in the 2019 challenge, we observed that
although there were many high-quality submissions at 4X, all of the
submissions began missing pathology at 8X acceleration [27]. Since
this time, 4X machine learning methods have been validated for
clinical interchangeability [48]. This suggests that the current
upper limit of 2D machine learning image reconstruction performance
remains between 4-fold and 8-fold acceleration rates. In order to
provide participants with both an obtainable target and a “reach”
goal, we kept the 4-fold and 8-fold tracks for the 2020
challenge.
This work is licensed under a Creative Commons Attribution 4.0
License. For more information, see
https://creativecommons.org/licenses/by/4.0/
This article has been accepted for publication in a future issue of
this journal, but has not been fully edited. Content may change
prior to final publication. Citation information: DOI
10.1109/TMI.2021.3075856, IEEE Transactions on Medical
Imaging
MUCKLEY, RIEMENSCHNEIDER et al.: PREPARATION OF PAPERS FOR IEEE
TRANSACTIONS ON MEDICAL IMAGING 3
One frequent feedback on the 2019 challenge was on
generalizability: despite the size of the data set, all of the data
and results were from studies performed on MRI scanners from a
single vendor at a single institution. To address this, we created
the new Transfer track at 4-fold acceleration (Transfer 4X). For
the Transfer track, participants were asked to run their models on
data from vendors outside the main fastMRI data set. There was a
caveat: we also restricted participants in the Transfer track to
train their models only using available fastMRI data to ensure
evaluation of transfer capability. At the time of the 2020
challenge announcement, we stated that these data would come “from
another vendor” but did not specify further. At the challenge
launch time, we revealed that the challenge data for this track was
a mix of data from GE and Philips, providing additional difficulty
for participants. As a result, submissions in the Transfer track
exhibited wide deviations in performance depending on vendor.
B. Data Set
For the 2020 challenge we used brain MRI data. The neuroimaging
subset of the fastMRI data has been described in an updated version
of the arXiv paper [25], with further information included in the
supplemental material for this paper. It includes 6,970 scans
(3,001 at 1.5 T, 3,969 at 3 T) collected at NYU Langone Health on
Siemens scanners using T1, T1 post-contrast, T2, and FLAIR
acquisitions. Unlike the knee challenge, this data set exhibits a
wide variety of reconstruction matrix sizes. A summary of the data
for the two main track splits is shown in Table I. Of these 6,970
scans, 565 were withheld for evaluation in the challenge. In
addition to standard HIPAA-compliant anonymization practices, all
scans were cropped at the level of the orbital rim, preserving only
the top part of the head.
TABLE I SUMMARY OF THE CHALLENGE DATA
Split T1 T1POST T2 FLAIR Total
Siemens/Main Tracks train 498 949 2,678 344 4,469 val 169 287 815
107 1,378 test (4X) 33 54 170 24 281 test (8X) 32 68 152 25 277
challenge (4X) 26 67 192 18 303 challenge (8X) 24 65 159 14
262
Transfer Track (4X, all challenge) GE 22 29 83 77 211 Philips 18 0
50 50 118
For the challenge, the 565 scans were augmented further by 329
non-Siemens scans for the Transfer track. GE data were collected at
NYU Langone Health and Philips data were collected on volunteers by
clinical partner sites of Philips Healthcare of North America.
Since the Philips data was collected on volunteers, this subsplit
had no post-contrast imaging. One difficulty of the Transfer track
was the fact that the GE data did not contain frequency
oversampling. The lack of frequency oversampling was due to
automatic removal
during the analog-to-digital conversion process on the GE
scanner.
In total the 2020 challenge had 6,405 scans available for training
and validation (train, val, test) and there were 894 total scans
for evaluation in the final challenge phase. This marked a
substantial increase in scale from the 2019 challenge. For
reference, the multicoil data from the 2019 challenge on knee data
had 1,290 scans for training and validation (train, val, test) and
104 scans for the challenge, so the data for training increased by
roughly 5-fold and the data for challenge evaluation increased by
roughly 8-fold.
Participants were restricted to only use this data set for training
the weights of their models. We did permit partic- ipants to use
their own data as validation data, but not for backpropagating
gradients.
C. Evaluation Process
Submissions were processed via https://fastmri. org. This site
maintains a submission system for both the challenge and the public
test leaderboard. Currently, there are no plans to make the
challenge split of the data available in order to maintain the
integrity of the results, but research groups may submit to the
public leaderboard using the test set via the website at any
time.
After submission, evaluation followed a two-stage process of
comparisons to the fully-sampled “ground truth” images. For the
ground truth images, we followed the previous conven- tion [27] to
use root sum-of-squares images. The advantage of this approach is
that it does not bias to any one method for coil sensitivity
estimation. There are some drawbacks to RSS, including 1)
discarding of the phase information and 2) RSS images can have
substantial noise in the background. The phase is not typically
used for anatomical evaluation. The issue with noise is more
fundamental as it is treated as ground truth in our quantitative
evaluation, and any deviations from it influence our ranking. This
is counterbalanced by using radiologist evaluation for declaring
the challenge winner. In planning for the challenge, we were unable
to build consensus on an alternative ground truth calculation
technique, but this topic could be re-examined in future
challenges. For the quantitative evaluation metric, we chose to use
SSIM [45], with a script showing the script used for evaluation in
the fastMRI repository at https://github.
com/facebookresearch/fastMRI. SSIM has several parameters. We
investigated adjusting these parameters prior to challenge launch,
but found that they generally did not alter the ranking of methods
evaluated in our quality control phase, and so as a result we used
the default parameters in scikit-image [49].
For the qualitative assessment phase, a board-certified
neuroradiologist selected six (two T1 post-contrast, two T2, and
two FLAIR) cases from the challenge data set in each of the three
tracks. Cases were specifically selected to represent a broad range
of neuroimaging pathologies from intracranial tumors and strokes to
normal and age-related changes. The selection process favored cases
with more subtle pathologies for the 4X track and more obvious
pathologies for the 8X track with the objective that this might
yield better granularity for
This work is licensed under a Creative Commons Attribution 4.0
License. For more information, see
https://creativecommons.org/licenses/by/4.0/
This article has been accepted for publication in a future issue of
this journal, but has not been fully edited. Content may change
prior to final publication. Citation information: DOI
10.1109/TMI.2021.3075856, IEEE Transactions on Medical
Imaging
4 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. XX, NO. XX, XXXX
2020
separating methods in the 4X track. Selected cases included both
intraaxial and extraaxial tumors, strokes, microvascular ischemia,
white matter lesions, edema, surgical cavities, as well as
postsurgical changes and hardware including craniotomies and
ventricular shunts. The Philips data set was constructed from
images of volunteers. Therefore small age-related imaging changes
were used for ranking in place of pathology.
Six radiologists with 9-16 years of experience (two of whom are
radiology division chiefs) were asked to evaluate the 18 selected
image volumes for each team, basing their overall ranking on the
quality of the depiction of the pathology using the ground truth as
a reference. Radiologists came from a wide set of institutions,
including the Mayo Clinic, Baylor College of Medicine, NYU Langone
Health, the University of Pittsburgh Medical Center, Stanford
University, and the University of California, Los Angeles. None of
these institutions had finalist submissions. All radiologists
looked at all images in the selected cases during the qualitative
evaluation phase, and results were averaged. Radiologists were
aware of the overarching goals of the challenge but were blinded as
to which teams submitted the images. In addition, we also asked
radiologists to score each case in terms of artifacts, sharpness
and contrast-to-noise ratio (CNR) using a Likert-type scale. On the
Likert scale, 1 was the best (e.g., no artifacts) and 4 was the
worst (e.g., unacceptable artifacts). A Likert score of 3 would
affect diagnostic image quality.
