Quick Start Guide - Dexcom › sites › dexcom.com › files › cgm... · 2014-06-04 · the flat...

Post on 05-Jul-2020

3 views 0 download

Transcript of Quick Start Guide - Dexcom › sites › dexcom.com › files › cgm... · 2014-06-04 · the flat...

Inserting the Sensor (continued)f. Hold the applicator, and pull the

safety lock straight out away from the applicator, in the direction of the arrows in the picture. Give the safety lock to the patient to help remove the transmitter at the end of the sensor session if appropriate.

g. Placethefingersofonehandattheedgeofthewhiteadhesive(attheopposite side of the sensor from the transmitter latch). You may pinch up on your skin using this hand.

h. While still pinching, use your other handtoplacetwofingersabovethecollar(seepicture).Placeyourthumbon the white plunger and push down completely. You should hear 2 clicks.

i. Moveyour2fingersfromabovethecollartobelowthecollar.Keepyourthumblightlyontopofthewhiteplungerandpullthecollarbacktowardsyourthumbuntilyou hear 2 clicksorcannotpullbackanymore.

j. Makesurethetransmitterlatchisdown(againstyourbody).Squeezethecenteroftheribbedtabsonthesidesofthesensorpod.Rocktheapplicatorbarrelforwardandoutawayfromyourbody.

Attaching the Transmitter:a. Cleanthebackofthetransmitterwithanalcoholwipe.Letdry.b. Placethetransmitterinthesensorpod(with

theflatsidedown,andthenarrowersideaway from the transmitter latch).

c. Placeonefingeronthetransmittertokeep it in place. With your other hand, pull the transmitter latch up and forward until you hear 2 clicks.

Setting the Alert Profile (continued)Usethe“TryIt”featureundertheProfilesoptionintheMainMenutohearanexampleofeachalertprofile.Followthesestepstosetthealertprofile:a. From any trend graph, press the SELECTbuttontoget

to the Main Menu.b.PresstheDOWNbuttontohighlight“Profiles.”Pressthe

SELECTbutton.c. Press the UP or DOWNbuttontohighlightthealertprofileyouwantto

use.d. Press the SELECTbutton.A willappeartotherightoftheprofileyou

chose.NOTE:Nomatterwhatprofileyouset,allalertsandalarmswillnotifyyoubyvibratingfirst.Therewillbenobeepifyouconfirmthealertoralarmafterthefirstvibration.

Step 4Inserting the SensorBeforeyoubegin,makesureyouhavethetransmitter,receiver,asensorandalcohol wipes. Wash your hands thoroughly, and dry them.a. Chooseasiteonyourbellythatisatleast3inchesfromyourinsulin

pumpinfusionsetorinjectionsiteandoutofthewayofyourwaistband.Avoidareasthatarelikelytobebumpedorpushedorareasofyourskinwith scarring, tattoos or irritation.

b. Removethesensorfromitspackaging.c. Cleanyourskinatthesensorplacementsitewithanalcoholwipe.Letdry.d. Removetheadhesivebackingfromthesensorpodone

half at a time.

e. Placethesensorhorizontally,NOTvertically,onyourskin.Moveyourfingersaroundtheadhesivepatchtosecure the tape to your skin.

Preparing the Device for Patients (continued)

Setting the Receiver in Blinded Mode:Inordertoblindthereceiver,youmustactivatetheClinicalFeaturesthroughtheDexcomSTUDIO™software.SeetheDexcomG4PLATINUMProfessionalSystemUser’sGuideforinstructionstoactivatethisfeature.a.LaunchtheDexcomSTUDIOsoftwarefromacomputerinyouroffice.b.UsingtheUSBcable,connectthereceivertothecomputer.c. Click on the Tools drop down menu. d. Click on the option Blind Attached

Receiver.e. Confirmyourselectionbyclickingon“OK”

beforethereceiverdisplayisupdated.

NOTE: In some cases the receiver may already beblinded.IfyouclickontheToolsmenuoption and the Unblind Attached Receiver is“checked,”thereceiverisalreadyinblindmode.

