Post on 24-May-2020
MEDICAL POLICY – 1.01.30
Artificial Pancreas Device Systems
BCBSA Ref. Policy: 1.01.30
Effective Date: April 1, 2020
Last Revised: March 10, 2020
Replaces: N/A
RELATED MEDICAL POLICIES:
None
Select a hyperlink below to be directed to that section.
POLICY CRITERIA | DOCUMENTATION REQUIREMENTS | CODING
RELATED INFORMATION | EVIDENCE REVIEW | REFERENCES | HISTORY
∞ Clicking this icon returns you to the hyperlinks menu above.
Introduction
An artificial pancreas device system combines a glucose monitor and an insulin infusion pump.
The goal is to try to match how a normal pancreas would work. The pancreas releases insulin
based on changing levels of glucose in the blood. In this system, insulin is either withheld or
released based on the blood glucose level shown on the monitor. For those with type 1
diabetes, these systems may help improve overall glycemic control. They can be especially
helpful in controlling episodes of very low blood sugar at night. This policy discusses when an
artificial pancreas device system may be considered medically necessary.
Note: The Introduction section is for your general knowledge and is not to be taken as policy coverage criteria. The
rest of the policy uses specific words and concepts familiar to medical professionals. It is intended for
providers. A provider can be a person, such as a doctor, nurse, psychologist, or dentist. A provider also can
be a place where medical care is given, like a hospital, clinic, or lab. This policy informs them about when a
service may be covered.
Policy Coverage Criteria
Device Medical Necessity Artificial pancreas device
system
Use of a U.S. Food and Drug Administration (FDA) approved
automated insulin delivery system (artificial pancreas device
Page | 2 of 18 ∞
Device Medical Necessity system) with a low-glucose suspend feature may be considered
medically necessary in patients with type 1 diabetes who meet
all of the following criteria:
• Age 14 years of age or older
AND
• Glycated hemoglobin (hemoglobin A1c) value between 5.8%
and 10.0%
AND
• Used insulin pump therapy for more than 6 months
AND
• At least two documented nocturnal hypoglycemic events in a
two-week period (see definition below)
Hybrid closed loop insulin
delivery system
Use of a U.S. Food and Drug Administration (FDA) approved
automated insulin delivery system (artificial pancreas device
system) designated as hybrid closed-loop insulin delivery
system (with low glucose suspend and suspend before low
features) may be considered medically necessary in patients
with type 1 diabetes who meet all of the following criteria:
• Age 7 and older
AND
• Glycated hemoglobin level between 5.8% and 10.0%
AND
• Used insulin pump therapy for more than 6 months
AND
• At least 2 documented nocturnal hypoglycemic events in a 2-
week period
Device Investigational Automated insulin delivery
system
Use of an automated insulin delivery system (artificial
pancreas device system) is considered investigational for
individuals who do not meet the above criteria
Use of an automated insulin delivery system (artificial
pancreas device system) not approved by the Food and Drug
Administration is considered investigational.
Page | 3 of 18 ∞
Documentation Requirements The patient’s medical records submitted for review should document that medical necessity
criteria are met. The record should include clinical documentation of:
• Diagnosis/condition
• History and physical examination documenting the severity of the condition
• Hemoglobin A1c (glycated hemoglobin) results
• History of insulin pump usage
• Documentation of nighttime hypoglycemia events
Coding
Code Description
HCPCS S1034 Artificial pancreas device system (eg, low glucose suspend [LGS] feature) including
continuous glucose monitor, blood glucose device, insulin pump and computer
algorithm that communicates with all of the devices
S1035 Sensor; invasive (eg, subcutaneous), disposable, for use with artificial pancreas device
system, 1 unit = 1 day supply
S1036 Transmitter; external, for use with artificial pancreas device system
S1037 Receiver (monitor); external, for use with artificial pancreas device system
Note: CPT codes, descriptions and materials are copyrighted by the American Medical Association (AMA). HCPCS
codes, descriptions and materials are copyrighted by Centers for Medicare Services (CMS).
Related Information
Consideration of Age
The ages stated in this policy for which the artificial pancreas system may be considered
medically necessary is based on the FDA approved indications for the device.
Evidence Review
Page | 4 of 18 ∞
Description
Automated insulin delivery systems, also known as artificial pancreas device systems, link a
glucose monitor to an insulin infusion pump that automatically takes action (eg, suspends or
adjusts insulin infusion) based on the glucose monitor reading. These devices are proposed to
improve glycemic control in patients with insulin-dependent diabetes, in particular, reduction of
nocturnal hypoglycemia.
Background
Diabetes and Glycemic Control
Tight glucose control in patients with diabetes has been associated with improved health
outcomes. The American Diabetes Association has recommended a glycated hemoglobin level
below 7% for most patients. However, hypoglycemia, may place a limit on the ability to achieve
tighter glycemic control. Hypoglycemic events in adults range from mild to severe based on a
number of factors including the glucose nadir, the presence of symptoms, and whether the
episode can be self-treated or requires help for recovery. Children and adolescents represent a
population of type 1 diabetics who have challenges in controlling hyperglycemia and avoiding
hypoglycemia. Hypoglycemia is the most common acute complication of type 1 diabetes (T1D).
Table 1 is a summary of selected clinical outcomes in T1D clinical management and research.