D. Timeline The 2020 challenge had the following timeline: •
December 19, 2019 - Release of the brain data set and
update to the arXiv reference [25]. • July 9, 2020 - Announcement
of the 2020 challenge. • October 1-15, 2020 - Release of the
challenge data set
and submission window. • October 16-19, 2020 - Calculation of SSIM
scores.
We selected the top 3 submissions for each track and forwarded them
to a panel of radiologists for qualitative evaluation.
• October 19-November 1, 2020 - Radiologists evaluated submissions.
They were asked to complete a score sheet for each of the 3 tracks
which included ranking the submissions for each individual case in
terms of overall quality of depiction of pathology.
• December 5, 2020 - Publication of the challenge leader- board
with results.
• December 12, 2020 - Official announcement of the winners of the
three tracks with presentations at the Medical Imaging Meets
NeurIPS Workshop.
E. Overview of Submission Methodologies Here we share a brief
description of the methodologies
behind each of the submissions that made it to the finalist round
for radiologist evaluation. The developers of these submissions are
included as co-authors on this paper.
A brief overview of finalist model properties is shown in Table II.
(Team names: “AIRS” is AIRS Medical, “ATB” is ATB, “MRR” is
MRRecon, “Nspin” is Neurospin, “Res” is
TABLE II FINALIST MODEL PROPERTIES
Team # Params Init. Coil Meth. GPUs Tr. Time
AIRS 200 M GRAPPA ESPIRiT 4 (V100) 7 days ATB 21 M Zero-Filled
U-Net 8 (TITAN) 10 days MRR 16 M Zero-Filled ESPIRiT 1 (V100) 14
days Nspin 155 M Zero-Filled U-Net 1 (V100) 7 days Res 841 k
ESPIRiT U-Net 4 (RTX 8000) 21 days
ResoNNance.) The number of model parameters ranged from 841,000 in
the case of ResoNNance to 200 million in the case of AIRS. Teams
applied GRAPPA [38], ESPIRiT [40], or simple zero-filled
initializations. For coil estimation, teams used either ESPIRiT
[40] or a simple center-based estimation with U-Net refinement
similar to that in the End-to-End Variational Network [14]. Teams
used 1-8 GPUs for training, and training time was between 7 and 21
days.
AIRS Medical: The AIRS Medical model used a combi- nation of image-
and k-space-domain processing in a fashion analogous (but distinct)
from that used in KIKI-Net [8]. The model included a data
consistency cascade with 4 U-Net stages. At each convolutional
layer of the U-Net [41], the multi-domain processing split the
channels into one group that operated in image space and one group
that operated in k-space [8]. Data consistency was enforced at each
layer. The network was initialized with a GRAPPA estimate [28]
(preprocessed into reconstruction + residual), and coil
sensitivities were estimated using ESPIRiT [40]. Since the sampling
pattern was pseudo-equispaced, multiple GRAPPA kernels were used to
calculate the GRAPPA images. AIRS optimized their model using Adam
[50] over 20 epochs using a batch size of 4 at a learning rate of
10−3 (decayed to 10−4 after 15 epochs) with SSIM as the loss
function. Optimization took approximately 7 days on four NVIDIA
V100 GPUs. Code is not publicly available.
ATB: The ATB model, called “Joint-ICNet” [51], was a 10- iteration
unrolled algorithm with CNNs replacing regularization terms in a
fashion similar to other recent methods [5], [14], [8]. Joint-ICNet
used the U-Net [41] at each convolutional layer with the
dual-domain processing previously introduced in KIKI-net [8].
Joint-ICNet used a zero-filled reconstruction as the initial
estimate and and coil sensitivities were calculated by refining a
rough central k-space estimate with a U-Net [41], [14]. ATB
optimized Joint-ICNet using Adam [50] over 50 epochs using a batch
size of one at a learning rate of 10−4
with SSIM as the loss function. Training took approximately 10 days
using 8 NVIDIA TITAN GPUs. Code is not publicly available.
MRRecon: The MRRecon model, called “Momen- tum DIHN,” unrolled the
Nesterov momentum algorithm with CNN-based regularization for 12
cascades, 6 pre-cascades, and 0 or 1 post-cascade. The CNN module
is a “Deep, Iterative, Hierarchical Network” (DIHN) that extends
the Down-Up network [52] with a hierarchical block design,
facilitating memory efficiency over a standard U-Net [41]. Momentum
DIHN used a zero-filled image as the initial estimate and ESPIRiT
for calculating coil sensitivities. To
This work is licensed under a Creative Commons Attribution 4.0
License. For more information, see
https://creativecommons.org/licenses/by/4.0/
This article has been accepted for publication in a future issue of
this journal, but has not been fully edited. Content may change
prior to final publication. Citation information: DOI
10.1109/TMI.2021.3075856, IEEE Transactions on Medical
Imaging
MUCKLEY, RIEMENSCHNEIDER et al.: PREPARATION OF PAPERS FOR IEEE
TRANSACTIONS ON MEDICAL IMAGING 5
improve transfer track performance, models with several
hyperparameters were ensembled to generate the final images.
MRRecon optimized Momentum DIHN using Adam [50] for less than 5
epochs using a batch size of one at a learning rate of 10−4 with a
compound L1/MS-SSIM loss function [42]. Training took approximately
14 days on an NVIDIA V100 GPU. Code is not publicly
available.
Neurospin: The Neurospin model, called XPDNet [53], [54], is a
modular neural network unrolling the Chambolle-Pock algorithm [55]
for 25 iterations. The model was inspired by the primal-only
version of the Primal-Dual net [56], replacing the vanilla CNN with
a multi-level wavelet CNN [57]. XPDNet used a zero-filled image as
the initial estimate and calculated coil sensitivities using a
rough central k-space estimate refined by a U-Net [41], [14].
Neurospin optimized XPDNet using Rectified Adam [58] over 100
epochs using a batch size of one at a learning rate of 10−4 with a
compound L1/MS-SSIM loss function [42] (98% MS-SSIM weight).
Training took approximately 7 days on an NVIDIA V100 GPU. Code is
available at https://github.com/zaccharieramzi/
fastmri-reproducible-benchmark.
ResoNNance: ResoNNance used a Recurrent Inference Machine (RIM)
that has been previously described [59], [60], [44], [31]. Coil
sensitivities were calculating using the center of k-spaced
followed by U-Net refinement [41]. RIM used ESPIRiT as the model
input calculated from the BART toolbox [61]. ResoNNance optimized
RIM using Adam [50] over 90 epochs using a batch size of one at a
learning rate of 10−3 with an SSIM loss function. Separate models
were trained for every field strength (1.5 T, 3 T) and contrast
(FLAIR, T1/T1PRE/T1POST, and T2). Code for the RIM, data loaders,
and documentation can be found through the DIRECT reposi- tory at
https://github.com/directgroup/direct.
III. RESULTS
A. Submission Overview For the 2020 challenge we received a total
of 19 submissions
from eight different groups. Seven groups submitted to the
Multi-Coil 4X and Multi-Coil 8X tracks. One of these groups chose
not to submit to the Transfer track, while an eighth group
submitted only to the Transfer track. As previously, we encourage
all submitting groups to publish papers and code used to generate
their results.
Figure 1 shows an overview of images submitted to the 4X track of
the challenge with Siemens data that were forwarded to
radiologists. All three top performing submissions were able to
successfully reconstruct the T2 and FLAIR images with minimal
artifact presentation. For some images in this track’s evaluation,
radiologists had difficulty perceiving substantive differences
between the three top performing reconstructions in terms of their
overall ability to depict the pathology. Overall, the results were
better on the high signal-to-noise T2 and FLAIR contrasts compared
with those on the T1POST. In the case in Figure 1, the ATB and
Neurospin methods struggled with a strong susceptibility effect,
introducing false vessels between the susceptibility and the
lateral ventricular wall.