Placing the Receiver in the Receiver Shield:The receiver shield is necessary to help ensure the receiver is clear of contamination.

a. Wash hands and put on sterile gloves. Peel the acoustic spacerfromthepaperbackingandsecureontheupperspeaker edge.

Getting to Know the CGM Device (continued)The sensor is the piece that is inserted and worn onyourbellyforthewholesession(upto7days).Ifyouhaveaproblemwithasensorsession,keepthe sensor until you speak with Dexcom Technical Support.

Step 2Preparing the Device for PatientsNOTE: For multiple patient use, it is important that each user’s data is erased from the receiver data memory after each use. This will ensure patient privacy with subsequent users. See the Dexcom G4 PLATINUM Professional System User’s Guide for instructions on erasing patient data from the receiver after each use.

Charge Receiver:Afullchargecantakeupto5hoursandwilllastabout3days.Keep the USB port cover open after charging.Tochargethereceiver,plugtheUSBcableintotheACpoweradapter.ThenplugthisadapterintoanACpoweroutlet(suchasawalloutlet),andplugtheotherendofthecableintothereceiver.Thebatterychargingscreenwillappearfirst.Afterafewseconds,thebatterychargingsymbol will show in the upper left corner ofthe3-hourtrendgraph(homescreen).Whenthereceiverisfullycharged,thebatterychargingsymbolwillbecompletelygray.

Confirm Time/Date and Transmitter ID is Correct:a. Press the SELECTbuttontogettotheMainMenu.b.PresstheDOWNbuttontoscrolldowntothe“Settings”menu.c. Press the SELECTbuttontoentertheSettingsmenu.

Preparing the Device for Patients (continued)d. The“Time/Date”optionwillbehighlighted.PresstheSELECTbutton.e. Press the UP or DOWNbuttontoentereachnumberorvalue.f. Press the RIGHT or SELECTbuttontomovetothenextsection.g. Aftersettingthe“AM/PM,”presstheSELECT

buttontoacceptchanges.

h.Scrolldowntothe“TransmitterID”menuoption.i. Press the SELECTbuttontoentertheTransmitterIDmenu.j. FindthetransmitterIDonthebackofthe

transmitter.

k. Press the UP or DOWNbuttontoentertheletterornumberineachspace.

l. Press the RIGHT or SELECTbuttontomovetothenext space.

m.Afterenteringthelastspace,presstheSELECT buttontoacceptchanges.

NOTE: You can only enter the transmitter ID when you are not in an active sensor session.

In blinded mode, setting up the high/low glucose alert settings is only necessary during the Setup Wizard, which only appears during the first time you use the receiver.WARNING:Low/highglucosealertsandthe55mg/dLlowglucosealarmarealldisabledwhenusingtheDexcomG4PLATINUMProfessionalSysteminblindedmodeandwillnotalertyouwhentheglucoseisloworhigh.Checkbloodglucosemeterifthereareanysignsorsymptomsofloworhighglucose.

Your healthcare professional has recommended you use the Dexcom G4 PLATINUM Professional CGM System in blinded mode. This means you will not be able to see your glucose values and trends or receive any glucose alerts or alarms during the sensor session, but the Dexcom G4 PLATINUM Professional System will be recording your glucose values. When you return the CGM system, your healthcare professional will be able to download your sensor glucose readings and review your glucose patterns and trends.Your healthcare professional will train you on the basic use of the Dexcom G4 PLATINUM Professional System in blinded mode. Anytime you have questions, review the Dexcom G4 PLATINUM Professional System User’s Guide, this quick start guide or call Dexcom® Technical Support at 1.877.339.2664 or 1.858.200.0200 24 hours a day, 7 days a week.

Using the Dexcom G4® PLATINUM Professional System in Blinded Mode

Step 1Getting to Know the CGM DeviceTheDexcomG4PLATINUMProfessionalContinuousGlucoseMonitoringSystemismadeupof4components:thereceiver,thetransmitter,thesensorand the receiver shield pack.

Dexcom G4 PLATINUM ReceiverYou will see this screen when you turn on the receiver during a sensor session.