Table 1. Outcome Measures for Type 1 Diabetes
Measure Definition Guideline type Organization Date
Hypoglycemia
Stakeholder survey,
expert opinion with
evidence review
Type 1 Diabetes
Outcome Programa1
2017
Level 1
Level 2
Level 3
Glucose
Page | 5 of 18 ∞
Measure Definition Guideline type Organization Date
Hypoglycemia Same as Type 1 Diabetes
Outcome Programa
Professional Practice
Committee with
systematic literature
review
ADA2 2019
Hypoglycemia
Clinical alert for
evaluation and/or
treatment
Clinically important or
serious
Severe hypoglycemia
Glucose 250 mg/dL
Type 1 Diabetes
Outcome Programa1
2017
Time in Rangeb Percentage of glucose
readings in the range of 70–
180 mg/dL per unit of time
Type 1 Diabetes
Outcome Programa
2017
Diabetic ketoacidosis
(DKA)
Elevated serum or urine
ketones > ULN
Serum bicarbonate
Page | 6 of 18 ∞
Hypoglycemia
Hypoglycemia affects many aspects of cognitive function, including attention, memory, and
psychomotor and spatial ability. Severe hypoglycemia can cause serious morbidity affecting the
central nervous system (eg, coma, seizure, transient ischemic attack, stroke), heart (eg, cardiac
arrhythmia, myocardial ischemia, infarction), eye (eg, vitreous hemorrhage, worsening of
retinopathy), as well as cause hypothermia and accidents that may lead to injury. Fear of having
hypoglycemia symptoms can also cause decreased motivation to adhere strictly to intensive
insulin treatment regimens.
The definition of a hypoglycemic episode is not standardized. In the pivotal Automation to
Simulate Pancreatic Insulin Response randomized controlled trial, a nocturnal hypoglycemic
episode was defined as a sensor glucose value of 65 mg/dL or less between 10 PM and 8 AM for
more than 20 consecutive minutes in the absence of a pump interaction within 20 minutes. In
2017, the American Diabetes Association defined serious, clinically significant hypoglycemia as
glucose levels
Page | 7 of 18 ∞
minimize the incidence and severity of hypoglycemic and hyperglycemic events. An APDS
control algorithm is embedded in software in an external processor or controller that receives
information from the CGM and performs a series of mathematical calculations. Based on these
calculations, the controller sends dosing instructions to the infusion pump.
Threshold Suspend Device System
Different APDS types are currently available for clinical use. Sensor augmented pump therapy
(SAPT) with low glucose suspend (LGS) (suspend on low) may reduce the likelihood or severity of
a hypoglycemic event by suspending insulin delivery temporarily when the sensor value reaches
(reactive) a predetermined lower threshold of measured interstitial glucose. Low glucose
suspension (LGS) automatically suspends basal insulin delivery for up to two hours in response
to sensor-detected hypoglycemia.
A sensor augmented pump therapy with predictive low glucose management (PLGM) (suspend
before low) suspends basal insulin infusion with the prediction of hypoglycemia. Basal insulin
infusion is suspended when sensor glucose is at or within 70 mg/dL above the patient-set low
limit and is predicted to be 20 mg/dL above this low limit in 30 minutes. In the absence of a
patient response, the insulin infusion resumes after a maximum suspend period of two hours. In
certain circumstances, auto-resumption parameters may be used.
When a sensor value is above or predicted to remain above the threshold, the infusion pump
will not take any action based on CGM readings. Patients using this system still need to monitor
their blood glucose concentration, set appropriate basal rates for their insulin pump, and give
premeal bolus insulin to control their glucose levels.
Control-to-Range System
A control-to-range system reduces the likelihood or severity of a hypoglycemic or
hyperglycemic event by adjusting insulin dosing only if a person's glucose levels reach or
approach predetermined higher and lower thresholds. When a patient's glucose concentration is
within the specified range, the infusion pump will not take any action based upon CGM
readings. Patients using this system still need to monitor their blood glucose concentration, set
appropriate basal rates for their insulin pump, and give premeal bolus insulin to control their
glucose levels.
Page | 8 of 18 ∞
Control-to-Target System
A control-to-target system sets target glucose levels and tries to maintain these levels at all
times. This system is fully automated and requires no interaction from the user (except for
calibration of the CGM). There are two subtypes of control-to-target systems: insulin-only and
bihormonal (eg, glucagon). There are no systems administering glucagon marketed in the
United States.
An APDS may also be referred to as a “closed-loop” system. A closed-loop system has
automated insulin delivery and continuous glucose sensing and insulin delivery without patient
intervention. The systems utilize a control algorithm that autonomously and continually
increases and decreases the subcutaneous insulin delivery based on real-time sensor glucose
levels. There are no completely closed-loop insulin delivery systems marketed in the United
States.
A hybrid closed-loop system also uses automated insulin delivery with continuous basal insulin
delivery adjustments. However, at mealtime, the patient enters the number of carbohydrates
they are eating in order for the insulin pump to determine the bolus meal dose of insulin. A
hybrid system option with the patient administration of a premeal or partial premeal insulin
bolus can be used in either control-to-range or control-to-target systems.
Source: https://www.fda.gov/medical-devices/artificial-pancreas-device-system/what-pancreas-what-
artificial-pancreas-device-system Accessed January 2020
https://www.fda.gov/medical-devices/artificial-pancreas-device-system/what-pancreas-what-artificial-pancreas-device-systemhttps://www.fda.gov/medical-devices/artificial-pancreas-device-system/what-pancreas-what-artificial-pancreas-device-system
Page | 9 of 18 ∞
These systems are regulated by the FDA as class III device systems.
Summary of Evidence
The following conclusions are based on a review of the evidence, including but not limited to,
published evidence and clinical expert opinion, solicited via BCBSA’s Clinical Input Process.
Low-Glucose Suspend Device
For individuals who have type 1 diabetes (T1D) who receive an artificial pancreas device system
with a low-glucose suspend feature, the evidence includes two randomized controlled trials
(RCTs) conducted in home settings. The relevant outcomes are symptoms, change in disease
status, morbid events, resource utilization and treatment-related morbidity. Primary eligibility
criteria of the key RCT, the Automation to Simulate Pancreatic Insulin Response (ASPIRE) trial,
were ages 16-to-70 years old, T1D, glycated hemoglobin levels between 5.8% and 10.0%, and at
least 2 nocturnal hypoglycemic events (≤65 mg/dL) lasting more than 20 minutes during a 2-
week run-in phase. Both trials required at least six months of insulin pump use. Both RCTs
reported significantly less hypoglycemia in the treatment group than in the control group. In
both trials, primary outcomes were favorable for the group using an artificial pancreas system;
however, findings from one trial were limited by nonstandard reporting of hypoglycemic
episodes, and findings from the other trial were no longer statistically significant when two
outliers (children) were excluded from analysis. The RCT limited to adults showed an
improvement in the primary outcome (area under the curve for nocturnal hypoglycemic events).
