Medical Physics secondaty level slides

Post on 16-Jan-2016

10 views 0 download

Tags:

description

Ultrasound, Xray imaging notes

Transcript of Medical Physics secondaty level slides

Incident Wave I0

transmitted Wave I2 transmitted

Wave I3

Reflected waves

Reflected wave R1 Reflected wave R2

Antenuation causedBy scattering, diffraction,Absorption/dissipated

I23

I12= I1 x e- µ1x1 I23= I2 x e- µ 2x2

Thickness x1,Acoustic Impedance Z1

Attenuation Coff µ1

Thickness x2,Acoustic Impedance Z2

Attenuation Coff µ2

R0

Attenuated R1Attenuated R2

t

Intensity

T1=2 X1/(V1)

T1=2 X1/(V1) + 2 X2/(V2)

Reflection of R1 at medium 0 and 1 boundary

Reflection of R2 at medium 0 and 1 boundary

I12

R0

R0 = α1 = (Z1-Z0)2

I0 (Z1+Z0)2 (simple ratio, dimension, remember the square)

0<= α<=1; If Z1=1/2 Z0=> α=1/9

I1 = (1-α1 ) I0= 1- (Z1-Z0)2

(Z1+Z0)2

= I0-R0

I0

I2 = (1-α2 ) I12= 1- (Z2-Z1)2

(Z2+Z1)2

= I12-R1

= I1 x e- µ1x1 1- (Z2-Z1)2

(Z2+Z1)2

I12

Ultrasound –Ultrasound –Reflection/echo;Reflection/echo; attenuation attenuationAcoustic impedanceAcoustic impedance

Key Physical theory:Reflection at bounduaryAttenuation and Acoustic impedance

f↑ resolution improves; Power ↑ but µ ↑ penetration depth ↓

Impedance ≠ attentuation Z=ρC (unit Rayl, kgm-2s-1)

(attenutaion lead to energy loss (as heat etc)Difference in impedance leads to reflection of sound energy)

µ x X is pure number

X in cm=> µ in 1/cmI (intensity) in Watt/m2

transmitted Wave I1

Key Word: TransducerPiezoelectric effect

A-Scan – ultrasound transducer + signal generator+ oscilloscopePrinciple – pulse echo (due to reflection) distance measurement

Show objects of different range and depthsElectirc pulse generates a ultrasound pulse by transducer and measure the time lag of echo (echo deforms transducer and generate electric signal)

B- ScanRepresent reflected intensity (spikes) AS brightness.Spots of different brightness proportional to Intensity of echo. By sweeping the transducer through an arc or use an array of transduceer a series of strips of scan over an area forms 2 D image

Acoustic shadow

Highly reflecting Gall stone

f↑ resolution improves; Power ↑ but µ ↑ penetration depth ↓

Incident X Ray I0

I1= I0 x e- µ1x1

Thickness x1, Attenuation Coff µ1

X-Ray; CT Scan Anttenuation causedBy scattering, diffraction,Absorption; for X Ray by ionising the tissue the X-Ray photon is absorbed-attenuated

Bone/

Air/low anttenuationorgan

Arteries/Intestine with artificial Contrast media

I0

ab

I1≈ I0 x e- µ1(a+b)

X-Ray partiallystopped by Bone/tissue with high attenuation

Metal MetalX-Ray stopped by Contrast media/Metal

E1Highest exposure

E2- Highexposure

E3 Lowexposure

E4 Mediumexposure

Exposure E1>E2>E4>E3

Transmitted beam vs reflected beam in US

RadionuclidesCharacteristics of Radionuclides used:-Non-toxic-produce Gamma ray only, no alpha and beta (range, damage)-Physical half life a few hours: long enough to allow imaging, not to long to cause lasting problem to body and risk of disposal-Decay to stable nuclide- Availability (technicium cow) and cost Technetium 99m decay into stable technetium (ie, not radiating), 6 hour half life), excreted in urine emits gamma only.

Problem of radiation hazard after excreted by body-still raidating for a few years in environment

Technetium -99m

Safety precaution:Limit the number of examinationKeep distance from radio active source except during exmainationShielding of source(Cow made of lead container)Containment- room with –ve pressure, proper disposal,