Thereceiveristhesmallhand-helddevicethatlookslikeacellphone.Itwillrecordyoursensorglucosereadingsevery5minutes,documentyourglucosepatternsandtrendsandcanbedownloadedforyouandyourhealthcare professional to review. The receiver will also alert you when troubleshootingmaybeneeded.Press the UP and DOWNbuttonstoscrollthroughthe1-hour,3-hour,6-hour,12-hourand24-hourtrendscreens,highlightmenuitems,orsetvalues.Press the SELECTbuttontoturnthereceiveronorselectthehighlightedoption.Press the LEFTbuttontogobacktothelastitemorscreen.Press the RIGHTbuttontohighlightthenextitem.

Plunger

Applicatorbarrel (containssmall introducer needle and sensor)

Transmitter latch

Safetylock

Collar

Sensorpod

Releasetab

Getting to Know the CGM Device (continued)

Dexcom G4 PLATINUM Transmitter

The transmitter is the grey “chip” that snaps into the sensor pod on your body.ThetransmitterhasauniqueIDthatmakesitpossibleforthetransmitter and receiver to communicate to each other. Keepthetransmitterwithin20feetofthereceiverwithoutobstructionthroughouttheentire7daysensorsessioninordertocaptureyoursensorglucosereadingsevery5minutes.Thetransmitterhasaninternalbattery.Whenthebatteryislow,theTransmitter Low Battery screen appears.Thismeansthereisabout1weekofbatterylife left.Contact your healthcare professional immediately if youseetheTransmitterLowBatteryscreensoanewtransmittercanbeordered.

Dexcom G4 PLATINUM Sensor Applicator

Preparing the Device for Patients (continued)b. NotethelocationoftheUSBopeningonthereceiver

shield.Stretchthereceivershieldoverthereceiver.

c. Peelthetrianglesealfromthepaperbackingandsecurethesealovertheopeningonthebackofthereceiver.

d. Peel1squaresealfromthepaperbackingandsecurethesealovertheUSBdooronthesideofthereceiver.SavetheadditionalUSBsealsforrecharging.

Step 3Setting the Alert ProfileYoucanchoosereceiver“alertprofiles”withdifferentsoundsandvolumestofityourneeds.Inblindedmode,theonlyalertsandalarmsthatwillbeactivatedwhilewearingthesystemarelowbattery,outofrange,troubleshooting/erroralertsandwhenacalibrationupdateisrequired.Thealertprofileoptionsare:

“Vibrate”canbeusedwhenyouwanttosilencethereceiverandbealertedbyvibration.

“Soft”canbeusedwhenyouwantthealerttobediscreet.Thissetsallthealertsandalarmstolowervolumebeeps.

“Normal” is the default. This sets all the alerts and alarms to higher volumebeeps.

“Attentive”canbeusedwhenyouwantthealerttobenoticeable.This sets all the alerts and alarms to loud and with distinctive melodies. “HypoRepeat”issimilartothenormalprofile,butifusedinrealtime,will provide a repetitive alarms when glucose is severely low. This profile is not an appropriate option in blinded mode.

Step d

Step e

Step f

Step h

“2 CLICKS”

Transmitter Low Battery screen

Sensor

Setting Transmitter ID screen

Profiles menu, Try It highlighted

CONTINUOUS GLUCOSE MONITORING SYSTEM

Quick Start Guide

WARNING: Review all indications, contraindications, warnings, precautions and detailed procedures in the Dexcom G4 PLATINUM Professional System User’s Guide prior to using the Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System.

Receiver shield pack

The information on the back of the transmitter may not exactly match

the picture above.

Step a

Step b

Step c

Step d

Receiver USB port cover

USB cable plugged into receiver

Setting Time/Date screen

Blind Attached Receiver

Unblind Attached Receiver

Step i

“ 2 CLICKS”

Step j

Install transmitter in sensor pod

Inserting the Sensor (continued)d. Hold the sides of the sensor pod with one hand.

Removethetransmitterlatchwiththeotherhandbyquicklytwistingoffthelatchawayfromyourbody.

