Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 20041
Radiation Directivity of
Human and Artificial Speech
Teemu Halkosaari 1 & Markus Vaalgamaa 2
1 Helsinki University of Technology, Laboratory of Acoustics and Audio Signal Processing
2 Nokia Technology, Audio Quality
e-mail: markus.vaalgamaa@nokia.com
Workshop of Wideband Speech Quality in Terminals and Networks: Assessment and Prediction, 8th-9th June
2004, Mainz, Germany
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 20042
Agenda
1. Motivation and Background
2. Goal
3. Measurements
4. Analyzes
5. Modelling
6. Results
7. Improvements to telephonometry
8. Conclusions
Part I: WHY?
Part II: The STUDY
Part III: The OUTCOME
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 20043
Part I
WHY?
1. Motivation and Background
2. Goal
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 20044
1. Motivation and Background
• Artificial mouths and HATS were originally designed for narrowband speech measurements with big landline handsets.
• In past years, the size of phone has become smaller and use of headsets is becoming common.
• In future, wideband phones are coming…
• It seems that HATS and artificial mouths will be used in future, although
• we do not know exactly the directivities of artificial mouths or• human mouth especially on near field locations
• Additional interesting topic is that how does the speech content affect to the directivity?
• Extensive literature on the subject is not available• Standardization is inadequate
• ITU-T P.58 Head and torso simulator for telephonometry• ITU-T P.51 Arficial mouth• narrow band: 300-3400Hz, wide band: 150-7000Hz
Part I: WHY?
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 20045
2. Goal
• Measure frequency responses for artificial mouths to positions where microphones of handsets and headsets are situated.
• Repeat the same measurements for a group of test subjects
• Most important part of the study is to
analyze the difference between artificial mouths and an average person.
• In practice a set of transfer functions is acquired.
Part I: WHY?
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 20046
Part II
The STUDY
3. Measurements
4. Analyzes
5. Modelling
6. Results
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 20047
• Two separate systems• Test subjects were measured in Helsinki University of Technology,
anechoic chamber, 8 channel parallel recording system.• Artificial mouths were measured in Nokia facilities in Salo, Finland,
anechoic chamber, Audio Precision APwin measurement system, 2 channel parallel MLS measurement.
• Probes• B&K 1/8 and ¼ inch measurement microphones• 8 Sennheiser KE-4-211-2 electret microphones• In addition, some phones: large and small handset, and accessories:
boom headset, mono and stereo headsets, were measured
3. Measurements: ArrangementsPart II: The STUDY
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 20048
3. Measurements: Positions
• Referred to• lip plane and its perpendicular• on chest: axes from throat downwards and sidewards
• Two reference positions for transfer functions• MRP = 25 mm in front of mouth for artificial mouths• 0.5 m in front of mouth for test subjects.
• Measurement positions:
Part II: The STUDY
Near cheek In front of mouth On chest
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 20049
3. Measurements: Setup for artificial mouths
Part II: The STUDY
• Two B&K measurement microphones.
• Two impulse responses parallel (reference and point of interest)
• Apwin measurement system, MLS
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200410
3. Measurements: Positioners
• Helmet and chest positioner were developed to keep microphones in right positions.
Part II: The STUDY
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200411
3. Measurements: Setup for test subjects (1)
IOtechWaveBook/516
LapTopcomputer
pre-amp E.A.A.2 channels
pre-amp E.A.A.2 channels
pre-amp E.A.A.2 channels
E.A.A. pre-amp..
8 channels
Biasing unit
. .
8 ch
anne
ls
Mixer
DisplayVHSRecorder
Camera
Loudspeaker
.
.
8 channels
Test subject or HATS
A/V-monitoring andTest Subject guidance
Data acquisition system
Microphones
Monitoring roomAnechoic chamber
LPT
Part II: The STUDY
• Parallel 8 channel audio recording at 32kHz sample rate
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200412
3. Measurements : Setup for test subjects (2)
Part II: The STUDY
• Small anechoic chamber at Helsinki University of Technology and monitoring system
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200413
3. Measurements: Test subjects
• Test subjects: • 8 male, 5 female, mainly young students, Finnish as mother tongue
• Speech material:• ”Kaksi vuotta sitten kävimme ravintola Gabrielissa Helsingissä ja
söimme siellä padallisen fasaania banaanilla höystettynä” (includes all Finnish phonemes)
• Translation: “Two years ago we went to restaurant Gabriel and we ate there a pot of pheasant larded with banana “
• Separate phonemes \n m s r\ and \a e i o u y ä ö\• Different speech volumes: loud, normal, and silent
Part II: The STUDY
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200414
4. Analyzes
• Starting point:• parallel 8-channel 32kHz 16bit raw audio data.• Impulse responses in measurements for artificial mouths -> directly
converted to transfer functions.
