Post on 19-Jan-2015
description
Rapid Particle Characterization of Therapeutic Protein Formulations Using
Continuous Digital Imaging
Lew Brown
Fluid Imaging Technologies, Inc.Yarmouth, ME USAwww.fluidimaging.com
Outline:
• Historical Usage of Microscope Images
• Limitations
• New Technologies for Analysis of Particulates (Benefits/Limitations)
• Fluid Imaging FlowCAM®: Architecture and Features
• Demonstration
Limitations of Traditional Microscopy
• Static Sample– Only a small amount of sample can be processed
at a time– Small sample = decreased reliability/repeatability
• Time to Prepare and Analyze Sample– Time required to prepare sample onto microscope
slide– Each particle counted and measured one at a time
• Human Error– “Tired eyes”– Interruptions
New Technologies for Analysis of Particulates (Benefits/Limitations)
• “Particle Analyzers”– Operate on “Particle Volume” Principle– Benefits:
• Non-static sampling• Very rapid counting
– Drawbacks:• “Particle Volume” only• No shape measurements• Limited information (“Relative Measurements”)• No picture!
New Technologies for Analysis of Particulates (Benefits/Limitations)
• Optical Automated Micrography– Originally photographic, now (mostly) digital– Benefits:
• “A picture is worth a thousand words!”• Automates measurements• Allows subjective review by scientist
– Drawbacks:• Static sample• Still time consuming
The Big Question:
FlowCAM Architecture
Continuous Imaging Particle Analysis: How it Works
Continuous Imaging Particle Analyzer: Features
• Continuous imaging• Fluorescence/scatter triggered imaging• Can collect upwards of 25,000 images/min• Intuitive software (including pattern
recognition) allows users to quickly navigate tens of thousands of images
–Qualify, isolate and quantify sub-populations based on any combination of the 23 measured parameters
–Remove unwanted particles from data sets
Analysis of Two Drug Formulations
• Goal–Provide a concentration of particles >10um
and >25um w/o counting the Si oil droplets that are common to these samples
–Provide images of these particles• Sample analysis
–10X Objective (~100X Overall Magnification), and a 90um deep flow cell
–Post-processed to remove Si oil droplets and air bubbles
Typical Results
Removing Unwanted Particles
• Display all particles >10um
• Use sorting or pattern matching to isolate unwanted images
• Delete unwanted images
• Display summary statistics for particles >10um and >25um
Effects of Processing• Before Droplet Removal
• After Droplet Removal
Summary of Results Sample A:
Summary of Results Sample B:
Formulations that contain high concentrations of particles larger than 10um have a higher chance of causing complications in patients than those that do not
Sample B has a many more outsized particles (492 p/ml) than sample A (95 p/ml)
Sample B is more apt to cause complications than than A
Conclusions
Flow Microscopy offers the Speed of Rapid P.A., With the Differentiation of
Microscopy
Speed enables the collection of statistically significant data populations
Instant visual feedback from particle images; great for outlier quantification and qualification
More data/particle provides an enhanced ability to identify and quantify different particle types
Flow Microscopy Offers the Speed of Rapid P.A., With the Differentiation of
Microscopy
All particle images saved; can be re-analyzed at any time, and can serve as concrete documentation
Morphological pattern matching allows for rapid data processing
Image libraries can be compiled and used as references for pattern matching (normalize out human error)
Thank You!
www.fluidimaging.com