The area under the curve is not used for assessment in clinical practice but the current
technology does allow user and provider review of similar trend data with continuous glucose
monitoring. Results from the ASPIRE study suggested that there were increased risks of
hyperglycemia and potential diabetic ketoacidosis in subjects using the threshold suspend
feature. This finding may be related to whether or not actions are taken by the user to assess
glycemic status, etiology of the low glucose (activity, diet or medication) and to resume insulin
infusion. Both retrospective and prospective observational studies have reported reductions in
rates and severity of hypoglycemic episodes in automated insulin delivery system users. The
evidence is sufficient that the magnitude of reduction for hypoglycemic events in the T1D
population is likely to be clinically significant. Limitations of the published evidence preclude
determining the effects of the technology on overall glycemic control as assessed by HbA1c and
other parameters and thus, net health outcomes. Evidence reported through clinical input
supports that the outcome of hypoglycemia prevention provides a clinically meaningful
Page | 10 of 18 ∞
improvement in net health outcome, and this use is consistent with generally accepted medical
practice. The evidence is sufficient to determine that the technology results in a meaningful
improvement in the net health outcome.
Hybrid Closed-Loop Insulin Delivery System
For individuals who have T1D who receive an artificial pancreas device system with a hybrid
closed-loop insulin delivery system, the evidence includes multicenter pivotal trials using devices
cleared by the Food and Drug Administration, supplemental data and analysis for expanded
indications and more recent studies focused on children and adolescents. Three crossover RCTs
using a similar first-generation device approved outside the United States have been reported.
Relevant outcomes are symptoms, change in disease status, morbid events, resource utilization
and treatment-related morbidity. Of the three crossover RCTs assessing a related device
conducted outside the United States, two found significantly better outcomes (ie, time spent in
nocturnal hypoglycemia and time spent in preferred glycemic range) with the device than with
standard care and the other had mixed findings (significant difference in time spent in nocturnal
hypoglycemia and no significant difference in time spent in preferred glycemic range). For the
U.S. regulatory registration pivotal trial, the primary outcomes were safety and not efficacy.
Additional evidence from device performance studies and clinical studies all demonstrate
reductions in time spent in various levels of hypoglycemia, improved time in range (70-
180mg/dl), rare diabetic ketoacidosis and few device-related adverse events. The evidence is
sufficient that the magnitude of reduction for hypoglycemic events in the T1D population is
likely to be clinically significant. The variations in the definition of primary and secondary
outcomes in the study design and conduct of the published evidence are limitations that
preclude determining the effects of the technology on net health outcomes. Evidence reported
through clinical input supports that the use of hybrid closed loop APDS systems provides a
clinically meaningful improvement in net health outcome and is consistent with generally
accepted medical practice. Reduction in the experience of hypoglycemia and inappropriate
awareness of hypoglycemia and glycemic excursions were identified as important acute clinical
outcomes in children, adolescents, and adults and are related to the future risk for end-organ
complications. The evidence is sufficient to determine that the technology results in meaningful
improvement in the net health outcome.
Ongoing and Unpublished Clinical Trials
Some currently unpublished trials that might influence this review are listed in Table 2.
Page | 11 of 18 ∞
Table 2. Summary of Key Trials
NCT No. Trial Name Planned
Enrollment
Completion
Date
Ongoing
NCT02748018 Multi-center, Randomized, Parallel, Adaptive, Controlled
Trial in Adult and Pediatric Patients With Type 1 Diabetes
Using Hybrid Closed Loop System and Control (CSII, MDI,
and SAP) at Home
1500 Dec 2021
NCT03859401 Hypoglycemia Prevention During and After Moderate
Exercise in Adults With Type 1 Diabetes Using anArtificial
Pancreas With Exercise Behavior Recognition
33 Dec 2019
NCT02733211 An Open-label, Two-center, Randomized, Cross-over
Study to Evaluate the Safety and Efficacy of Night Closed-
loop Control Using the MD-LogicAutomated Insulin
Delivery SystemCompared to Sensor Augmented Pump
Therapy in Poorly Controlled Patients With Type 1
Diabetes at Home
28 Dec 2019
NCT03739099 Assessment of the Efficacy of Closed-loop Insulin Therapy
(Artificial Pancreas) on the Control of Type 1 Diabetes in
Prepubertal Child in Free-life: Comparison Between
Nocturnal and 24-hour Use on 18 Weeks, Followed by an
Extension on 18 Weeks
120 Sep 2020
NCT03844789a The International Diabetes Closed Loop (iDCL) Trial:
Clinical Acceptance of theArtificial Pancreas in Pediatrics:
A Study of t:Slim X2 With Control-IQ Technology
101 Mar 2020
NCT: national clinical trial.
a Denotes industry-sponsored or cosponsored trial.
Clinical Input from Physician Specialty Societies and Academic Medical
Centers
While the various physician specialty societies and academic medical centers may collaborate
with and make recommendations during this process, through the provision of appropriate
reviewers, input received does not represent an endorsement or position statement by the
physician specialty societies or academic medical centers, unless otherwise noted.
https://clinicaltrials.gov/ct2/show/NCT02748018https://clinicaltrials.gov/ct2/show/NCT03859401?cond=NCT03859401&draw=1&rank=1https://clinicaltrials.gov/ct2/show/NCT02733211?cond=NCT02733211&draw=1&rank=1https://clinicaltrials.gov/ct2/show/NCT03739099?cond=NCT03739099&draw=1&rank=1https://clinicaltrials.gov/ct2/show/NCT03844789?cond=NCT03844789&draw=1&rank=1
Page | 12 of 18 ∞
2019
In response to requests, while this topic was under review in 2019, clinical input on the use of an
artificial pancreas device system with a hybrid closed-loop insulin delivery system for individuals
with type 1 diabetes was received from 4 respondents, including 4 physician-level responses
identified through 2 specialty societies including physicians with academic medical center
affiliations. Evidence from clinical input is integrated within the Summary of Evidence.