Figure 2 shows example images for radiologist evaluation from the
8X track with Siemens data. In this track, artifacts
T1POST T2 FLAIR
A T
in 0.964 0.968 0.913
Fig. 1. Examples of 4X submissions evaluated by radiologists with
slice- level SSIM scores. All methods reasonably reconstructed T2
and FLAIR images. The ATB and Neurospin methods struggled with a
susceptibility region, exaggerating the focus of susceptibility and
introducing a few false vessels between the susceptibility and the
lateral ventricular wall. In other cases, radiologists observed
mild smoothing of white matter regions on T1POST images.
are seen to be more severe and pronounced. For some cases
radiologists stated that they were hesitant to accept any of the
submissions at 8X. Over-smoothing is readily apparent in T1POST
reconstructions from all three of the top performers. We noticed at
this acceleration level that so-called horizontal “banding” effects
[62] could be appreciated in the FLAIR images due to the extreme
acceleration and the anisotropic sampling pattern.
Example images from the 4X Transfer track are shown in Figure 3.
For this track, we observed the lowest SSIM values (Section III-B).
Of note, there is a divergence between performance of methods on GE
versus Philips data. This can be seen the image submitted by
ResoNNance in Figure 3, which introduces artifacts in its
reconstructions of the GE images (T1POST and T2 in Figure 3), but
less so in its Philips reconstruction (FLAIR in Figure 3). Most
participant models (trained on Siemens data) were able to
reconstruct Philips data with higher fidelity than GE, likely due
to the fact that Philips and Siemens followed the same protocol for
writing frequency-oversampled data to their raw data files. An
additional factor is that GE uses a T1-based FLAIR, whereas Philips
and Siemens use a T2-based FLAIR.
This work is licensed under a Creative Commons Attribution 4.0
License. For more information, see
https://creativecommons.org/licenses/by/4.0/
This article has been accepted for publication in a future issue of
this journal, but has not been fully edited. Content may change
prior to final publication. Citation information: DOI
10.1109/TMI.2021.3075856, IEEE Transactions on Medical
Imaging
6 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. XX, NO. XX, XXXX
2020
T1POST T2 FLAIR G
0.904 0.935 0.836
Fig. 2. Examples of 8X submissions evaluated by radiologists with
slice- level SSIM scores. At this level of acceleration fine
details are smoothed and obscured for all contrasts. On T1POST
images, AIRS Medical was relatively more successful than ATB and
Neurospin in showing fine details of the mass, particularly in its
periphery. Noticeable on the FLAIR images are horizontal “banding”
effects that arise from how neural networks interact with
anisotropic sampling patterns.
B. Quantitative Results
Figure 4 shows an overview of SSIM scores across group rankings.
SSIM values were highly clustered in the 4X track, with all top 4
participants scoring between 0.955 and 0.965. We observed greater
variation between submissions in the 8X track, with the top
participant scoring 0.952 and the others scoring below 0.944. The
greatest variation occurred in the Transfer track. Many
participants struggled to adapt their models to GE data. These data
did not include frequency oversampling in the raw k-space data,
which we have observed can decrease SSIMs for models by as much as
0.1-0.4 if no other adjustments are made. On the other hand, the
Philips data did include frequency oversampling, so adaptation here
was more straightforward.
Table III summarizes results by contrast for the finalists in each
competition track with means and 95% confidence intervals based on
2.5% and 97.5% quantiles. The strongest SSIM scores were usually
recorded on T1 post-contrast images (T1POST), while the weakest
scores were typically on FLAIR images. The same participant
recorded the top average SSIM score for every contrast in every
track except the Transfer track for T1 contrast. In this case, two
other participants posted higher SSIM scores.
T1POST T2 FLAIR
0.919 0.930 0.932
Fig. 3. Examples of 4X Transfer submissions evaluated by
radiologists with slice-level SSIM scores. The T1POST and T2
examples are from GE scanners, whereas the FLAIR example is from a
Philips scanner. All methods introduced blurring to the images.
Several methods had trouble adapting to the GE data while
performing relatively well on the Philips data, as seen in the form
of aliasing artifacts in one of the T1POST images.
One team, HungryGrads, submitted to all tracks and received a very
low SSIM score between 0.4 and 0.5. This team set the background
air to nearly 0s, which led to a clinically irrelevant SSIM loss of
approximately 0.3 for their submissions. The HungryGrads submission
prompted our team to perform a post-hoc analysis where we masked
both the submission and the reference RSS ground truth before
calculating SSIM, with results plotted in Figure S1 in the
supplementary material. Applying this mask markedly improved the
SSIM scores of HungryGrads, although it would not have made this
team a finalist. Applying the mask would have enabled ATB to enter
the finalist round for the Transfer track. Our custom mask would
not have changed finalist rankings otherwise.
C. Radiologist Evaluation Results
Radiologist rankings based on quality of pathology depiction were
concordant with SSIM scores for the top submissions as shown in
Figure 5. The second and third place performers for both 4X and 8X
tracks were flipped between the quantitative ranking based on SSIM
and the qualitative ranking based on radiologists. The SSIM
difference between these two constructions methods was relatively
small, out to the third
This work is licensed under a Creative Commons Attribution 4.0
License. For more information, see
https://creativecommons.org/licenses/by/4.0/
This article has been accepted for publication in a future issue of
this journal, but has not been fully edited. Content may change
prior to final publication. Citation information: DOI
10.1109/TMI.2021.3075856, IEEE Transactions on Medical
Imaging
MUCKLEY, RIEMENSCHNEIDER et al.: PREPARATION OF PAPERS FOR IEEE
TRANSACTIONS ON MEDICAL IMAGING 7
TABLE III SUMMARY OF AVERAGE SSIM SCORES WITH 95% CONFIDENCE
INTERVALS
Team AVG T1 T1POST T2 FLAIR
4X Track AIRS Medical 0.964 (0.914, 0.987) 0.967 (0.926, 0.985)
0.969 (0.935, 0.986) 0.965 (0.933, 0.988) 0.930 (0.862, 0.976) ATB
0.960 (0.908, 0.985) 0.964 (0.922, 0.983) 0.965 (0.928, 0.984)
0.961 (0.926, 0.986) 0.924 (0.850, 0.971) Neurospin 0.959 (0.904,
0.984) 0.963 (0.920, 0.982) 0.965 (0.926, 0.984) 0.960 (0.924,
0.986) 0.920 (0.844, 0.970)
8X Track AIRS Medical 0.952 (0.894, 0.979) 0.953 (0.906, 0.979)
0.963 (0.939, 0.981) 0.951 (0.899, 0.978) 0.918 (0.857, 0.973) ATB
0.944 (0.881, 0.973) 0.943 (0.892, 0.973) 0.954 (0.926, 0.977)
0.943 (0.889, 0.973) 0.905 (0.835, 0.965) Neurospin 0.942 (0.874,
0.973) 0.940 (0.888, 0.972) 0.953 (0.924, 0.975) 0.942 (0.887,
0.972) 0.898 (0.824, 0.957)
4X Transfer Track AIRS Medical 0.940 (0.802, 0.989) 0.902 (0.795,
0.983) 0.960 (0.942, 0.969) 0.975 (0.934, 0.992) 0.910 (0.728,
0.979) MRRecon 0.930 (0.846, 0.981) 0.946 (0.910, 0.975) 0.956
(0.935, 0.969) 0.950 (0.889, 0.984) 0.897 (0.671, 0.956) ResoNNance
0.913 (0.791, 0.977) 0.936 (0.892, 0.967) 0.939 (0.920, 0.956)
0.957 (0.911, 0.980) 0.855 (0.695, 0.951)
decimal place. In the Transfer track, radiologist rankings matched
ranking based on SSIM.
