PMT Readout and Floor Triggering Charge estimation using the times over the thresholds Event...

Post on 18-Jan-2016

214 views 0 download

Transcript of PMT Readout and Floor Triggering Charge estimation using the times over the thresholds Event...

PMT Readout and Floor TriggeringPMT Readout and Floor Triggering• Charge estimation using the times over the thresholds• Event Building and Triggering

+

multiplicity

George BourlisGeorge Bourlis

ReadOut Electronics

GPS Input

USB PortTrigger Output5 PMT Signal Inputs

25ps accuracy TDC

HPTDC

• 32 channels (LR) – 8 Channels (HR)

•25ps (HR) to 100 ps (LR) accuracy

•Self Calibrating

ReadOut Electronics

EZ-USB FX2 HPTDC CONTROL

Card Debugging Configuration

Trigger

Input

splitter

splitter

discriminator

- Offset of each input

- Scale of each input

- Timing among input channels

Sample PulsesInput 1

Sample PulsesInput 1

Sample PulsesInput 1

The Photomultiplier Tube:

PH: XP1912

Charge (pCb)

Single p.e

Gain vs HV

Calibration

@ “nominal” H.V.

gain: ~ 4 105

<charge>/p.e. ~ 0.07pCb

<pulse height>/p.e. ~ 1.05mV

Rise Time: 1.2 ns

Charge (in units of mean p.e. charge)

At the Detector Center

Data

- Monte Carlo Prediction

Charge versus Time Over ThresholdC

ha

rge

(p

C)

Time Over Threshold (s)

1st Threshold only 1st & 2nd Threshold 1st, 2nd & 3rd Threshold

4mV

50mV

15mV

Charge Parameterization

1st & 2nd Threshold 1st, 2nd & 3rd Threshold

kkCharge= a (tot)

Charge Estimation

1st & 2nd Threshold 1st, 2nd & 3rd Threshold

Measured Charge - Estimated Charge

Measured Charge

11% 8%

- Estimated Resolution ~10%

- Better if all thresholds are crossed

Charge Estimation

1st & 2nd Threshold 1st, 2nd & 3rd Threshold

Measured Charge - Estimated Charge

Estimated Error

- σ=(1.01 ± 0.01) (2 thresholds)

- σ=(1.1 ± 0.1) (3 thresholds)

Time (ns)

Trigger

Input