2015
In response to requests, input on artificial pancreas device systems was received from 2
physician specialty societies and 4 academic medical centers when the policy was under review
in 2015. Input was mixed on whether artificial pancreas systems, including closed-loop
monitoring devices with a low-glucose suspend threshold feature, are considered medically
necessary. Most reviewers thought there was sufficient supportive data on devices with a low-
glucose suspend feature in patients at high risk of hypoglycemia, but some thought the data
insufficient.
Practice Guidelines and Position Statements
American Diabetes Association
The American Diabetes Association has released multiple publications on controlling type 1
diabetes (see Table 3).
Table 3. Recommendations on Diabetes
Date Title Publication
Type
Recommendation LOE
2019 Standards of Medical
Care in Diabetes
Guideline
standard
Automated insulin delivery systems improve
glycemic control and reduce hypoglycemia in
adolescents and should be considered in
adolescents with type 1 diabetes
B
Page | 13 of 18 ∞
Date Title Publication
Type
Recommendation LOE
2017 Standardizing
Clinically Meaningful
Outcome Measures
Beyond HbA1c for
Type 1 Diabetes
Consensus
report21,a
Developed definitions for hypoglycemia,
hyperglycemia, time in range, and diabetic
ketoacidosis in type 1 diabetes
N/A
HbA1c: hemoglobin A1c; LOE: Level of Evidence.
a Jointly published with American Association of Clinical Endocrinologists, the American Association of Diabetes
Educators, the Endocrine Society, JDRF International, The Leona M. and Harry B. Helmsley Charitable Trust, the
Pediatric Endocrine Society, and the T1D Exchange.
American Association of Clinical Endocrinologists et al
The American Association of Clinical Endocrinologists and American College of Endocrinology
(2018) published a joint position statement on the integration of insulin pumps and continuous
glucose monitoring in patients with diabetes.22 The statement emphasized the use of continuous
glucose monitoring and insulin pump therapy for type 1 diabetes patients who are not in
glycemic target ranges despite intensive attempts at self-blood glucose monitoring and multiple
insulin injection therapy.
Medicare National Coverage
There is no national coverage determination.
Regulatory Status
Table 4 summarizes the FDA-approved automated insulin delivery systems.
Page | 14 of 18 ∞
Table 4. FDA-Approved Automated Insulin Delivery Systems (Artificial
Pancreas Device Systems)
Device Age
Indication
Manufacturer Date
Approved
PMA No./
Device Code
MiniMed 530G Systema (open-
loop, LGS)
≥16 y Medtronic Jul 2013 P120010/OZO
MiniMed 630G System with
SmartGuard™b (open-loop, LGS)
≥16 y
≥14 y
Medtronic Aug 2016
Jun 2017
P150001/OZO
P150001/S008
MiniMed 670GSystemc (hybrid
closed-loop, LGS or PLGM)
≥14 y
≥7-13 y
Medtronic Sep 2016
Jul 2018
P160017/OZP
P160017/S031
FDA: Food and Drug Administration; LGS: low glucose suspend; OZO: Artificial Pancreas Device System, threshold
suspend; OZP: Automated Insulin Dosing Device System, Single Hormonal Control; PMA: premarket approval; PLGM:
predictive low glucose management.
a MiniMed 530G System consists of the following devices that can be used in combination or individually: MiniMed
530G Insulin Pump, Enlite™ Sensor, Enlite™ Serter, the MiniLink Real-Time System, the Bayer Contour NextLink
glucose meter, CareLink® Professional Therapy Management Software for Diabetes, and CareLink® Personal Therapy
Management Software for Diabetes (at time of approval).
b MiniMed 630G System with SmartGuard™ consists of the following devices: MiniMed 630G Insulin Pump, Enlite®
Sensor, One-Press Serter, Guardian® Link Transmitter System, CareLink® USB, Bayer’s CONTOUR ® NEXT LINK 2.4
Wireless Meter, and Bayer’s CONTOUR® NEXT Test Strips (at time of approval).
c MiniMed 670G System consists of the following devices: MiniMed 670G Pump, the Guardian Link (3) Transmitter, the
Guardian Sensor (3), One-Press Serter, and the Contour NEXT Link 2.4 Glucose Meter (at time of approval).
The MiniMed® 530G System includes a threshold suspend or LGS feature.5 The threshold
suspend tool temporarily suspends insulin delivery when the sensor glucose level is at or below
a preset threshold within the 60- to 90-mg/dL range. When the glucose value reaches this
threshold, an alarm sounds. If patients respond to the alarm, they can choose to continue or
cancel the insulin suspend feature. If patients fail to respond, the pump automatically suspends
action for two hours, and then insulin therapy resumes.
The MiniMed® 630G System with SmartGuard™, which is similar to the 530G, includes updates
to the system components including waterproofing.6 The threshold suspend feature can be
programmed to temporarily suspend delivery of insulin for up to two hours when the sensor
glucose value falls below a predefined threshold value. The MiniMed 630G System with
SmartGuard™ is not intended to be used directly for making therapy adjustments, but rather to
provide an indication of when a finger stick may be required. All therapy adjustments should be
based on measurements obtained using a home glucose monitor and not on the values
Page | 15 of 18 ∞
provided by the MiniMed 630G system. The device is not intended to be used directly for
preventing or treating hypoglycemia but to suspend insulin delivery when the user is unable to
respond to the SmartGuard™ Suspend on Low alarm to take measures to prevent or treat
hypoglycemia themselves.
The MiniMed® 670G System is a hybrid closed-loop insulin delivery system consisting of an
insulin pump, a glucose meter, and a transmitter, linked by a proprietary algorithm and the
SmartGuard Hybrid Closed Loop.7, The system includes an LGS feature that suspends insulin
delivery; this feature either suspends delivery on low-glucose levels or suspends delivery before
low-glucose levels, and has an optional alarm (manual mode). Additionally, the system allows
semiautomatic basal insulin-level adjustment (decrease or increase) to preset targets (automatic
mode). As a hybrid system; basal insulin levels are automatically adjusted, but the patient needs
to administer premeal insulin boluses. The CGM component of the MiniMed 670G System is not
intended to be used directly for making manual insulin therapy adjustments; rather it is to
provide an indication of when a glucose measurement should be taken.
