Analysis of Cations in Hydraulic Fracturing Flowback Water from the Marcellus Shale Using Ion Chromatography
2
Environmental Impact of Hydraulic Fracturing
• Challenges
• Potential for contamination of water, soil, and air
• Compliance to clean water act and shale gas regulations
• Optimization of fracturing processes
• Solution
• Water quality monitoring
• Anion and cations
• Metals
• Organics
• Radiation
3
Hydraulic Fracturing Workflow
Well 1
Desalination
Frack Chemicals Pre-Injection Site
assessment
Recycle
Flowback / Produced
Brin
es
Waste
Disposal
Deep Well
Injection
Gas Production
Monitoring Well Monitoring Well
Frack Chemicals
Well 2
Water
Fresh Water
4
Hydraulic Fracturing Workflow Monitoring
Inorganic
Organic
Metals
Anions
Surfactants
Cl-, Br-, SO4-
IC, Discrete Analyzer
Ethoxylated phenols, acrylamide
LC-MS/MS, LC-CAD
Sr, Ba, Ca, Mn, Ar, etc.
IC, AAS, ICP-OES, ICP-MS, HR-ICP-MSCations
Analy
tes
Radiation
Water Chemistry
Sediments HF Water
Composition
Frack Design
Flowback and
Wastewater
Produced WaterSite Monitoring
Natural Gas Methane, BTEX
GC
Gross Alpha, Beta, Gamma, Radium 226, 228
GM, NaI
Isotopes ratios
Organic acids
IC
Brines
TDS, alkalinity, pH, conductivity, DO
multiple
13C-CH4 , 18O 87Sr/86Sr
stable gas IRMS HR-ICP-MS, TIMS, MC-ICP-MS
Instrumentation
5
Analytes in Flowback Wastewater Measured by IC
• Inorganic anions
• Chloride
• Impacts effectiveness of additives (reuse)
• Disrupts nitrification processes
• Bromide
• Ozonation, chlorination -> disinfection by-products: brominated
trihalomethanes, bromate
Carcinogenic
• Sulfate
• Can disrupt anaerobic digestion processes
• Organic acids
• Formic and acetic acids
• pH balance is important for efficient fracking
6
Superior Resolution Using 4 µm Particle Size Capillary Columns
Dionex IonPac AS11-HC-4µm (0.4 × 250 mm Column)
Column: Thermo Scientific™ Dionex™
IonPac™ AG11-HC-4µm/
AS11-HC-4µm (0.4 × 250 mm)
Eluent Source: Thermo Scientific Dionex
EGC-KOH Cartridge (Capillary)
Eluent: 1 mM KOH for 0.01 min,
1–5 mM KOH in 15 min;
5–55 mM KOH in 25 min
Flow Rate: 15 µL/min
Inj. Volume: 0.4 µL
Temperature: 30 °C
Detection: Suppressed Conductivity,
Thermo Scientific™ Dionex™
ACES™ 300 Anion Capillary
Electrolytic Suppressor, recycle
mode
400 10 20 30-2
20
µS
Minutes
1
2
34
56
7
8
9
10
11 12
1314
15
1617
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
3637
38
39
40
1. Quinate
2. Fluoride
3. Lactate
4. Acetate
5. 2-Hydroxybutyrate
6. Propionate
7. Formate
8. Butyrate
9. Methylsulfonate
10. Pyruvate
11. Isovalerate
12. Valerate
13. Monochloroacetate
14. Bromate
15. Chloride
16. 2-Oxovalerate
17. Nitrite
18. Ethylphosphate