TABLE IV SUMMARY OF QUALITY RANKS AND LIKERT SCORES
Team Rank Artifacts Sharpness CNR
4X Track AIRS 1.36 ± 0.64 1.53 ± 0.70 1.53 ± 0.51 1.53 ± 0.51 Nspin
1.94 ± 0.86 1.81 ± 1.01 1.72 ± 0.66 1.75 ± 0.84 ATB 2.22 ± 0.87
1.75 ± 0.97 1.97 ± 0.65 1.86 ± 0.80
8X Track AIRS 1.28 ± 0.64 1.67 ± 0.68 1.89 ± 0.75 1.94 ± 0.75 Nspin
2.25 ± 0.77 1.86 ± 0.83 2.72 ± 0.81 2.28 ± 0.81 ATB 2.28 ± 0.70
1.92 ± 0.94 2.56 ± 0.77 2.42 ± 0.84
4X Transfer Track AIRS 1.11 ± 0.32 1.42 ± 0.50 1.83 ± 0.65 1.81 ±
0.62 MRR 1.97 ± 0.56 1.61 ± 0.55 2.41 ± 0.69 2.22 ± 0.64 Res. 2.78
± 0.54 3.08 ± 0.84 2.86 ± 0.76 3.06 ± 0.86
A summary of the ranks and Likert scores with with means and
standard deviations is shown in Table IV (AIRS = AIRS Medical,
Nspin = Neurospin, MRR = MRRecon, and Res. = ResoNNance). We
applied standard deviations for Table IV (instead of quantiles as
used in Table III) to show more information on the variability.
Across all metrics AIRS Medical separated itself from the other
submissions with the highest SSIM and best image quality. Aside
from this single team, differentiation among the other teams was
not strong. Of note, both the Neurospin and ATB teams had nearly
identical average SSIM scores for the quantitative evaluation, with
ATB presenting a slightly higher score (0.960 vs. 0.959 in 4X,
0.944 vs. 0.942 in 8X). In the radiologist evaluation phase, these
ranks flipped, with Neurospin receiving slightly higher ranks (1.94
vs. 2.22 in 4X, 2.25 vs. 2.28 in 8X).
A case-wise breakdown of the ranks for all 3 finalists and all
rated cases is shown in Figure 5. For second and third- place
metrics as rated by SSIM, radiologist assessment was discordant
between the two methods. However, in 16 out of 18 cases the highest
SSIM score within the finalists’ batches also received the highest
radiologists’ rating. A similar relation - not shown here - was
found for the other used metrics such as normalized mean-squared
error (NMSE) and peak signal-to-
noise ratio (PSNR). Radiologist agreement according to Kendall’s
coefficient of
concordance generally improved as SSIM scores diverged. We
calculated the concordance using radiologist rankings of teams for
quality of depiction of pathology vs. the ground truth. For each
case in the radiologist evaluation phase, we evaluated Kendall’s
coefficient of concordance with tie correction, and then aggregated
over all cases by averaging. This resulted in values of 0.457 for
the 4X track, 0.386 for the 8X track, and 0.781 for the 4X Transfer
track (where 0 indicates complete disagreement and 1 indicates
complete agreement). In the 4X and 8X tracks, discordance was
primarily driven by two submissions (Neurospin and ATB) that were
very close in SSIM score. For the Transfer track, separation among
the teams was more clear, and we observed corresponding increases
in concordance.
Radiologists did take note of hallucinatory effects introduced by
the submission models. Figure 6 shows hallucination examples from
all three tracks. In some cases methods created artifact-mimics. In
other examples, models morphed an abnormality into a more normal
brain structure, such as a sulcus or vessel. Finally, we observed
at least one example combining these two where an artifact was
created at some intermediate layer of a model and then processed by
the remaining portions of the network into a normal structure
mimic.
IV. DISCUSSION
A. Submission Overview In the 2019 challenge all three tracks were
very closely
contested, with little separation between teams either in the
quantitative or the radiologist evaluation phases. We observed this
pattern to be reversed in the 2020 challenge, with one team
assertively scoring the best in all evaluation phases. For some
images in the 4X track, multiple radiologists said that they did
not observe major differentiating aspects affecting the depiction
of pathology in the submissions. However, when averaging the
radiologists’ rankings, radiologists preferred the method that had
the highest-scoring on SSIM from AIRS Medical. We further observed
that the AIRS model scored highest on Likert-type ratings of
artifacts, sharpness, and CNR. This model also provided improvement
over the baseline [14], which had
This work is licensed under a Creative Commons Attribution 4.0
License. For more information, see
https://creativecommons.org/licenses/by/4.0/
This article has been accepted for publication in a future issue of
this journal, but has not been fully edited. Content may change
prior to final publication. Citation information: DOI
10.1109/TMI.2021.3075856, IEEE Transactions on Medical
Imaging
8 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. XX, NO. XX, XXXX
2020
1 2 3 4 5 6 7 Team SSIM Rank
0.80
0.83
0.85
0.88
0.90
0.93
4X SSIM 8x SSIM
1 2 3 4 5 6 7 Team SSIM Rank
0.80
0.83
0.85
0.88
0.90
0.93
0.95
Philips AVG GE
Fig. 4. Summary of SSIM values across contestants. (top) Model
perfomance for teams submitting to the main 4X and 8X Siemens
competition tracks. (bottom) Model performance for teams submitting
to the Transfer track (combination of GE and Philips data). The
“AVG” model score for the Transfer track was a simple average
across all volumes in the Transfer track.
previously been demonstrated for clinical interchangeability at 4X
for knee imaging [48]. Outside of the AIRS model, in the 4X and 8X
tracks the second and third-place models scored very close together
in both the quantitative and the qualitative evaluation phase. In
some cases the SSIM scores for these two models were identical out
to three decimal places.
We observed decreases in performance in the Transfer track. Many
participants struggled to adapt their models to the GE data with
its lack of disk-written frequency oversampling. Although
technically the GE scanner did not operate in any majorly different
way than Philips and Siemens scanners (all use frequency
oversampling), this simple aspect rendered many models useless in
this track without modification. Another factor was a divergence in
FLAIR methodology: our Philips and Siemens data used T2 FLAIR
images, whereas the GE data had T1 FLAIR images. Modifications for
correcting these effects seem not to be straightforward. We note
that as designed the Transfer track primarily evaluated one type of
transfer: generalization across vendors. This was the most
commonly- cited type of transfer in feedback from the 2019
challenge, but future challenges may investigate other types of
transfer.
In terms of radiologist evaluations, despite the drawbacks of SSIM
and RSS ground truths, we observed a correlation
4X Main | 8X Main | 4X Transfer0
Reviewed Individual Cases
1
2
3
M ea n R a d io lo g is t R a ti n g
Overall Rating
Determined Case-Wise:
Highest SSIM
Middle SSIM
Lowest SSIM
Fig. 5. Scatter plot of mean radiologist rank across cases. The
horizontal axis has a separate tick for each case evaluated by the
radiologist cohort. The scatter plot markers indicate whether that
method was from the team with the highest, middle, or lowest SSIM
scores. We generally observed radiologists awarding the best ranks
to models with the best SSIM score.
between radiologist scores and SSIM scores for large SSIM
separations. Multiple radiologists found images at 4X to be similar
in terms of depiction of the pathology, although artifacts tended
to be more problematic in T1POST images. When it came to the 8X and
Transfer tracks, radiologist sentiment became more negative.