The most recent supplemental approval for the MiniMed® 670G System in July 2018 followed
the granting a designation of breakthrough device status.
On June 21, 2018, the FDA approved the t:slim X2 Insulin Pump with Basal-IQ Technology (PMA
P180008) for individuals who are 6 years of age and older. The System consists of the t:slim X2
Insulin Pump paired with the Dexcom G5 Mobile CGM (Continuous Glucose Monitor), as well as
the Basal-IQ Technology. The t:slim X2 Insulin Pump is intended for the subcutaneous delivery of
insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring
insulin. The t:slim X2 Insulin Pump can be used solely for continuous insulin delivery and as part
of the System as the receiver for a therapeutic CGM. The t:slim X2 Insulin Pump running the
Basal-IQ Technology can be used to suspend insulin delivery based on CGM sensor readings.
Introduction into clinical care is planned for summer 2019.
References
1. Agiostratidou G, Anhalt H, Ball D, et al. Standardizing clinically meaningful outcome measures beyond HbA1c for type 1
diabetes: A Consensus Report of the American Association of Clinical Endocrinologists, the American Association of Diabetes
Educators, the American Diabetes Association, the Endocrine Society, JDRF International, The Leona M. and Harry B. Helmsley
Charitable Trust, the Pediatric Endocrine Society, and the T1D Exchange. Diabetes Care. Dec 2017;40(12):1622-1630. PMID
29162582.
2. American Diabetes Association. 6. Glycemic Targets: Standards of Medical Care in Diabetes-2019. Diabetes Care. Jan
2019;42(Suppl 1):S61-s70. PMID 30559232.
Page | 16 of 18 ∞
3. Abraham MB, Jones TW, Naranjo D, et al. ISPAD Clinical Practice Consensus Guidelines 2018: Assessment and management of
hypoglycemia in children and adolescents with diabetes. Pediatr Diabetes. Oct 2018;19 Suppl 27:178-192. PMID 29869358.
4. Food and Drug Administration (FDA). Guidance for Industry and Food and Drug Administration Staff: The Content of
Investigational Device Exemption (IDE) and Premarket Approval (PMA) Applications for Artificial Pancreas Device Systems
[draft]. 2012;
https://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/UCM259305.pdf
Accessed January 2020.
5. Food and Drug Administration (FDA). Premarket Approval (PMA): MINIMED 530G SYSTEM. 2013;
https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P120010 Accessed January 2020.
6. Food and Drug Administration (FDA). Premarket Approval (PMA): MINIMED 630G SYSTEM WITH SMARTGUARD(TM). 2016;
https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?ID=320606 Accessed January 2020.
7. Food and Drug Administration (FDA). Premarket Approval (PMA): MiniMed 670G System. 2016;
https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P160017 Accessed January 2020.
8. Blue Cross and Blue Shield Technology Evaluation Center (TEC). Artificial Pancreas Device Systems. TEC Assessments.
2013;Volume 28:Tab 14. PMID.
9. Bergenstal RM, Klonoff DC, Garg SK, et al. Threshold-based insulin-pump interruption for reduction of hypoglycemia. N Engl J
Med. Jul 18 2013;369(3):224-232. PMID 23789889.
10. Garg S, Brazg RL, Bailey TS, et al. Reduction in duration of hypoglycemia by automatic suspension of insulin delivery: the in-
clinic ASPIRE study. Diabetes Technol Ther. Mar 2012;14(3):205-209. PMID 22316089.
11. Ly TT, Nicholas JA, Retterath A, et al. Effect of sensor-augmented insulin pump therapy and automated insulin suspension vs
standard insulin pump therapy on hypoglycemia in patients with type 1 diabetes: a randomized clinical trial. JAMA. Sep 25
2013;310(12):1240-1247. PMID 24065010.
12. Agrawal P, Zhong A, Welsh JB, et al. Retrospective analysis of the real-world use of the threshold suspend feature of sensor-
augmented insulin pumps. Diabetes Technol Ther. May 2015;17(5):316-319. PMID 25611577.
13. Gomez AM, Marin Carrillo LF, Munoz Velandia OM, et al. Long-term efficacy and safety of sensor augmented insulin pump
therapy with low-glucose suspend feature in patients with type 1 diabetes. Diabetes Technol Ther. Feb 2017;19(2):109-114.
PMID 28001445.
14. Bergenstal RM, Garg S, Weinzimer SA, et al. Safety of a hybrid closed-loop insulin delivery system in patients with type 1
diabetes. JAMA. Oct 4 2016;316(13):1407-1408. PMID 27629148.
15. Garg SK, Weinzimer SA, Tamborlane WV, et al. Glucose outcomes with the in-home use of a hybrid closed-loop insulin delivery
system in adolescents and adults with type 1 diabetes. Diabetes Technol Ther. Mar 2017;19(3):155-163. PMID 28134564.
16. Forlenza GP, Deshpande S, Ly TT, et al. Application of zone model predictive control artificial pancreas during extended use of
infusion set and sensor: a randomized crossover-controlled home-use trial. Diabetes Care. Aug 2017;40(8):1096-1102. PMID
28584075.
17. Tauschmann M, Thabit H, Bally L, et al. Closed-loop insulin delivery in suboptimally controlled type 1 diabetes: a multicentre,
12-week randomised trial. Lancet. Oct 13 2018;392(10155):1321-1329. PMID 30292578.
18. Abraham MB, Nicholas JA, Smith GJ, et al. Reduction in Hypoglycemia With the Predictive Low-Glucose Management System: A
Long-term Randomized Controlled Trial in Adolescents With Type 1 Diabetes. Diabetes Care. Feb 2018;41(2):303-310. PMID
29191844.