Step 5Starting the Sensor SessionOnceyouhaveinsertedthesensorandattachedthetransmitter,youneedtostart the sensor session on the receiver.a. From any trend graph, press the SELECTbuttontogettotheMainMenu.b. PresstheDOWNbuttontohighlight“StartSensor.”

Press the SELECTbutton.TheStartSensor“thinking”screenwillletyouknowthesensor2-hourstartupperiodhasbegun.

c. Checkthereceiver10minutesafterstartingthesensorsessiontomakesurethereceiverandtransmitterarecommunicating.Theantennasymbol( )shouldbeintheupperleftcornerofthetrendgraph.IftheOutofRangesymbol( ) shows in the upper right corner of the trend graph instead,seetheDexcomG4PLATINUMProfessionalSystemUser’sGuide for more information.

d. A2-hourcountdownsymbolwillshowinthestatusareaofthereceiverscreenandwillfillinduringthe2-hourstartupperiod.

NOTE:YouwillNOTgetanysenorglucosereadingsorglucosealertsfromtheDexcomG4PLATINUMProfessionalSystemwheninblindedmode.

[0-24minutes]

[24-48minutes]

[72-96minutes]

Ready for calibration

[48-72minutes]

Start

Countdown symbol fills in during sensor startup. When you see any of these symbols, you will get NO sensor glucose readings or alerts.

Countdown symbol appears in upper right corner

UPbutton

DOWNbutton

SELECTbutton

LEFTbutton

Battery

Trend graph hours

Current time

RIGHTbutton

SpeakerStatusarea

Antenna

Transmitter InstalledTransmitterLatch

Twist transmitter latch up or down

Side view of transmitter in sensor pod

Start Sensor “thinking” screen

Frequently Asked Questions (continued)When will the receiver need to be recharged?Thereceiverwillshowthelowbatteryalert.Youcanalsochargethereceiverifyounoticethebatterysymbol is low.

How do I re-charge the Receiver during the sensor session?PeelbackandremovethesquaresealoverthereceiverUSBport.PlugtheUSBcableintotheACpoweradapter,andplugtheadapterintoawalloutlet.InsertthemicroUSBcableintothereceiverUSBporttorechargethereceiver.Whenchargingiscomplete,useanewsquaresealfromthereceivershieldpackandplaceitovertheUSBport(seetheDexcomG4PLATINUMProfessionalReceiverShieldPackageInsertformoredetailedinstructions).Ifyouuseall3squaresealsduringthe7daysensorsessionandthereceiverneedstobecharged,returnthereceivertoyourhealthcareprofessionaloffice.

How will I be able to see my glucose patterns and trends if the receiver is in blinded mode?ReceiverscanbedownloadedusingtheDexcomSTUDIOsoftwareinyourhealthcareprofessional’sfacility.Followupwithyourhealthcareprofessionalto review your glucose patterns and trends from the sensor session.

Can the Dexcom G4 PLATINUM Professional CGM System get wet?ThereceiverCANNOTgetwetandisNOTwaterresistant.Showering,bathingandswimmingisOKwhilewearingthesensorandtransmitteronthebody.SeetheDexcomG4PLATINUMProfessionalSystemUser’sGuideformore details.

What can I do if the sensor tape starts to come loose?Medicaltape(suchasBlenderm™, Tegaderm™,Smith&NephewIV3000®, 3M™tape)canbeusedforextrasupport.Ifyouusetape,onlytapeoverthewhiteadhesivepatchonallsidesforevensupport.SeetheDexcomG4PLATINUMProfessionalSystemUser’sGuideformoredetails.