Multiple radiologists in both of these tracks offered feedback that
none of the submitted images would be acceptable, indicating that
these two tasks may remain open problems going into 2021.
We note that the results of this paper were observed within the
regime of retrospective undersampling. Retrospective undersampling
does not consider potential differences in signal relaxation along
the echo trains. Even though echo trains can be designed for most
sequences in such a manner that undersampling does not lead to a
change in overall relaxation weighting along the phase dimension,
we have not shown equivalency in our work. We would recommend that
researchers confirm the results of the methods in this paper with
prospective sampling prior to clinical use.
The results of the challenge suggests a few conclusions on
approaches. The first: cascaded models with a data fidelity term
and CNN regularization continue to dominate the submission field,
as occurred in the previous challenge [11]. Second: AIRS, the team
that won all three tracks, had the largest model. However, large
models were not always better, with ATB having a model with similar
performance to Neurospin despite having 87% less parameters.
Lastly, we note the AIRS model used a normalization routine to get
the data into a consistent format for all coil configurations, as
well as being the only team to use a GRAPPA [38] reconstruction as
the initialization.
B. Quantitative Evaluation Process Discussions around the
quantitative evaluation process pri-
marily concerned the presence of background noise during both
This work is licensed under a Creative Commons Attribution 4.0
License. For more information, see
https://creativecommons.org/licenses/by/4.0/
This article has been accepted for publication in a future issue of
this journal, but has not been fully edited. Content may change
prior to final publication. Citation information: DOI
10.1109/TMI.2021.3075856, IEEE Transactions on Medical
Imaging
MUCKLEY, RIEMENSCHNEIDER et al.: PREPARATION OF PAPERS FOR IEEE
TRANSACTIONS ON MEDICAL IMAGING 9
Ground Truth Reconstruction Residual 4X
Tr ac
k 0.964
8X Tr
ac k
k
0.938
Fig. 6. Examples of reconstruction hallucinations among challenge
submissions with SSIM scores over residual plots (residuals
magnified by 5). (top) A 4X submission from Neurospin generated a
false vessel, possibly related to susceptibilities introduced by
surgical staples. (middle) An 8X submission from ATB introduced a
linear bright signal mimicking a cleft of cerebrospinal fluid, as
well as blurring of the boundaries of the extra-axial mass.
(bottom) A submission from ResoNNance introduced a false sulcus or
prominent vessel.
the planning and execution stage of the challenge. The influence of
background noise on SSIM scores is substantial. One participant in
the 2019 challenge had a dedicated style transfer model in order to
add this noise back into the reconstructed images [63]. Despite the
drawbacks to SSIM, we were unable to agree on an alternative for
the 2020 challenge.
In the 2020 challenge, the HungryGrads team submitted images with
backgrounds of nearly zeroes, which penalized their scores.
Prompted by this submission we investigated the effect a masked
metric might have had on their scores in Section III-B. We opted
for a masking algorithm that removes most background pixels and
altered the algorithm parameters for low-SNR edge cases where it
did not perform well. Due to the relatively small size of the
challenge data set visual inspection of the validity of the masks
was feasible. The ranking of the challenge did not change
dramatically due to masking, but masking made metrics less prone to
a specific reconstruction method’s impact on the background.
Another intriguing alternative would be to use alternative
reconstruction techniques such as adaptive combine that implicitly
suppress background noise [47]. We considered using adaptive
combine reconstructions in our evaluation, but concluded that the
results were not particularly meaningful as most models had been
trained with RSS backgrounds. Another alternative to adaptive
combine would be to use other metrics that are more aware of the
noise properties, such as Stein’s Unbiased Risk Estimate
(SURE).
One area lacking in our quantitative analysis was hallu-
cination detection. This is an area of great interest to the
community, but as of the end of our challenge we were unaware of
automated, quantitative methods for detecting lesions or
characterizing stability (although some methods can demonstrate
instability qualitatively [64]). Detection of automated
stability/hallucination analysis remains a topic of great interest
for future challenges.
C. Qualitative Radiologist Evaluation
For the 2020 challenge, we altered the radiologist question- naire
to focus their ranking on the depiction of pathologies rather than
general image quality. Some radiologists found the focus helpful,
commenting specifically that the images at 8X and in the 4X
Transfer track might not be acceptable for clinical use. As this
task aligns more closely with the normal clinical workflow, we
would encourage future competition organizers to use this approach
for their radiologist evaluation procedures.
In the 4X track, there were specific cases where the radiologist
rankings were concordant and others where the rankings were
discordant. Discordant cases tended, upon review, to show that the
main abnormalities were similarly well depicted across the top 3
reconstructions, though there were oftentimes concordant
estimations of differences between reconstructions in terms of
artifacts, sharpness and CNR.
Radiologist sentiment was affected by hallucinations such as those
in Figure 6. Such hallucinatory features are not acceptable and
especially problematic if they mimic normal structures that are
either not present or actually abnormal. These images had high SSIM
scores, indicating that even though these images are considered
well-optimized according to this metric, they are not optimized
regarding hallucination features. Neural network models can be
unstable as demonstrated via adversarial perturbation studies [64].
Despite the lack of realism in some of these perturbations, our
results indicate that hallucination and artifacts remain a real
concern, particularly at higher accelerations. This topic is in
major need for further development.
We note that we did not perform intensity correction for either the
participant submissions or the ground truth recon- structions. None
of the radiologists commented on intensity inhomogeneity in either
the initial testing phase or the final evaluation phase. However,
intensity correction is routinely done by vendor scanners and could
have affected the evaluation, as some pathologies manifest via
varying tissue intensities.
D. Feedback from Participants
We asked participants for feedback regarding challenge or-
ganization and future organization. Participants were generally
enthusiastic about being able to participate in the challenge. We
received positive feedback on our communication via the fastMRI
GitHub repository at https://github.com/ facebookresearch/fastMRI
and the forum associated with the web site at https://fastmri.org.
We also received positive feedback around the challenge’s realism
in focusing on multi-coil data, as well as the challenge’s
generalizability initiative in focusing on the Transfer
track.
This work is licensed under a Creative Commons Attribution 4.0
License. For more information, see
https://creativecommons.org/licenses/by/4.0/
This article has been accepted for publication in a future issue of
this journal, but has not been fully edited. Content may change
prior to final publication. Citation information: DOI
10.1109/TMI.2021.3075856, IEEE Transactions on Medical
Imaging
10 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. XX, NO. XX, XXXX
2020
Still, participants felt the realism could be improved in other
areas. In particular, the sampling mask used for the challenge used
pseudo-regular sampling in order to achieve exact 4X and 8X
sampling rates. This sampling pattern is not equivalent to the
perfectly equidistant sampling pattern used on MRI systems, which
give acceleration rates slightly less than the target rate due to
the densely-sampled center. As a result, challenge models are
likely to require further fine-tuning training before application
to clinical data.
Another point of feedback centered on the storage and compute
resources necessary to participate in the challenge. In the 2019
challenge, the storage aspect was mitigated by the inclusion of the
single-coil track (which had a smaller download size). The
single-coil track attracted a lot of engagement, with 25 out of 33
groups submitting to it [27]. From the compute angle, the trend
towards larger models requires costly hardware. Training the
baseline End-to-End Variational Network [14] requires 32 GPUs, each
with 32 GB of memory, for about 3.5 days. This level of compute
power is not available at many academic centers. By comparison,
multiple participants submitted models trained on only a single
GPU. This was also a topic of feedback from non-participants, with
some telling us informally that they did not participate due to
compute or storage requirements. For the future, researchers felt
it would be helpful for the barriers to entry were lower,
particularly for academic groups that might have innovative methods
but less compute or storage.