19. Forlenza GP, Pinhas-Hamiel O, Liljenquist DR, et al. Safety Evaluation of the MiniMed 670G System in Children 7-13 Years of
Age with Type 1 Diabetes. Diabetes Technol Ther. Jan 2019;21(1):11-19. PMID 30585770.
20. Wood MA, Shulman DI, Forlenza GP, et al. In-Clinic Evaluation of the MiniMed 670G System "Suspend Before Low" Feature in
Children with Type 1 Diabetes. Diabetes Technol Ther. Nov 2018;20(11):731-737. PMID 30299976
https://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/UCM259305.pdfhttps://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P120010https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?ID=320606https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P160017
Page | 17 of 18 ∞
21. Messer LH, Forlenza GP, Sherr JL, et al. Optimizing Hybrid Closed-Loop Therapy in Adolescents and Emerging Adults Using the
MiniMed 670G System. Diabetes Care. Apr 2018;41(4):789-796. PMID 29444895.
22. American Diabetes Association. 7. Diabetes Technology: Standards of Medical Care in Diabetes-2019. Diabetes Care. Jan
2019;42(Suppl 1):S71-s80. PMID 30559233.
History
Date Comments 03/10/15 New Policy. Policy created with information on this topic previously addressed in Policy
No. 1.01.522 and a literature review through December 20, 2014. FDA-approved
artificial pancreas device system with low glucose suspend feature may be considered
medically necessary for patients with type 1 diabetes who meet criteria; otherwise
artificial pancreas device systems are considered investigational.
01/12/16 Annual Review. Added Related Policy 1.01.522 Continuous or Intermittent Monitoring
of Glucose in Interstitial Fluid. Policy updated with literature review through October 1,
2015; references added. Policy statements unchanged.
04/12/16 Minor update. Removal of related policy 1.01.522, policy was archived on April 30,
2016.
11/08/16 Minor update. Language added to support that this policy applies only to those age 16
and older as indicated by FDA approval for the use of the device.
02/01/17 Annual Review, approved January 10, 2017. Policy updated with literature review
through October 4, 2016; references added. Policy statements unchanged.
04/11/17 Policy moved into new format; no change to policy statements. Evidence Review
section reformatted.
02/01/18 Annual Review, approved January 16, 2018. Policy updated with literature review
through October 2017; references updated. Policy statement added that use of hybrid
closed loop insulin delivery system as an artificial pancreas device system (age 14 and
older) is considered investigational.
9/01/18 Minor update. Re-added language supporting that this policy applies to those age 16
and older; it was inadvertently removed in a previous update.
03/01/19 Minor update, added Documentation Requirements section.
07/01/19 Annual Review, approved June 11, 2019. Policy updated with literature review through
March 2019, references 1, 3-7, 13, 17, 18, and 20-24 added. Policy statements changed:
the age criterion changed in the first medically necessary statement; medically
necessary statement added on FDA-approved automated insulin delivery system
(artificial pancreas device system) designated as hybrid closed loop insulin delivery
system in patients with type 1 diabetes who meet specified criteria; and investigational
Page | 18 of 18 ∞
Date Comments statement added on use of an automated insulin delivery system (artificial pancreas
device system) for individuals who have not met specified criteria.
02/01/20 Annual Review, approved January 9, 2020. Policy updated with literature review
through September 2019; references added. Policy statements unchanged.
04/01/20 Delete policy, approved March 10, 2020. This policy will be deleted effective July 2,
2020, and replaced with InterQual criteria for dates of service on or after July 2, 2020.
Disclaimer: This medical policy is a guide in evaluating the medical necessity of a particular service or treatment. The
Company adopts policies after careful review of published peer-reviewed scientific literature, national guidelines and
local standards of practice. Since medical technology is constantly changing, the Company reserves the right to review
and update policies as appropriate. Member contracts differ in their benefits. Always consult the member benefit
booklet or contact a member service representative to determine coverage for a specific medical service or supply.
CPT codes, descriptions and materials are copyrighted by the American Medical Association (AMA). ©2020 Premera
All Rights Reserved.
Scope: Medical policies are systematically developed guidelines that serve as a resource for Company staff when
determining coverage for specific medical procedures, drugs or devices. Coverage for medical services is subject to
the limits and conditions of the member benefit plan. Members and their providers should consult the member
benefit booklet or contact a customer service representative to determine whether there are any benefit limitations
applicable to this service or supply. This medical policy does not apply to Medicare Advantage.
Discrimination is Against the Law
Premera Blue Cross complies with applicable Federal civil rights laws and does not discriminate on the basis of race, color, national origin, age, disability, or sex. Premera does not exclude people or treat them differently because of race, color, national origin, age, disability or sex.
Premera: • Provides free aids and services to people with disabilities to communicate
effectively with us, such as: • Qualified sign language interpreters • Written information in other formats (large print, audio, accessible
electronic formats, other formats) • Provides free language services to people whose primary language is not
English, such as: • Qualified interpreters• Information written in other languages
If you need these services, contact the Civil Rights Coordinator.
If you believe that Premera has failed to provide these services or discriminated in another way on the basis of race, color, national origin, age, disability, or sex, you can file a grievance with: Civil Rights Coordinator - Complaints and Appeals PO Box 91102, Seattle, WA 98111 Toll free 855-332-4535, Fax 425-918-5592, TTY 800-842-5357 Email AppealsDepartmentInquiries@Premera.com
You can file a grievance in person or by mail, fax, or email. If you need help filing a grievance, the Civil Rights Coordinator is available to help you.
You can also file a civil rights complaint with the U.S. Department of Health and Human Services, Office for Civil Rights, electronically through the Office for Civil Rights Complaint Portal, available at https://ocrportal.hhs.gov/ocr/portal/lobby.jsf, or by mail or phone at: U.S. Department of Health and Human Services 200 Independence Avenue SW, Room 509F, HHH Building Washington, D.C. 20201, 1-800-368-1019, 800-537-7697 (TDD) Complaint forms are available at http://www.hhs.gov/ocr/office/file/index.html.