19. Trifluoroacetate
20. Bromide
21. Nitrate
22. Citramalate
23. Malate
24. Carbonate
25. Malonate
26. Citraconitate
27. Maleate
28. Sulfate
29. ɑ-Ketoglutarate
30. Oxalate
31. Fumarate
32. Tungstate
33. Phosphate
34. Phthalate
35. Arsenate
36. Citrate
37. Chromate
38. Isocitrate
39. cis-Aconitate
40. trans-Aconitate
7
Fast Run on the Dionex IonPac AS18-4µm Column
0 1 2 3 4 5
-0.5
5.5
µS
Minutes
1
2
3
4
5 67
Column: Dionex IonPac AS18-4µm, 0.4 × 250 mm
Eluent Source: Dionex EGC-KOH Cartridge (Capillary)
Eluent: 35 mM KOH
Flow Rate: 30 µL/min
Inj. Volume: 0.4 µL
Col. Temp.: 30 °C
IC Cube Temp.: 15 C
Detection: Suppressed conductivity,
Dionex ACES 300 Suppressor, recycle
mode
Peaks: 1. Fluoride 0.2 mg/L (ppm)
2. Chloride 0.5
3. Nitrite 1.0
4. Sulfate 1.0
5. Bromide 1.0
6. Nitrate 1.0
7. Phosphate 2.0
8
Analytes in Flowback Wastewater Measured by IC
• Cations
• Potassium, sodium
• Impacts effectiveness of additives (reuse)
• Lithium
• Human toxicity
• Ammonium
• Corrosive
• Magnesium, calcium, barium, strontium
• Scale buildup
9
Challenge of Wastewater Analysis
High concentrations of dissolved salts:
• Exceed column capacity
• Poor chromatography
• Peak suppression
• Inaccurate reporting
• Exceed linear calibration range
• Analyte specific
• Inaccurate results
• Decrease column lifetime
0 2 4 11
0
12,000
µS
Minutes
6 8 10
0
50
µS
0 2 4 116 8 10
Minutes
Undiluted
Diluted
10
Obtaining Accurate Data From High Salt Samples
Manual Analysis
• Post-run
• Determine concentration from chromatogram peak area
• Exceed limit → dilute → re-run sample
• Pre-run
• Manual conductivity measurement
• Exceed limit → dilute → run sample
• Tedious
• Dilution prone to errors
11
Obtaining Accurate Data From High Salt Samples
Automated Analysis
• “AutoDilution”
• Post-run analysis using ion chromatograph software
• Exceeding peak height or area -> re-run with less sample loaded
• In-line Conductivity
• Conductivity measured prior to loading sample onto column
• Exceeding upper limit -> less sample loaded
Injecting Less Sample
• Smaller sample loop
• Partial loop
• Automated sample dilution
• Lower amount of sample loaded
Thermo Scientific™
Dionex™
Chromeleon™
Chromatography Data
System (CDS) Software
12
Automated Dilution: Precision and Accuracy
Cl (g/L) Avg. %RSD
2 6545 0.058
0.02 683.1 0.034
Draw/Dispense
Speed
(μL/sec)
Set Volume
(µL)
Liquid
Dispensed
(µg)
%RSD %Accuracy
50/25 4950 4926.9 0.0032 99.5
50/25 1980 1968.9 0.0747 99.4
10/5 50 50.5 0.38 99.0
10/5 20 20.1 0.67 98.4
N = 5 injections
N = 10 injections
13
Using a Gradient for Faster Cation Separation
Column: Dionex IonPac CG16/CS16
columns, 5 mm i.d.
