HEMODYNAMIC MONITORING:The Fundamentals

2

Goal:

Provide the participant with the basic knowledge required to care for a patient with an arterial line, central venous pressure/right atrial pressure line, and/or a pulmonary artery catheter.

3

Objectives: Define hemodynamic monitoring. State indications for hemodynamic monitoring. Identify the correct reference point for leveling &

zeroing the hemodynamic monitoring system. Identify components of normal arterial, right atrial,

pulmonary artery, & pulmonary artery wedge waveforms. Identify factors that can affect the accuracy of the

waveforms. Describe common complications associated with

hemodynamic monitoring. Discuss the nursing care for patients with hemodynamic

monitoring.

4

Hemodynamic Monitoring:

The use of a device or instrument to provide physiological measurements in order to more closely and accurately monitor a patient’s condition.

Usually involves the use of an invasive catheter inserted into a body cavity or organ system.

5

Hemodynamic Monitoring - Examples:

Intra-arterial pressureCentral venous pressureIntra-cardiac pressures:

Right atrial pressurePulmonary artery pressurePulmonary artery wedge pressure

6

Indications:

Intra-arterial:Continuous blood

pressure monitoringBlood sampling

Frequent arterial or venous sampling

Ability to quickly assess the effects of medical interventionsMedications or fluid

Intra-cardiac:Continuous monitoring of

both right & left heart fluid status

Cardiac outputMixed venous blood

samplingAbility to quickly assess

the effects of medical managementMedications or fluid

7

Common Components:

Monitoring device:AmplifierOscilloscope

Transducer cableTransducerTransducer holder

Flush system:Pressure bagFluid:

Heparin vs. NS

Pressure tubingCatheterCarpenter’s level

Common System Components:

9

Accurate & Reliable Waveforms/Values:

Technical factors:Patient positioning:

Supine Head of bed: 0-45°

Leveling: Eliminates effects of hydrostatic forces on the observed

hemodynamic pressures Ensure air-fluid interface of the transducer is leveled

before zeroing and/or obtaining pressure readings Phlebostatic axis:

Level of left atrium4th ICS & MAL (technically ½ AP diameter)Mark the chest with washable felt pen

Phlebostatic Axis:

Leveling:

12

Accurate & Reliable Waveforms/Values:

Technical factors:Zeroing:

Negates the force exerted by the atmosphere (760 mmHg at sea level)

Pressure transducers can be affected by changes in temperature

“Drift” will occur from the zero baseline over time

13

Accurate & Reliable Waveforms/Values:

Technical factors:Pressure tubing & transducer system

System free of airLength of tubing; correct tubingSystem connectors tight; stopcocks Luer-lockedAdequate fluid in flush systemFlush solution pressurized to 300 mmHgDynamic response testing

Square wave, fast flush or snap test

Dynamic Response Test:

Figure A – Expected square wave test

Figure B – Overdamped

Figure C – Underdamped

15

Accurate & Reliable Waveforms/Values:

Overdamped:Sluggish, artificially rounded & blunted

appearanceSBP erroneously low; DBP erroneously highCauses: large air bubbles in system, too

compliant of tubing, loose/open connectionsOverdamped? Shock states, vasodilation,

aortic stenosis, thrombus on catheter tip, catheter kinked or against vessel wall

Overdamped Waveform:

17

Accurate & Reliable Waveforms/Values:

Underdamped or ringing:Overresponsive, exaggerated, artificially

spiked waveformSBP erroneously high; DBP erroneously low Causes: small air bubbles, too long of

tubing, defective transducer Underdamped? Vasoconstriction,

hypertension, atherosclerosis, aortic regurgitation, hyperdynamic states (fever)

Underdamped Waveform:

Intra-arterial Pressure Monitoring

20

Blood Pressure Monitoring:

Non-invasive blood pressure:Dependent on blood flow

Invasive, intra-arterial monitoring:Dependent on pressure changesTypically more accurate than NIBPAllows for convenient arterial blood

sampling

21

Arterial Line Insertion Sites:

Radial:Most common siteGood collateral circulationAccessibility & ease of maintenanceModified Allen’s test

BrachialFemoralDorsalis Pedis

22

Arterial Waveform:

Peak systolic pressure:Generated by left ventricular contractionPeak systole correlates with QRS on ECG

rhythm stripNormal: 100-140 mmHg

Arterial Waveform:

24

Arterial Waveform:

Dicrotic notch:Closure of the aortic valveMarks the end of ventricular systole & the

beginning of diastoleDiastole:

Lowest pressure in arterial systemMeasured just before systolic upstrokeNormal: 60-80 mmHg

Arterial Waveform:

26

Arterial Waveform:

Catheter location:The more distal the catheter is placed, in

relation to the aorta, the higher the systolic pressure & the lower the diastolic pressure

28

Complications:

Infection:LocalizedSystemic

HematomaHemorrhageThrombosis/embolizationIschemiaNerve damage

29

Nursing Care: Assess arterial line flush system:

Adequate, appropriate fluid Pressurized to 300 mmHg No air bubbles Connections tight Infection control

Level & zero q shift & PRN Continuously observe arterial waveform quality;

over/underdampened Correlation to NIBP Site care Assess circulation distal to insertion site

Central Venous Pressure&

Right Atrial Pressure

31

Central Venous Pressure & Right Atrial Pressure:

Indications:Assessment of intra-vascular volume status:

Preload: the volume or pressure generated at end-diastole

Assess right ventricular function:Follow trends

Secure access; often only IV access available

32

Central Venous Pressure & Right Atrial Pressure:

Measured through a catheter tip placed within the right atrium (RAP) or just outside RA in the vena cava (CVP):Single-lumen, double-lumen, triple-lumen

cathetersPulmonary artery catheter

Access Sites:SubclavianInternal jugular veinFemoral vein

33

CVP/RAP Pressures:

Normal values:2–6 mmHg

Increased CVP/RAP:HypervolemiaHyperdynamic states (increased cardiac output)Cardiac tamponadeConstrictive pericarditisPulmonary hypertensionHeart failurePulmonary embolusPositive pressure ventilation

34

CVP/RAP Pressures:

Normal values:2–6 mmHg

Decreased CVP/RAP:Hypovolemia:

DehydrationHemorrhage

Decreased mean systolic pressureVasodilation (specifically, venodilation):

SepsisVasodilators

35

CVP/RAP Waveform:

Waveform Components: “a” wave:

Atrial contraction (atrial kick)Within the PR interval

“x” descent: atrial diastole “c” wave:

Rise in pressure w/ tricuspid valve closureNear the end of the QRS

“v” wave:Passive atrial filling during ventricular systoleFollowing the T wave, within the T-P interval

“y” descent: passive atrial emptying

CVP/RAP Waveform:

37

Abnormal Waveforms:

Large “a” waves:Loss of A-V synchrony (Cannon Waves)Tricuspid valve stenosis

Loss of “a” waves:Dysrhythmias resulting in loss of P wave

Giant “v” waves:Tricuspid insufficiencyRight ventricular failure

38

Complications:PneumothoraxInfection:

Localized, endocarditis, systemicBleeding/hemorrhageDysrhythmias:

Premature ventricular contractionsVentricular fibrillationHeart block

Thrombus/embolusPerforation of cardiac chamber

39

Nursing Care: Assess CVP/RAP line flush system:

Adequate, appropriate fluid Pressurized to 300 mmHg No air bubbles Connections tight Infection control

Level & zero q shift & PRN Continuously observe CVP/RAP waveform quality;

over/underdampened Correlation to clinical picture Site care Assess for complications

Pulmonary Artery Pressure&

Pulmonary Artery Wedge Pressure

41

Indications: Assess left & right heart function:

Preload:Right heart: RAPLeft heart: pulmonary artery diastolic & pulmonary artery wedge

pressures Cardiac output: thermodilution Afterload:

Right heart: Pulmonary vascular resistance (PVR)Left heart: Systemic vascular resistance (SVR)

Contractility: Stroke work index (SWI)

Assess response to therapeutic interventions Atrial and ventricular pacing Mixed venous blood gas Continuous venous oxygen saturation

42

Clinical Indications:Complicated MIEnd-stage heart failureAcute pulmonary edemaPulmonary embolusAcute respiratory distress syndromeShockAcute renal failureComplex fluid managementCardiac surgeryHigh-risk surgical patients

43

PA Pressure Monitoring:

Pulmonary artery catheter:Standard is a #7-French, multi-lumen,

radiopaque catheterMarked in 10-cm incrementsMultiple ports & openings1.5-cc balloon at distal end

Pulmonary Artery Catheter:

45

Anatomy of a PA Catheter:

Components:ThermistorBalloon inflation port w/ gate valveProximal injectate portTypically has three infusion ports:

RA infusion port & lumen openingRV infusion port & lumen openingPA distal infusion port & lumen opening

Balloon

Insertion & Placement:

48

Insertion:

Access:Right internal jugularLeft Subclavian veinFemoral

Method:Patient positioningPercutaneousVessel dilator & introducer sheath

RA

RV

PA

PAWP

Insertion:

RA

RV

PA

PAWP

55

PA Pressures:

Pulmonary artery systolic (PAS):Normal PAS: 15–30 mmHg

Pulmonary artery diastolic (PAD):Normal PAD: 5–16 mmHg

Pulmonary arterial wedge pressure:Normal PAWP: 5–12 mmHg

All pressures measured at end-expiration

56

PA Pressures:

Increased PA pressures:Pulmonary hypertensionPulmonary diseaseMitral valve diseaseLeft ventricular failurePulmonary embolus

Decreased PA pressures:HypovolemiaPulmonary artery vasodilation (meds, SIRS, sepsis)

PA Waveform:

PA Waveform Components:

1. PA systolic peak

2. Dicrotic notch:Closure of PA valve

3. PA diastole

4. Anacrotic notch:Opening of PA valve

59

Pulmonary Artery Wedge Pressure:

Also: PCWP or PAOPNormal PAWP: 5–12 mmHgIndirect measurement of mean left atrial pressure

(LAP) which is an indirect measure of left ventricular end-diastolic pressure (LVEDP) or left ventricular preload

Normally, PAWP is 1–3 mmHg lower than PADPhysiologically impossible for PAWP to be higher

than the PAD

PA

PAWP

PAWP:

PAWP Balloon Inflation:

PAWP Waveform Components:

“a” wave:Atrial contraction (atrial kick)

“x” descent:Atrial diastole

“c” wave:Closure of mitral valve

“v” wave:Atrial filling

“y” descent:Passive atrial emptying

PAWP Waveform Components:

67

Obtaining PAWP:

Continuous, close observation of the ECG and the PAP waveform during procedure

Inflate balloon: Open gate valve after syringe attached Slowly & gently inflate with only enough air to “wedge” the

balloon (typically 1.25-1.5 cc) Wedge for no more than 10-15 seconds or 2-3 respirations If strong resistance is met – do not inflate If no resistance is met or blood in lumen - STOP

Measure PAWP at end-expiration “a” wave method – most accurate

Obtaining PCWP:

Respiratory Variation:

PAPwaveform in a spontaneously

breathing patient

Respiratory Variation:PAWP waveform in a mechanically

ventilated patient

71

Monitoring Problems:

Air bubbles in systemBlood in systemThrombus at tip of catheterSpontaneous wedging & overwedgingLoss of pressure tracingPressures that do not match the clinical

picture

Overwedging:

73

Complications:

InfectionDysrhythmiasPneumothoraxHematoma & hemorrhageValve rupturePulmonary artery rupturePulmonary infarctPulmonary embolus

74

Nursing Care:

Goals of therapyPrevention of infectionPatient positioningAssuring accuracy of dataContinuous waveform monitoringActivity

75

Cardiac Output:

Cardiac output:Amount of blood pumped by the heart per

minuteCO = SV x HRFour physiologic factors affect CO:

PreloadAfterloadContractilityHeart rate

76

Cardiac Output:

Preload:Load (volume) that stretches the ventricles prior to

contraction:Right heart – CVP/RALeft heart – PAD or PAWP

Afterload:Impedance to the ejection of blood from ventricles:

Depends on: volume & mass of blood ejected; and compliance & size of vascular space into which blood is being ejected

Right heart – PVRLeft heart - SVR

77

Cardiac Output:

Contractility:Ability of myocardial tissues to shorten and develop

tension; “squeeze”:Cannot directly measure clinicallyIndirectly: LVSWI or RVSWI

Heart rate

Stroke volume: average volume of blood ejected per cardiac contraction

78

Cardiac Output:

Thermodilution:Equipment:

Computer module/monitor w/ softwarePA catheter w/ thermistor & injection set-up

5-10 cc fluid bolus (injectate)Room temperature vs. iced injectateInject smoothly & quickly w/ end-expirationCalibration factors: computation constantPatient positioning

79

Cardiac Output:

Normal values:Cardiac Output: 4–6 L/minCardiac Index: 2.5–4 L/min