• Phonemes and active speech were segmented manually (CoolEdit Pro, around 10000 manual marks by the way).
• All analyzes were made in Matlab environment:• SNR and Coherences to ensure reliability• Power spectra estimates• Transfer function estimates• Confidential intervals consideration for transfer functions
• FFT 1024 samples that corresponds to 32 ms, Hanning-window, 50% overlap
• Averaging and Smoothed to 1/3-octave bands• Speech was averaged weighting with coherence
Part II: The STUDY
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200415
5. Modelling
• In addition to measurements two models were built based on acoustic field theory
• Results were validated by modelling: Sphere and piston source• Head = sphere• Mouth = radially vibrating round surface of the sphere (piston).
• An infinite baffle can be added to simulate the upper body.• Mirror source method
• Numerical implementation in Matlab
0R 0
u 0
Part II: The STUDY
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200416
6. Results: HATS Nearfield
• Measured and modelled results for B&K 4128 HATS• Transfer functions from Mouth Reference Position (MRP) to other positions
Part II: The STUDY
150 1000 7000-25
-20
-15
-10
-5
0
Hz
dB
150 1000 7000-30
-25
-20
-15
-10
-5
Hz
dB
Transfer functions from MRP to near cheek positions:• Large handset (blue circles)• Small handset (red squares)• Boom headset (green triangles)• Corresponding models (black dashed)
Transfer functions from MRP to chest positions:• Near throat with vest (blue circles, solid)• Near throat without (blue circles, dashed)• Middle of chest with vest (red squares, solid)• Middle of chest without vest (red squares, dashed)• Corresponding models (black dashed)
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200417
6. Results: HATS Far field (1)
• Measured results with B&K 4128 HATS
• Transfer function from MRP to positions in the front of mouth:• 10 cm (blue circles)• 25 cm (red squares)• 50 cm (green triangles)• 100 cm (violet quadrangles)• with (solid) and without vest (dotted)
Part II: The STUDY
150 1000 7000
-25
-20
-15
-10
-5
0
Hz
dB
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200418
6. Results: HATS Far field (2)
• Measured and modelled for B&K 4128 HATS
• Transfer function for positions in front of mouth on 50 cm distance.• Vest on (blue circles)• Vest off (red squares)• Bare head, torso unattached (green triangles)• Model with the baffle (black dotted, thin)• Model without the baffle (black dasked, thick)
Part II: The STUDY
150 1000 7000-28
-26
-24
-22
-20
-18
-16
Hz
dB
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200419
6. Results: Human – B&K HATS
• Comparison between transfer functions, referred to 0.5m in front of mouth position. Positions, colors and symbols are same as in the slide 16.
• TFHuman / TFHATS• 1/3 mouth cross-section area ratio in models
• 0 dB values indicate that there is no difference between human and HATS
• Positive values indicate that HATS is more directive, i.e. has less SPL on measured microphone positions on next to cheek or on chest.
Next to cheek On chest
Part II: The STUDY
150 1000 7000
-2
0
2
4
6
8
10
12
Hz
dB
150 1000 7000
-2
0
2
4
6
8
10
12
Hz
dB
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200420
6. Results: B&K 4227 and Head Acoustics HMS II.3 HATS
• Comparison between transfer functions, referred to 0.5m position• TFHuman / TFHATS
• Positive values indicate that HATS is more directive, i.e. has less SPL on measured microphone positions on side.