Getting Help in Other Languages
This Notice has Important Information. This notice may have important information about your application or coverage through Premera Blue Cross. There may be key dates in this notice. You may need to take action by certain deadlines to keep your health coverage or help with costs. You have the right to get this information and help in your language at no cost. Call 800-722-1471 (TTY: 800-842-5357).
አማሪኛ (Amharic): ይህ ማስታወቂያ አስፈላጊ መረጃ ይዟል። ይህ ማስታወቂያ ስለ ማመልከቻዎ ወይም የ Premera Blue Cross ሽፋን አስፈላጊ መረጃ ሊኖረው ይችላል። በዚህ ማስታወቂያ ውስጥ ቁልፍ ቀኖች ሊኖሩ ይችላሉ። የጤናን ሽፋንዎን ለመጠበቅና በአከፋፈል እርዳታ ለማግኘት በተውሰኑ የጊዜ ገደቦች እርምጃ መውሰድ ይገባዎት ይሆናል። ይህን መረጃ እንዲያገኙ እና ያለምንም ክፍያ በቋንቋዎ እርዳታ እንዲያገኙ መብት አለዎት።በስልክ ቁጥር 800-722-1471 (TTY: 800-842-5357) ይደውሉ።
( ةالعربي :(. امةھ ماتولعم اإلشعار ھذا يحوي
خالل من ھاعلي صولحلا تريد لتيا التغطيةلل أو ةصحيلاكطيتتغ لىع اظلحفل نةعيم يخراوت في إجراء خاذتال تحتاج وقد .اإلشعار ھذا في
تكلفة أية بدتك دون بلغتك مساعدةوال تاوملالمع ھذه على ولحصال لك يحق .800-722-1471 (TTY: 800-842-5357)
أو طلبك وصخصب مةمھ ماتوعلم عارشإلا ھذا ويحي قدةمھم يخراوت ھناك تكون قد .Premera Blue Cross
اعدةمس تصلايفكالتال دفع فيبـ
.
Arabic
Oromoo (Cushite): Beeksisni kun odeeffannoo barbaachisaa qaba. Beeksisti kun sagantaa yookan karaa Premera Blue Cross tiin tajaajila keessan ilaalchisee odeeffannoo barbaachisaa qabaachuu danda’a. Guyyaawwan murteessaa ta’an beeksisa kana keessatti ilaalaa. Tarii kaffaltiidhaan deeggaramuuf yookan tajaajila fayyaa keessaniif guyyaa dhumaa irratti wanti raawwattan jiraachuu danda’a. Kaffaltii irraa bilisa haala ta’een afaan keessaniin odeeffannoo argachuu fi deeggarsa argachuuf mirga ni qabaattu. Lakkoofsa bilbilaa 800-722-1471 (TTY: 800-842-5357) tii bilbilaa.
Français (French): Cet avis a d'importantes informations. Cet avis peut avoir d'importantes informations sur votre demande ou la couverture par l'intermédiaire de Premera Blue Cross. Le présent avis peut contenir des dates clés. Vous devrez peut-être prendre des mesures par certains délais pour maintenir votre couverture de santé ou d'aide avec les coûts. Vous avez le droit d'obtenir cette information et de l’aide dans votre langue à aucun coût. Appelez le 800-722-1471 (TTY: 800-842-5357).
Kreyòl ayisyen (Creole): Avi sila a gen Enfòmasyon Enpòtan ladann. Avi sila a kapab genyen enfòmasyon enpòtan konsènan aplikasyon w lan oswa konsènan kouvèti asirans lan atravè Premera Blue Cross. Kapab genyen dat ki enpòtan nan avi sila a. Ou ka gen pou pran kèk aksyon avan sèten dat limit pou ka kenbe kouvèti asirans sante w la oswa pou yo ka ede w avèk depans yo. Se dwa w pou resevwa enfòmasyon sa a ak asistans nan lang ou pale a, san ou pa gen pou peye pou sa. Rele nan 800-722-1471 (TTY: 800-842-5357).
Deutsche (German): Diese Benachrichtigung enthält wichtige Informationen. Diese Benachrichtigung enthält unter Umständen wichtige Informationen bezüglich Ihres Antrags auf Krankenversicherungsschutz durch Premera Blue Cross. Suchen Sie nach eventuellen wichtigen Terminen in dieser Benachrichtigung. Sie könnten bis zu bestimmten Stichtagen handeln müssen, um Ihren Krankenversicherungsschutz oder Hilfe mit den Kosten zu behalten. Sie haben das Recht, kostenlose Hilfe und Informationen in Ihrer Sprache zu erhalten. Rufen Sie an unter 800-722-1471 (TTY: 800-842-5357).
Hmoob (Hmong): Tsab ntawv tshaj xo no muaj cov ntshiab lus tseem ceeb. Tej zaum tsab ntawv tshaj xo no muaj cov ntsiab lus tseem ceeb txog koj daim ntawv thov kev pab los yog koj qhov kev pab cuam los ntawm Premera Blue Cross. Tej zaum muaj cov hnub tseem ceeb uas sau rau hauv daim ntawv no. Tej zaum koj kuj yuav tau ua qee yam uas peb kom koj ua tsis pub dhau cov caij nyoog uas teev tseg rau hauv daim ntawv no mas koj thiaj yuav tau txais kev pab cuam kho mob los yog kev pab them tej nqi kho mob ntawd. Koj muaj cai kom lawv muab cov ntshiab lus no uas tau muab sau ua koj hom lus pub dawb rau koj. Hu rau 800-722-1471 (TTY: 800-842-5357).
Iloko (Ilocano): Daytoy a Pakdaar ket naglaon iti Napateg nga Impormasion. Daytoy a pakdaar mabalin nga adda ket naglaon iti napateg nga impormasion maipanggep iti apliksayonyo wenno coverage babaen iti Premera Blue Cross. Daytoy ket mabalin dagiti importante a petsa iti daytoy a pakdaar. Mabalin nga adda rumbeng nga aramidenyo nga addang sakbay dagiti partikular a naituding nga aldaw tapno mapagtalinaedyo ti coverage ti salun-atyo wenno tulong kadagiti gastos. Adda karbenganyo a mangala iti daytoy nga impormasion ken tulong iti bukodyo a pagsasao nga awan ti bayadanyo. Tumawag iti numero nga 800-722-1471 (TTY: 800-842-5357).