As always, the selection of best quantitative evaluation metrics to
use is extremely difficult and there are potential drawbacks to
many or all. Participants did provide feedback concerning the use
of SSIM and the use of RSS for ground truth images. Although groups
acknowledged efforts to seek superior metrics, they felt that
settling for this particular metric was disappointing. Some
participants felt there was a tradeoff between optimizing for SSIM
(which promotes smoothing) vs. radiologist interpretation. Most
vendors have variations in their post-processing pipelines for
precisely this reason. Some vendor post-processing methods even
allow for radiologists to adjust the strength of the
regularization. We did not allow secondary submissions from
participants that might enhance the images for human perception,
such as those based on noise dithering or inspired by stochastic
resonance [65], [48]. Allowing secondary submissions for
radiologist interpretation may be beneficial for future challenges,
provided ground truth images are also included to allow
radiologists to watch for hallucination. In compiling the results
for this challenge we have attempted to investigate some other
options that would at least mitigate the effects of background
noise and feel that this is an important topic for further
investigation. Consensus around evaluations–for ground truth
calculations, metrics, and radiologist presentation–would
substantially aid the organization of future challenges.
V. CONCLUSION
The 2020 fastMRI reconstruction challenge featured two core
modifications from its 2019 predecessor: 1) a new competition
Transfer track to evaluate model generalization and 2)
adjusting
the radiologist evaluation to focus on pathology depiction. In
addition to these, we extended our competition to a new anatomy
with much larger data sets for both training and competition
evaluation. The competition resulted in a new state-of-the-art
model. Our challenge confirmed areas in need of research,
particularly those along the lines of evaluation metrics, error
characterization, and AI-generated hallucinations. Radiologist
sentiment was mixed for images submitted to the 8X and the Transfer
tracks; these may remain open research frontiers going into 2021.
We hope that researchers and future challenge organizers find the
results of the 2020 fastMRI challenge helpful in their future
endeavors.
ACKNOWLEDGMENT
We would like to thank all research groups that participated in the
challenge. We thank the scientific advisors that gave feedback
during challenge organization. We thank Girish M. Fatterpekar, Nael
Kambiz, Amy L. Kotsenas, Christie Mary Lincoln, Joseph Mettenburg,
and Max Wintermark for judging the submissions and the contribution
of their clinical expertise and time. We would also like to thank
Tullie Murrell, Mark Tygert, and C. Lawrence Zitnick for many
valuable contributions to the launch and execution of this
challenge. We give special thanks to Philips North America and
their clinical partner sites for the data they contributed for use
as part of the Transfer track. The NYU authors thank NIH grants
P41EB017183, R01EB024532, and R21EB027241 for financial support.
The ResoNNance team’s submission was conducted under the STAIRS
project under the TKI-PPP program. The ResoNNance collaboration
project is co-funded by the PPP Allowance made available by Health
Holland, Top Sector Life Sciences & Health, to stimulate
public-private partnerships.
REFERENCES
[1] S. Chetlur, C. Woolley, P. Vandermersch, J. Cohen, J. Tran, B.
Catanzaro, and E. Shelhamer, “cuDNN: Efficient primitives for deep
learning,” arXiv preprint arXiv:1410.0759, 2014.
[2] M. Abadi, P. Barham, J. Chen, Z. Chen, A. Davis, J. Dean, M.
Devin, S. Ghemawat, G. Irving, M. Isard, M. Kudlur, J. Levenberg,
R. Monga, S. Moore, D. G. Murray, B. Steiner, P. Tucker, V.
Vasudevan, P. Warden, M. Wicke, Y. Yu, and X. Zheng, “TensorFlow: A
system for large-scale machine learning,” in Proc. USENIX Conf.
Operating Sys. Design and Impl., 2016, pp. 265–283.
[3] A. Paszke, S. Gross, F. Massa, A. Lerer, J. Bradbury, G.
Chanan, T. Killeen, Z. Lin, N. Gimelshein, L. Antiga, A. Desmaison,
A. Kopf, E. Yang, Z. DeVito, M. Raison, A. Tejani, S. Chilamkurthy,
B. Steiner, L. Fang, J. Bai, and S. Chintala, “PyTorch: An
imperative style, high- performance deep learning library,” in Adv.
Neural Inf. Process. Sys., 2019, pp. 8026–8037.
[4] Y. Yang, J. Sun, H. Li, and Z. Xu, “Deep ADMM-Net for
compressive sensing MRI,” in Adv. Neural Inf. Process. Sys., 2016,
pp. 10–18.
[5] K. Hammernik, T. Klatzer, E. Kobler, M. P. Recht, D. K.
Sodickson, T. Pock, and F. Knoll, “Learning a variational network
for reconstruction of accelerated MRI data,” Magn. Res. Med., vol.
79, no. 6, pp. 3055–3071, 2018.
[6] J. Schlemper, J. Caballero, J. V. Hajnal, A. N. Price, and D.
Rueckert, “A deep cascade of convolutional neural networks for
dynamic MR image reconstruction,” IEEE Trans. Med. Imag., vol. 37,
no. 2, pp. 491–503, 2017.
[7] G. Yang, S. Yu, H. Dong, G. Slabaugh, P. L. Dragotti, X. Ye, F.
Liu, S. Arridge, J. Keegan, Y. Guo, and D. Firmin, “DAGAN: Deep de-
aliasing generative adversarial networks for fast compressed
sensing MRI reconstruction,” IEEE Trans. Med. Imag., vol. 37, no.
6, pp. 1310– 1321, 2017.
This work is licensed under a Creative Commons Attribution 4.0
License. For more information, see
https://creativecommons.org/licenses/by/4.0/
This article has been accepted for publication in a future issue of
this journal, but has not been fully edited. Content may change
prior to final publication. Citation information: DOI
10.1109/TMI.2021.3075856, IEEE Transactions on Medical
Imaging
MUCKLEY, RIEMENSCHNEIDER et al.: PREPARATION OF PAPERS FOR IEEE
TRANSACTIONS ON MEDICAL IMAGING 11
[8] T. Eo, Y. Jun, T. Kim, J. Jang, H.-J. Lee, and D. Hwang,
“KIKI-net: Cross- domain convolutional neural networks for
reconstructing undersampled magnetic resonance images,” Magn. Res.
Med., vol. 80, no. 5, pp. 2188– 2201, 2018.
[9] H. K. Aggarwal, M. P. Mani, and M. Jacob, “MoDL: Model-based
deep learning architecture for inverse problems,” IEEE Trans. Med.
Imag., vol. 38, no. 2, pp. 394–405, 2018.
[10] B. Zhu, J. Z. Liu, S. F. Cauley, B. R. Rosen, and M. S. Rosen,
“Image reconstruction by domain-transform manifold learning,”
Nature, vol. 555, no. 7697, pp. 487–492, 2018.
[11] F. Knoll, K. Hammernik, E. Kobler, T. Pock, M. P. Recht, and
D. K. Sodickson, “Assessment of the generalization of learned image
reconstruction and the potential for transfer learning,” Magn. Res.
Med., vol. 81, no. 1, pp. 116–128, 2019.
[12] F. Knoll, K. Hammernik, C. Zhang, S. Moeller, T. Pock, D. K.
Sodickson, and M. Akcakaya, “Deep-learning methods for parallel
magnetic resonance imaging reconstruction: A survey of the current
approaches, trends, and issues,” IEEE Signal Process. Mag., vol.
37, no. 1, pp. 128–140, 2020.