• Transfer functions from MRP to near cheek positions: Large handset (blue circles), Small handset (red squares), and Boom headset (green triangles)
Next to cheek for B&K 4227 Next to cheek for HA HATS
Part II: The STUDY
150 1000 7000-2
0
2
4
6
8
10
12
Hz
dB
150 1000 7000-2
0
2
4
6
8
10
12
Hz
dB
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200421
6. Results: Speech content
• Speech content changes the directivity pattern• Speech style: loud – silent etc..• Articulation: continuous phonation – natural speech• Phonemes (Smoothed Human/HATS difference for mic. pos. 1.3 below
with 95% confidence intervals)
• Mouth aperture size = steepness of the response on high frequencies
• Statistical approach: ANOVA model did not apply = speech data scattered
Part II: The STUDY
150 1000 7000-2
0
2
4
6
8
Hz
dB
TFs human vs. HATS by vowel groups
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200422
6. Results: Reliability
• Beforehand by careful design of the measurement processes
• Reliability of human measurements:• SNR and Coherence (estimates)
• SNR on wide-band: >25dB• Coherence on narrow-band: >0.8, on wide-band: >0.65
• Normal distribution hypothesis• 95% confidential intervals on wide-band: <±0.5dB
• Reliability of HATS measurements is better than for humans i.e. very good (and thus omitted here)
Part II: The STUDY
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200423
Part III
The OUTCOME
7. Improvements to telephonometry
8. Conclusions
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200424
150 1000 70000
2
4
6
8
10
12
Hz
dB
150 1000 70000
2
4
6
8
10
Hz
dB
7. Improvement proposal: Equalization (1)
• Artificial mouths (especially B&K 4128) does not correspond to average test subject
• Difference curves are ”well-behaving” for B&K HATS -> A low order filter could equalize the difference near cheek.
• Example: Yulewalk recursive IIR filter design (least-squares method) was applied in Matlab for smoothed and averaged differences (B&K HATS).
1.4
1.3
1.1
Average of curves on right (blue)IIR N = 3 (red)IIR N = 7 (green)
Part III: The OUTCOME
Difference and 95% confidence limits between human and B&K HATS near cheek positions: Large handset (blue circles), Small handset (red squares), and Boom headset (green triangles)
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200425
7. Improvement proposal: Equalization (2)
• Smoothed difference curves and 95% confidence limits between human - B&K 4227 (left) and Human - Head Acoustics HMS II.3 (right) on near cheek positions
• Large handset (blue circles)• Small handset (red squares)• Boom headset (green triangles)
Part III: The OUTCOME
150 1000 70000
2
4
6
8
10
12
Hz
dB
150 1000 70000
2
4
6
8
10
12
Hz
dB
Human - B&K 4227 Human - HA HATS
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200426
7. Improvement proposal: Mouth size
• B&K 4128 HATS
• Positions 1.3 (left) and 1.1 (right) where measured with 3 different mouth sizes
• small, 15 mm x 11 mm, adaptor partly blocked (blue circles)• normal, 30 mm x 11 mm, adaptor (red squares)• big, 42 mm x 16 mm, mouth adaptor removed (green triangles)
Part III: The OUTCOME
150 1000 70000
5
10
15
20
Hz
dB
150 1000 70000
5
10
15
20
Hz
dB
Position 1.3 Human - B&K HATS Position 1.1 Human – B&K HATS
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200427
7. Improvement proposal: Vest
• About 2cm thick vest should be used in B&K HATS measurements. How does it change the directivity in position on the chest?
• Let’s take off the vest and compare to average test subject (dashed lines)
• The near-chest reflection shifts to higher frequency
• Difference between human and HATS measurements on chest positions:• Near throat with (blue circles, solid) and without vest (blue circles, dashed)• Middle of chest with (red squares, solid) and without vest (red squares,
dashed)
Part III: The OUTCOME
vest on (solid)
vest off (dashed)
150 1000 7000-2
0
2
4
6
8
10
12
14
Hz
dB
4.2
4.1
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200428
8. Conclusions
• Key results:• On high frequencies B&K HATS is too directional.• Near cheek: the difference (Human/Artificial mouth) is slighly
increasing by moving away from the lip plane. • On averate the difference between human and artificial mouth is
between 1 and 6 dB.• Near chest: the differences in chest reflections are significant.• Speech content affects the directivity.
• Reasons:• Mouth cross-section area during speech is smaller than in artificial
mouths. Speech content and mouth cross-section area are linked.• The torso and head design of the artificial mouths do not correspond
to the acoustical characteristics of a real body.
• What could be improved?• Near cheek: simple equalization was proposed.• Improvement of the structure of HATS (mouth size, torso, etc.). This
path requires more measurements and analysis.• Vest usage should always be considered carefully.
Part III: The OUTCOME
Teemu Halkosaari & Markus Vaalgamaa: Radiation Directivity of Human and Artificial Speech 08-June-04© Nokia 200429
Agenda
1. Motivation and Background
2. Goals
3. Measurements
4. Analyzes
5. Modelling
6. Results
7. Improvements to telephonometry
8. Conclusions
Part I: WHY?
Part II: The STUDY
Part III: The OUTCOME