Calibration “Do’s”•Usethesamebloodglucosemeterforallcalibrationsduringasensor

session•Makesureyourbloodglucoseteststripsarecodedcorrectlyandnot

expired•Washanddryyourhandsbeforetestingyourbloodglucose•Useonlyfingerstickbloodglucosevaluesforcalibration

Calibration “Don’ts”•Don’tcalibratewhenyousee or inthestatusbar•Don’tcalibrateifyourbloodglucosevaluefromyourbloodglucosemeterishigherthan400mg/dLorlowerthan40mg/dL

•Don’tcalibratewhenacetaminophen(suchasTylenol®) is active in your body

Calibrating (continued)

Calibration Update:Thecalibrationsingleblooddroppromptwillappearonthereceiverscreentoletyouknowacalibrationupdateisneeded.Toconfirmthisprompt,pressSELECT. Afterthestartupcalibration,calibrationupdatesmustbeenteredonceevery12hoursataminimum.NOTE:Calibratinglessthantherecommendedamountmay result in inaccurate sensor glucose readings. You can enter more calibrationseachdayifyouwant.SeetheDexcomG4PLATINUMProfessionalSystemUser’sGuideformoredetailsoncalibrating.

Advanced Features of the Dexcom G4 PLATINUM Professional CGM SystemTheDexcomG4PLATINUMProfessionalSystemhasadvancedfeaturesand alerts that may provide more information to assess your glucose patterns and trends. These advanced features that may provide additional information inblindedmodeare:

Step 6CalibratingYouneedtouseyourbloodglucosemetertocalibratetheDexcomG4PLATINUMProfessionalSystem.Attheendofthe2-hourstartupperiod,youmustentertwocalibrationsbeforeanysensorglucosereadingswillberecorded.Calibrationbloodglucosevaluesneedtobeenteredintothereceiverwithin5minutesoftestingwithyourbloodglucosemeter.

Startup Calibration:a. Afterthe2-hourstartupperiod,thestartupcalibration

doubleblooddroppromptwillshowonthereceiverscreen. Press the SELECT buttontoclearthisalert.

b.Washanddryyourhands.c. Takeafingerstickbloodglucosemeasurementusingyourbloodglucose

meter.d. From any trend graph, press the SELECT buttontogettotheMainMenu.e. Press the DOWNbuttontohighlightEnterBG.PresstheSELECT button.f. Forthestartupcalibration,thedefaultbloodglucosevalueonthereceiver

is120mg/dL.PresstheUP or DOWNbuttontoentertheexactbloodglucosevaluefromyourbloodglucosemeter.PresstheSELECTbutton.

g. Press the SELECTbuttonagaintoacceptthecalibration.Thecalibrationscreenwillthenshow.

h. Repeatstepsc-gtoenterasecondbloodglucosevalue.Afterthesecondbloodglucosevaluehasbeenentered,the“DisplayOff”screenshouldbevisible.

Step 7Ending the Sensor SessionThesensorautomaticallyshutsoffafter7days.Thereceiverwillalertyouat6hours,2hoursand30minutesbeforethesensorsessionends.TheReplaceSensorNow screen will show the sensor session has ended. Thesensorpodandtransmittercanberemovedbythepatient or healthcare professional.

Removing the Sensor Pod and Transmitter:NOTE: Keep the sensor and transmitter together when removing from your skin.a. PeelthesensorpodoffyourbodylikeaBand-Aid®. Do not remove the

transmitterfromthesensorpoduntiltheentiresensorhasbeenremovedfromyourbody.

b. Placethesensorpod/transmitteronaflatsurface.Therearetwowaysyoucanremovethetransmitter:

Sensor TroubleshootingThesearethemostcommonlyseenerrormessagesfortheDexcomG4PLATINUMProfessionalSystem.Followtheinstructionstotroubleshoottheerror message. If other error messages come up on the receiver, refer to theDexcomG4PLATINUMProfessionalSystemUser’sGuideorcontactDexcomTechnicalSupportat1.877.339.2664 or 1.858.200.0200 for any unresolvedquestions/troubleshootingissues.Iftheerrormessagecannotberesolved, contact your healthcare professional to discuss next steps.

Dexcom,DexcomG4,DexcomG4PLATINUM,DexcomG4PLATINUMProfessional,DexcomSTUDIO,andOneStepAheadareeitherregisteredtrademarksortrademarksofDexcom,Inc.intheUnitedStatesand/orothercountries.Allotherproductorcompanynamesthatmaybementionedinthispublicationaretradenames,trademarksorregisteredtrademarksoftheirrespective owners.