[13] B. Yaman, S. A. H. Hosseini, S. Moeller, J. Ellermann, K.
Ugurbil, and M. Akcakaya, “Self-supervised learning of
physics-guided reconstruction neural networks without fully sampled
reference data,” Magn. Res. Med., vol. 84, no. 6, pp. 3172–3191,
2020.
[14] A. Sriram, J. Zbontar, T. Murrell, A. Defazio, C. L. Zitnick,
N. Yakubova, F. Knoll, and P. Johnson, “End-to-end variational
networks for accelerated MRI reconstruction,” in Intl. Conf. Med.
Imag. Comput. Comput.-Assisted Intervention, 2020, pp. 64–73.
[15] K. Fukushima and S. Miyake, “Neocognitron: A self-organizing
neural network model for a mechanism of visual pattern
recognition,” in Compet. Coop. Neural Nets, 1982, pp.
267–285.
[16] Y. LeCun, L. Bottou, Y. Bengio, and P. Haffner,
“Gradient-based learning applied to document recognition,” Proc.
IEEE, vol. 86, no. 11, pp. 2278– 2324, 1998.
[17] R. Raina, A. Madhavan, and A. Y. Ng, “Large-scale deep
unsupervised learning using graphics processors,” in Proc. Intl.
Conf. Mach. Learning, 2009, pp. 873–880.
[18] D. C. Ciresan, U. Meier, L. M. Gambardella, and J.
Schmidhuber, “Deep, big, simple neural nets for handwritten digit
recognition,” Neural Comput., vol. 22, no. 12, pp. 3207–3220,
2010.
[19] J. Deng, W. Dong, R. Socher, L.-J. Li, K. Li, and L. Fei-Fei,
“ImageNet: A large-scale hierarchical image database,” in Proc.
IEEE Conf. Comput. Vis. Pattern Recogn., 2009, pp. 248–255.
[20] A. Krizhevsky, I. Sutskever, and G. E. Hinton, “ImageNet
classification with deep convolutional neural networks,” in Adv.
Neural Inf. Process. Sys., 2012, pp. 1097–1105.
[21] K. Simonyan and A. Zisserman, “Very deep convolutional
networks for large-scale image recognition,” in Intl. Conf.
Learning Repr., 2015.
[22] K. He, X. Zhang, S. Ren, and J. Sun, “Deep residual learning
for image recognition,” in Proc. IEEE Conf. Comput. Vis. Pattern
Recogn., 2016, pp. 770–778.
[23] C. Szegedy, V. Vanhoucke, S. Ioffe, J. Shlens, and Z. Wojna,
“Rethinking the Inception architecture for computer vision,” in
Proc. IEEE Conf. Comput. Vis. Pattern Recogn., 2016, pp.
2818–2826.
[24] G. Huang, Z. Liu, L. Van Der Maaten, and K. Q. Weinberger,
“Densely connected convolutional networks,” in Proc. IEEE Conf.
Comput. Vis. Pattern Recogn., 2017, pp. 4700–4708.
[25] J. Zbontar, F. Knoll, A. Sriram, T. Murrell, Z. Huang, M. J.
Muckley, A. Defazio, R. Stern, P. Johnson, M. Bruno, M. Parente, K.
J. Geras, J. Katsnelson, H. Chandarana, Z. Zhang, M. Drozdzal, A.
Romero, M. Rabbat, P. Vincent, N. Yakubova, J. Pinkerton, D. Wang,
E. Owens, C. L. Zitnick, M. P. Recht, D. K. Sodickson, and Y. W.
Lui, “fastMRI: An open dataset and benchmarks for accelerated MRI,”
arXiv preprint arXiv:1811.08839, 2018.
[26] F. Knoll, J. Zbontar, A. Sriram, M. J. Muckley, M. Bruno, A.
Defazio, M. Parente, K. J. Geras, J. Katsnelson, H. Chandarana, Z.
Zhang, M. Drozdzal, A. Romero, M. Rabbat, P. Vincent, J. Pinkerton,
D. Wang, N. Yakubova, E. Owens, C. L. Zitnick, M. P. Recht, D. K.
Sodickson, and Y. W. Lui, “fastMRI: A publicly available raw
k-space and DICOM dataset of knee images for accelerated MR image
reconstruction using machine learning,” Radiol.: Artificial
Intell., vol. 2, no. 1, p. e190007, 2020.
[27] F. Knoll, T. Murrell, A. Sriram, N. Yakubova, J. Zbontar, M.
Rabbat, A. Defazio, M. J. Muckley, D. K. Sodickson, C. L. Zitnick,
and M. P. Recht, “Advancing machine learning for MR image
reconstruction with an open competition: Overview of the 2019
fastMRI challenge,” Magn. Res. Med., vol. 84, no. 6, pp. 3054–3070,
2020.
[28] W. A. Grissom, K. Setsompop, S. A. Hurley, J. Tsao, J. V.
Velikina, and A. A. Samsonov, “Advancing RF pulse design using an
open-competition format: Report from the 2015 ISMRM challenge,”
Magn. Res. Med., vol. 78, no. 4, pp. 1352–1361, 2017.
[29] K. G. Schilling, A. Daducci, K. Maier-Hein, C. Poupon, J.-C.
Houde, V. Nath, A. W. Anderson, B. A. Landman, and M. Descoteaux,
“Chal- lenges in diffusion MRI tractography–lessons learned from
international benchmark competitions,” Magn. Res. Imag., vol. 57,
pp. 194–209, 2019.
[30] O. Maier, S. H. Baete, A. Fyrdahl, K. Hammernik, S. Harrevelt,
L. Kasper, A. Karakuzu, M. Loecher, F. Patzig, Y. Tian, K. Wang, D.
Gallichan, M. Uecker, and F. Knoll, “CG-SENSE revisited: Results
from the first ISMRM reproducibility challenge,” Magn. Res. Med.,
2020.
[31] Y. Beauferris, J. Teuwen, D. Karkalousos, N. Moriakov, M.
Caan, L. Rodrigues, A. Lopes, H. Pedrini, L. Rittner, M. Dannecker,
V. Stu- denyak, F. Groger, D. Vyas, S. Faghih-Roohi, A. K. Jethi,
J. C. Raju, M. Sivaprakasam, W. Loos, R. Frayne, and R. Souza,
“Multi-channel MR reconstruction (MC-MRRec) challenge – comparing
accelerated MR reconstruction models and assessing their
generalizability to datasets collected with different coils,” arXiv
preprint arXiv:2011.07952, 2020.
[32] E. T. Whittaker, “XVIII.—On the functions which are
represented by the expansions of the interpolation-theory,” Proc.
Royal Soc. Edinburgh, vol. 35, pp. 181–194, 1915.
[33] H. Nyquist, “Certain topics in telegraph transmission theory,”
Trans. Am. Inst. Electr. Eng., vol. 47, no. 2, pp. 617–644,
1928.
[34] V. A. Kotelnikov, “On the carrying capacity of the ‘ether’ and
wire in telecommunications,” in Mat. First All-Union Conf. on
Questions of Comm. (Russian), 1933.
[35] C. E. Shannon, “Communication in the presence of noise,” Proc.
IRE, vol. 37, no. 1, pp. 10–21, 1949.
[36] D. K. Sodickson and W. J. Manning, “Simultaneous acquisition
of spatial harmonics (SMASH): Fast imaging with radiofrequency coil
arrays,” Magn. Res. Med., vol. 38, no. 4, pp. 591–603, 1997.
[37] K. P. Pruessmann, M. Weiger, M. B. Scheidegger, and P.
Boesiger, “SENSE: Sensitivity encoding for fast MRI,” Magn. Res.
