FATIMA - Fast Timing Array

Status Report

Alison Bruce on behalf of

L.M. Fraile for the FATIMA collaborationGrupo de Física Nuclear, Dpto. Física Atómica, Molecular y Nuclear,

Universidad Complutense, E-28040 Madrid, Spain

FATIMA Helsinki 2013

Tapered LaBr3(Ce) design

L.M. Fraile – GFN-UCM

� LaBr3(Ce) detectors delivered to Madrid last week

→ “Studsvik design” optimized for light collection

→ returned to Saint Gobain for repair (bad E resolution)

→ under tests…

38 mm

47 mm

19 mm

16.5 mm

30.5 mm

FATIMA Helsinki 2013

UK has ordered, received and tested 31 detectors

ø1.5”x2”

FATIMA Helsinki 2013

Detector Characterisation

J. McIntyre et al., NIM A 652, 1, 2011, 201-204

Activity: ~0.7 counts/sec./cm3 ~0.1 counts/sec/cm3

EC

β-decay

α

0-255 keV

788-1000 keV 1.5-3 MeV

O.J.Roberts et al

FATIMA Helsinki 2013

EXILL-FATIMA campaign, ILL Grenoble

� EXOGAM + Fast timing LaBr3(Ce) array→ Nuclear spectroscopy of prompt fission fragments

→ 19 proposals on 85 nuclides

→ Thermal neutron induced fission of 235U target.

→ EXOGAM + 16 LaBr3(Ce) with digital DAQ for energy and analog

TACs for timing

L.M. Fraile – GFN-UCM

FATIMA Helsinki 2013L.M. Fraile – GFN-UCM

Detectors in use at RIKEN

Eurica + 18 LaBr3

FATIMA Helsinki 2013L.M. Fraile – GFN-UCM

FATIMA Helsinki 2013

Beta-delayed gamma-ray spectrum in 104Zr

L.M. Fraile – GFN-UCM

F.Browne et al to be published

FATIMA Helsinki 2013

Half-life of the first 2+ in 104Zr

L.M. Fraile – GFN-UCM

Plastic – LaBr time

t1/2 in the literature is 2.0(3) ns

t1/2 = 2.0(5) ns

F.Browne et al to be published

FATIMA Helsinki 2013

Potential FATIMA configurations

� 5 mm Mu metal included: (73% Ni, 17% Fe, 6% Cu, 4% Cr).

� Pb shielding included: 5 mm

� Al Housing included: 0.5mm

� MgO Insulation included: 0.25 mm

� Five energies simulated: 50, 100, 150, 200, 250 and 500 keV

� Five source positions (“Cases”) simulated to determine response.

Case 2

Case 3

Case 4

Case 1

Case 5

Middle Det.

Back . Front

Simulations done by O.J.Roberts (Brighton)

FATIMA Helsinki 2013

Design 1

� Approx. 10 cm to front of crystals

� 3 rings of 12 detectors, (36 detectors in total).

� All detectors 90o relative to beam.

O.J.Roberts et al

FATIMA Helsinki 2013

Design 6

• 4 rings of 36 detectors.

• One ring of 12 at 15o.

• One ring of 12 at -15o.

• One ring of 8 at 45o.

• One ring of 4 at 20o.

• The front of each detector is 13.5 cm away from the middle of AIDA.

O.J.Roberts et al

FATIMA Helsinki 2013

Design 6 Results

� Since Case 2=Case 4=Case 3=Case 5, only look at Case 2 and Case 5 along with Case 1.

� Case 1 is slightly lower than the other cases.

� 106 events in each Case

Case 2

Case 5Case 3

Case 4Case 1 Back Det. Front Det.

Middle Det.

O.J.Roberts et al

FATIMA Helsinki 2013

Design 8

� 4 Rings of 32 detectors in a “tiled” arrangement:

→ 8 detectors at -45o

→ 12 detectors at 75o

→ 8 detectors at 45o

→ 4 detectors at 20o

Similar to Design 6, but with the

1st ring at -45o in approx. 2 cm

closer. For other rings, detector

distance is 13.5 cm.

O.J.Roberts et al

FATIMA Helsinki 2013

Design 8 Results

� Since Case 2=Case 4=Case 3=Case 5, only look at Case 2 and Case 5 along with Case 1.

� Case 1 is slightly lower than the other cases.

� 106 events in each Case

Case 2

Case 5Case 3

Case 4Case 1 Back Det. Front Det.

Middle Det.

O.J.Roberts et al

FATIMA Helsinki 2013

Conclusions

� Designs 8 and 6 are similar, and the best.

Design 6Design 8

• Designs 4 and 5 follow in terms of efficiency.

Design 4

Design 5

1st

2nd

3rd

O.J.Roberts et al

FATIMA Helsinki 2013

Fatima – provisional design

15°30° 28°

27°

Dave Seddon (Liverpool)/Ian Burrows (Daresbury)

FATIMA Helsinki 2013

Fatima + Aida

AIDA

SUPPORT STRUCTURES

EXISTING CARRIAGE SYSTEMS

FTA DETECTORS

MOUNTING BRACKETS

PROPOSED SETUP AT GSI

Dave Seddon (Liverpool)/Ian Burrows (Daresbury)

FATIMA Helsinki 2013

Timescales

L.M. Fraile – GFN-UCM

31 detector systems purchased by UK.

Spring 2014: Eurica + LaBr3 campaign (part 2) at RIKEN

Autumn 2014: Detectors used a focal plane of the

RITU/MARA separator at Jyvaskyla

Summer 2014: Test of array + AIDA + DAQ at GSI???