Calibration prompt

Replace Sensor Now screen

©2014Dexcom,Inc.Allrightsreserved. LBL-012214Rev05MT21694

Dexcom, Inc. I 6340 Sequence Drive I San Diego, CA 92121 Technical Support: 1.877.339.2664 / 1.858.200.0200 I www.dexcom.com

Enter BG “thinking” screen

Display Off (CGM in calibration and

recording readings)

Ending the Sensor Session (continued) •Useyourfingerstospreadthebacktabsofthesensorpod.The

transmitter will pop out OR •Usethesafetylock(fromthesensorapplicator).

Insert the jagged edges of the safety lock so they “hug” the wide end of the transmitter in the sensor pod. Press the safety lock down until you cannot press anymore, and then pull up. The transmitter will pop out. DO NOT discard the transmitter

c. Returnthefollowingcomponentstoyourhealthcareprofessional: •Receiverwithreceivershieldstillintact.DONOTremovethereceiver

shieldbeforereturningittoyourhealthcareprofessional •Transmitter •Sensorpod(ifrequestedbyhealthcareprofessionaland/ortherewere

technical issues with the sensor) •ACpoweradapter •ChargingUSBcableNOTE:Forproperremovalofthereceivershieldbythehealthcareprofessional,seetheDexcomG4PLATINUMProfessionalSystemUser’sGuide.

Blood Drop in Status Area

Calibration needed.• Enter a calibration as described in Step 6 of this quick

start guide

Out of Range

The receiver and transmitter are not communicating.• Make sure the receiver is within 20 feet of the transmitter• If the out of range icon continues to display, contact

Dexcom Technical Support

Wait 15 Minutes Calibration Error

Sensor is not calibrating correctly.• Press the SELECT button. will show in status bar• Wait 15 minutes, then enter 1 calibration• If error continues, enter 1 more calibration, and wait 15

minutes• If no sensor glucose readings show, contact Dexcom

Technical Support

Wait 1 Hour Calibration Error

Sensor is not calibrating correctly. • Press the SELECT button. will show in status bar• Wait about 1 hour, then enter 1 calibration• Wait 15 minutes. If no sensor glucose readings show,

contact Dexcom Technical Support

Glucose Reading Error in Status Area

Wait. usually clears up by itself within minutes or hours. If this continues after 3 hours, contact Dexcom Technical Support.• Do not calibrate • Make sure the sensor pod is sticking well to your body

and nothing is rubbing against it• Check that the transmitter is snapped in on both sides

Red stoplight in upper right corner

Main Menu, Events highlighted

Events menu, Carbs highlighted

Transmitter with safety lock snapped in

Frequently Asked QuestionsWhy can’t I take any acetaminophen containing medication while I am using the Dexcom G4 PLATINUM Professional System?AcetaminophenisacontraindicationoftheDexcomG4PLATINUMProfessionalSystemandshouldnotbetakenwhilethesensorisinserted.TheDexcomG4PLATINUMProfessionalSystemmayrecordinaccuratelyhighwhenyoutakeacetaminophen-containingdrugs.Thelevelanddurationofinaccuracymaybedifferentforeachperson.Tounderstandhowlongacetaminophen-containingmedicinesmayaffectyourbody,consultyourhealthcare professional.

Startup calibration prompt

Advanced Features of the Dexcom G4 PLATINUM Professional CGM System (continued) •EventMarkers(insulin,carbohydrates,exerciseandhealthrelatedissues)

•OutofRangeIconandAlert

Based on the recommendations of the healthcare professional, these featurescanbeturnedofforonfortheindividualpatient’sneedsandgoals.SeetheDexcomG4PLATINUMProfessionalSystemUser’sGuideforinstructions to activate and use these features.