Med., vol. 42, no. 5, pp. 952–962, 1999.
[38] M. A. Griswold, P. M. Jakob, R. M. Heidemann, M. Nittka, V.
Jellus, J. Wang, B. Kiefer, and A. Haase, “Generalized
autocalibrating partially parallel acquisitions (GRAPPA),” Magn.
Res. Med., vol. 47, no. 6, pp. 1202–1210, 2002.
[39] M. Lustig, D. L. Donoho, J. M. Santos, and J. M. Pauly,
“Compressed sensing MRI,” IEEE Sig. Process. Mag., vol. 25, no. 2,
pp. 72–82, 2008.
[40] M. Uecker, P. Lai, M. J. Murphy, P. Virtue, M. Elad, J. M.
Pauly, S. S. Vasanawala, and M. Lustig, “ESPIRiT—an eigenvalue
approach to autocalibrating parallel MRI: Where SENSE meets
GRAPPA,” Magn. Res. Med., vol. 71, no. 3, pp. 990–1001, 2014.
[41] O. Ronneberger, P. Fischer, and T. Brox, “U-Net: Convolutional
networks for biomedical image segmentation,” in Intl. Conf. Med.
Imag. Comput. Comput.-Assisted Intervention, 2015, pp.
234–241.
[42] N. Pezzotti, S. Yousefi, M. S. Elmahdy, J. V. Gemert, C.
Schulke, M. Doneva, T. Nielsen, S. Kastryulin, B. P. F. Lelieveldt,
M. V. Osch, E. Weerdt, and M. Staring, “An adaptive intelligence
algorithm for undersampled knee MRI reconstruction,” IEEE Access,
vol. 8, pp. 204 825–204 838, 2020.
[43] P. Wang, E. Z. Chen, T. Chen, V. M. Patel, and S. Sun,
“Pyramid convo- lutional RNN for MRI reconstruction,” arXiv
preprint arXiv:1912.00543, 2019.
[44] P. Putzky, D. Karkalousos, J. Teuwen, N. Miriakov, B. Bakker,
M. Caan, and M. Welling, “i-RIM applied to the fastMRI challenge,”
arXiv preprint arXiv:1910.08952, 2019.
[45] Z. Wang, A. C. Bovik, H. R. Sheikh, and E. P. Simoncelli,
“Image quality assessment: from error visibility to structural
similarity,” IEEE Trans. Imag. Process., vol. 13, no. 4, pp.
600–612, 2004.
[46] P. B. Roemer, W. A. Edelstein, C. E. Hayes, S. P. Souza, and
O. M. Mueller, “The NMR phased array,” Magn. Res. Med., vol. 16,
no. 2, pp. 192–225, 1990.
[47] D. O. Walsh, A. F. Gmitro, and M. W. Marcellin, “Adaptive
reconstruction of phased array MR imagery,” Magn. Res. Med., vol.
43, no. 5, pp. 682– 690, 2000.
[48] M. P. Recht, J. Zbontar, D. K. Sodickson, F. Knoll, N.
Yakubova, A. Sriram, T. Murrell, A. Defazio, M. Rabbat, L. Rybak,
M. Kline, G. Ciavarra, E. F. Alaia, M. Samim, W. R. Walter, D. J.
Lin, Y. W. Lui, M. Muckley, Z. Huang, P. Johnson, R. Stern, and C.
L. Zitnick, “Using deep learning to accelerate knee MRI at 3T:
Results of an interchangeability study,” Am. J. Roentgenology, vol.
215, no. 6, pp. 1421–1429, 2020.
This work is licensed under a Creative Commons Attribution 4.0
License. For more information, see
https://creativecommons.org/licenses/by/4.0/
This article has been accepted for publication in a future issue of
this journal, but has not been fully edited. Content may change
prior to final publication. Citation information: DOI
10.1109/TMI.2021.3075856, IEEE Transactions on Medical
Imaging
12 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. XX, NO. XX, XXXX
2020
[49] S. Van der Walt, J. L. Schonberger, J. Nunez-Iglesias, F.
Boulogne, J. D. Warner, N. Yager, E. Gouillart, and T. Yu,
“scikit-image: image processing in Python,” PeerJ, vol. 2, p. e453,
2014.
[50] D. P. Kingma and J. Ba, “Adam: A method for stochastic
optimization,” in Intl. Conf. Learning Repr., 2015.
[51] Y. Jun, H. Shin, and D. Hwang, “Joint deep model-based MR
image and coil sensitivity reconstruction network (Joint-ICNet) for
fast MRI,” in Proc. IEEE/CVF Conf. Comput. Vis. Pattern Recogn.,
June 2021, in press.
[52] S. Yu, B. Park, and J. Jeong, “Deep iterative down-up CNN for
image denoising,” in Proc. IEEE/CVF Conf. Comput. Vis. Pattern
Recogn. Workshops, June 2019.
[53] Z. Ramzi, P. Ciuciu, and J.-L. Starck, “Benchmarking MRI
reconstruction neural networks on large public datasets,” App.
Sci., vol. 10, no. 5, p. 1816, 2020.
[54] Z. Ramzi, J.-L. Starck, and P. Ciuciu, “XPDNet for MRI
reconstruction: An application to the 2020 fastMRI challenge,” in
Proc. Intl. Soc. Magn. Res. Med., 2021, in press.
[55] A. Chambolle and T. Pock, “A first-order primal-dual algorithm
for convex problems with applications to imaging,” J. of
Mathematical Imag. Vis., vol. 40, no. 1, pp. 120–145, 2011.
[56] J. Adler and O. Oktem, “Learned primal-dual reconstruction,”
IEEE Trans. Med. Imag., vol. 37, no. 6, pp. 1322–1332, 2018.
[57] P. Liu, H. Zhang, K. Zhang, L. Lin, and W. Zuo, “Multi-level
Wavelet- CNN for image restoration,” in Proc. IEEE Conf. Comput.
Vis. Pattern Recogn. Workshops, 2018, pp. 773–782.
[58] L. Liu, H. Jiang, P. He, W. Chen, X. Liu, J. Gao, and J. Han,
“On the variance of the adaptive learning rate and beyond,” in
Intl. Conf. Learning Repr., 2020.
[59] P. Putzky and M. Welling, “Recurrent inference machines for
solving inverse problems,” arXiv preprint arXiv:1706.04008,
2017.
[60] K. Lønning, P. Putzky, J.-J. Sonke, L. Reneman, M. W. Caan,
and M. Welling, “Recurrent inference machines for reconstructing
heteroge- neous MRI data,” Med. Imag. Anal., vol. 53, pp. 64–78,
2019.
[61] M. Uecker, F. Ong, J. I. Tamir, D. Bahri, P. Virtue, J. Y.
Cheng, T. Zhang, and M. Lustig, “Berkeley advanced reconstruction
toolbox,” in Proc. Intl. Soc. Magn. Res. Med., 2015.
[62] A. Defazio, T. Murrell, and M. P. Recht, “MRI banding removal
via adversarial training,” in Adv. Neural Inf. Process. Sys.,
2020.
[63] K. Hammernik, J. Schlemper, C. Qin, J. Duan, R. M. Summers,
and D. Rueckert, “Σ-net: Systematic evaluation of iterative deep
neural networks for fast parallel MR image reconstruction,” arXiv
preprint arXiv:1912.09278, 2019.
[64] V. Antun, F. Renna, C. Poon, B. Adcock, and A. C. Hansen, “On
instabilities of deep learning in image reconstruction and the
potential costs of AI,” Proc. Natl. Acad. Sci., vol. 117, no. 48,
pp. 30 088–30 095, 2020.