Eluent Source: Dionex EGC III MSA
cartridge
Eluent (A): 30 mM MSA
Gradient (B): 20–30 mM MSA (0–10 min)
30–55 mM MSA (10–18 min)
55 mM MSA (18–32 min)
20 mM MSA (32–38 min)
Flow Rate: 1 mL/min
Inj. Volume: 25 µL
Col. Temp.: 40 °C
Detection: Suppressed conductivity,
Thermo Scientific™ Dionex™
CERS™ 500 Cation Eletrolytically
Regenerated Suppressor,
recycle mode
Sample: Mix of standards
0
2.9
µS
Minutes
1 2
35
46
7
8
0 30 6010 40 50
1 2
3
54
6
7
8
A
B
20
Peaks:
1. Lithium 0.1 mg/L 5. Magnesium 0.5
2. Sodium 0.4 6. Calcium 1.0
3. Ammonium 0.5 7. Strontium 5.0
4. Potassium 1.0 8. Barium 5.0
14
High Resolution Cation Analysis Using a DionexIonPac CS16 Column at Different Flow Rates
Minutes
µS
A
B
0 40-1
7
1 2 3
4
5
20
6 C
10 µL/min 1200 psi
20 µL/min 2400 psi
30 µL/min 3600 psi
Column: Dionex IonPac CS16,
2 x 250 mm x 0.5 mm ID
Eluant: 30 mmol/L MSA (EG)
Flow rate: A: 10 µL/min
B: 20 µL/min
C: 30 µL/min
Inj. volume: 0.4 µL
Temperature: 40 °C
Detection: Suppressed conductivity
Thermo Scientific™ Dionex™
CCES™ 300 Cation Capillary
Electrolytic Suppressor,
AutoSuppression, recycle
mode
Peaks: 1. Lithium 0.5 mg/L
2. Sodium 2.0
3. Ammonium 2.5
4. Potassium 5.0
5. Magnesium 2.5
6. Calcium 5.0
15
Determination of Cations in Flowback Wastewater
Column: Dionex IonPac CG16/CS16 columns,
5 mm i.d.
Eluent Source: Dionex EGCIII MSA cartridge
Gradient: 20–30 mM MSA (0–10 min)
30–55 mM MSA (10–18 min)
55 mM MSA (18–32 min)
20 mM MSA (32–38 min)
Flow Rate: 1 mL/min
Inj. Volume: 25 µL
Col. Temp.: 40 °C
Detection: Suppressed conductivity,
Thermo Scientific™ Dionex™ CERS™
500 Cation Electrolytically
Regenerated Suppressor, recycle
mode
Sample: 100-fold diluted flowback water, filtered,
0.2 µm
0
6.8
µS
Minutes
0 5 10 25
0
475
µS
Minutes
3
1
2
3
4
5
15
5
4
1
2
6
20
67
8
7 8
32
0 5 10 2515 20 32
30
30
Peaks:
Measured Undiluted
1. Lithium 0.33 mg/L 33 mg/L
2. Sodium 330.0 33,000
3. Ammonium 1.8 180
4. Potassium 5.9 590
5. Magnesium 13.0 1,300
6. Calcium 130.0 13,000
7. Strontium 14.0 1,400
8. Barium 2.1 210
16
Change in Cation Concentration of Flowback Water
0
5,000
10,000
15,000
20,000
25,000
30,000
35,000
40,000
45,000
1 2 3 4 5 6 7 8 9 10
Co
nce
ntr
atio
n (
mg
/L)
Fraction
Sodium
Calcium
0 gallons 140,000
17
Change in Cation Concentration of Flowback Water
0
200
400
600
800
1,000
1,200
1,400
1,600
1,800
2,000
1 2 3 4 5 6 7 8 9 10
Con
ce
ntr
atio
n (
mg
/L)
Fraction
Strontium
Magnesium
Potassium
0
50
100
150
200
250
300
350
400
1 2 3 4 5 6 7 8 9 10
Con
ce
ntr
atio
n (
mg
/L)
Fraction
Barium
Ammonium
Lithium
Ion composition → wastewater reuse or treatment
18
Conclusion
• Wastewater containing high salt can be challenging to
analyze
Automated sample pre-screening and dilution
Accurate and consistent determination of anions,
organic acids, and cations
Formulate wastewater reuse or treatment strategy
19
Technical and Application Notes
• TN 138: Accurate and Precise Automated Dilution and In-line
Conductivity Measurement Using the AS-AP Autosampler
Prior to Analysis by Ion Chromatography
• TN 139: Determination of Anions in Fracking Flowback Water
From the Marcellus Shale Using Automated Dilution and Ion
Chromatography
• AN 1094: Determination of Cations in Hydraulic Fracturing
Flowback Water from the Marcellus Shale
www.thermoscientific.com/ic