Low battery alert

Main Menu, Alerts highlighted

Alerts menu, Advanced highlighted

Advanced menu, Out of Range highlighted

Indications for Use:TheDexcomG4PLATINUMProfessionalContinuousGlucoseMonitoringSystemisaglucosemonitoringdeviceindicatedfordetectingtrendsandtracking patterns in persons (age 18 and older)withdiabetes.Thesystemisintendedformultiplepatientuseandrequirespurchasebyahealthcareprovider. The system is not intended for use in a hospital setting.TheDexcomG4PLATINUMProfessionalSystemisindicatedforuseasanadjunctivedevicetocomplement,notreplace,informationobtainedfromstandard home glucose monitoring devices.TheDexcomG4PLATINUMProfessionalSystemaidsinthedetectionofepisodesofhyperglycemiaandhypoglycemia,facilitatingbothacuteandlong-termtherapyadjustments,whichmayminimizetheseexcursions.InterpretationoftheDexcomG4PLATINUMProfessionalSystemresultsshouldbebasedonthetrendsandpatternsseenwithseveralsequentialreadings over time.

Contraindications:•RemovetheDexcomG4PLATINUMSensor,Transmitter,and ReceiverbeforeMagneticResonanceImaging(MRI),Computed Tomography(CT)scan,ordiathermytreatment.ThedeviceisMR

Unsafe.DonotbringanyportionofthedeviceintotheMRenvironment.TheDexcomG4PLATINUMProfessionalSystemhasnotbeentestedduringMRIorCTscansorwithdiathermytreatment.Themagneticfieldsand heat could damage the device so that it might not display sensor glucosereadingsorprovidealerts,andyoumightmissaloworhighbloodglucose value.•Takingmedicationswithacetaminophen(suchasTylenol)whilewearing

the sensor may falsely raise your sensor glucose readings. The level of inaccuracydependsontheamountofacetaminophenactiveinyourbodyandmaybedifferentforeachperson.

Key Warnings - Risk of Blood Borne Pathogen Transmission:•UsersneedtoadheretoStandardPrecautionswhenhandlingorusingthisdevice.Allpartsofthesystemshouldbeconsideredpotentiallyinfectiousandarecapableoftransmittingblood-bornepathogensbetweenpatientsandhealthcareprofessionals.Formoreinformation,referto“GuidelineforIsolationPrecautions:PreventingTransmissionofInfectiousAgentsinHealthcareSettings2007,”http://www.cdc.gov/hicpac/2007ip/2007isolationprecautions.html.

Key Warnings - Risk of Blood Borne Pathogen Transmission (continued)•Thetransmittershouldbecleanedanddisinfectedafteruseoneachpatient.ThissystemmayonlybeusedfortestingmultiplepatientswhenStandardPrecautionsandthemanufacturer’scleaninganddisinfectionprocedures are followed.

•Replacethereceivershieldandacousticspacerwhentransferringthereceiver to a new patient. Wear gloves when handling the receiver. Handling the receiver without a shield or with a damaged shield could lead to contamination, which could cause infection.

Key Precautions - Risk of Blood Borne Pathogen Transmission:•Ifall3squaresealshavebeenusedbeforetheendofthesession,returnthereceivertothehealthcareprofessionaloffice.Donotchargethereceiverif you cannot seal it. Unsealed receivers could spread contamination.

•Donotpuncturethereceivershieldforanyreason.Puncturedshieldscouldspread contamination.

•Inspectthereceivershield,sealsandspacerfortearsorotherdamageordefectsbeforeyouplacethecomponentsonthereceiver.Donotuseany parts of the receiver shield pack that are damaged or defective. If the receiver shield, seals and spacer are torn, damaged or otherwise defective after giving to the patient, the patient should return the receiver to the healthcare professional for a replacement. The damaged or defectivereceivershieldshouldnotbetakenoffbythepatient.Ifadefector malfunction is suspected, the healthcare professional should retain the shieldandcontactDexcomTechnicalSupport.Theshieldhelpsprotectthe user and healthcare professionals from contamination and allows the receiver to function properly.

For complete details on the cleaning and disinfection procedures of the DexcomG4PLATINUMProfessionalCGMSystem,refertotheDexcomG4PLATINUMProfessionalSystemUser’sGuide.

Safetylock

Sensor Troubleshooting (continued)

Sensor Failed

Sensor shuts off before end of 7-day session.• Press the SELECT button to clear the alert/screen • Contact Dexcom Technical Support• Remove failed sensor and insert a new sensor

MR