Italiano ( ):Questo avviso contiene informazioni importanti. Questo avviso può contenere informazioni importanti sulla tua domanda o copertura attraverso Premera Blue Cross. Potrebbero esserci date chiave in questo avviso. Potrebbe essere necessario un tuo intervento entro una scadenza determinata per consentirti di mantenere la tua copertura o sovvenzione. Hai il diritto di ottenere queste informazioni e assistenza nella tua lingua gratuitamente. Chiama 800-722-1471 (TTY: 800-842-5357).
Italian
中文 (Chinese):本通知有重要的訊息。本通知可能有關於您透過 Premera Blue Cross 提交的申請或保險的重要訊息。本通知內可能有重要日期。您可能需要在截止日期
之前採取行動,以保留您的健康保險或者費用補貼。您有權利免費以您的母
語得到本訊息和幫助。請撥電話 800-722-1471 (TTY: 800-842-5357)。
037338 (07-2016)
https://www.hhs.gov/ocr/office/file/index.htmlhttps://ocrportal.hhs.gov/ocr/portal/lobby.jsfmailto:AppealsDepartmentInquiries@Premera.com
日本語 (Japanese):この通知には重要な情報が含まれています。この通知には、 Premera Blue Crossの申請または補償範囲に関する重要な情報が含まれている場合があります。この通知に記載されている可能性がある重要な日付をご確認くだ
さい。健康保険や有料サポートを維持するには、特定の期日までに行動を
取らなければならない場合があります。ご希望の言語による情報とサポー
トが無料で提供されます。800-722-1471 (TTY: 800-842-5357)までお電話ください。
한국어 (Korean): 본 통지서에는 중요한 정보가 들어 있습니다 . 즉 이 통지서는 귀하의 신청에 관하여 그리고 Premera Blue Cross 를 통한 커버리지에 관한 정보를 포함하고 있을 수 있습니다 . 본 통지서에는 핵심이 되는 날짜들이 있을 수 있습니다. 귀하는 귀하의 건강 커버리지를 계속 유지하거나 비용을 절감하기 위해서 일정한 마감일까지 조치를 취해야 할 필요가 있을 수 있습니다 . 귀하는 이러한 정보와 도움을 귀하의 언어로 비용 부담없이 얻을 수 있는 권리가 있습니다 . 800-722-1471 (TTY: 800-842-5357) 로 전화하십시오 .
ລາວ (Lao): ແຈ້ງການນີ້ ນສໍ າຄັນ. ແຈ້ງການນີ້ອາດຈະມີ ນສໍ າຄັນກ່ຽວກັບຄໍ າຮ້ອງສະ ກ ຫຼື ຄວາມຄຸ້ມຄອງປະກັນໄພຂອງທ່ານຜ່ານ Premera Blue Cross. ອາດຈະມີ ນທີ າຄັນໃນແຈ້ງການນີ້. ທ່ານອາດຈະຈໍ າເປັ ນຕ້ອງດໍ າເນີ ນການຕາມກໍ ານົດ ເວລາສະເພາະເພື່ອຮັກສາຄວາມຄຸ້ມຄອງປະກັນສຸຂະພາບ ຫຼື ຄວາມຊ່ວຍເຫຼື ອເລື່ອງ າໃຊ້ າຍຂອງທ່ານໄວ້ . ທ່ານມີ ດໄດ້ ບຂໍ້ ນນີ້ ແລະ ຄວາມຊ່ວຍເຫຼື ອເປັ ນພາສາ ຂອງທ່ານໂດຍບ່ໍ ເສຍຄ່າ. ໃຫ້ໂທຫາ 800-722-1471 (TTY: 800-842-5357).
ູຂໍ້
່
ສໍ ັ
ຈ
ໝ
ສິ
ັ
່
ວ
ຄ
ມ
ມູຮັ
ູມີ ມຂໍ້
ភាសាែខមរ ( ): ឹ
រងរបស់
Premera Blue Cross ។ របែហលជាមាន កាលបរ ិ ឆ ំខានេនៅកងេសចក
េសចកតជី ូ
ជាមានព័ ៌ ៉ ងសំ ់អពី ់ ៉ ប់
នដំ ងេនះមានព័ ី
តមានយា ខាន ំ ទរមងែបបបទ ឬការរា
ណ ត៌មានយ៉ា ំ ់ តងសខាន។ េសចក
េចទស ់ ន ុ ត
ណងេនះ។ អ វការបេញញសមតភាព ដលកណតៃថ ចបាស
កតាមរយៈ
ដំ ឹ នករបែហលជារតូ ច ថ ់ ំ ់ ងជាក់ ់
នដ
ន
ី ន
ូ
អ
ូ
ជ
ជ
ំណឹងេនះរបែហល
នានា េដើ ីនងរកសាទុ ៉ បរងស់ ុ ់ ក ឬរបាក់ ំ
អ
មប ឹ កការធានារា ខភាពរបស ជ
ធនកមានសិ ទទលព័ មានេនះ និ ំ យេនៅកុងភាសារបសទិ ួ ត៌ ងជ ននួ
ន
់ កេដាយម
អ
នអ
យេចញៃថល។ ួ
នអស
ន
ិ
លុ ើ ូ ូយេឡយ។ សមទ ទ រស័ព 800-722-1471 (TTY: 800-842-5357)។
Khmer
ਕਵਰਜ ਅਤ ਅਰਜੀ ਬਾਰ ਮਹ ਤਵਪਰਨ ਜਾਣਕਾਰੀ ਹ ਸਕਦੀ ਹ . ਇਸ ਨ ਿਜਸ ਜਵਚ ਖਾਸ
ਤਾਰੀਖਾ ਹ ਸਕਦੀਆ ਹਨ. ਜੇਕਰ ਤਸੀ ਜਸਹਤ ਕਵਰਜ ਿਰਖਣੀ ਹਵ ਜਾ ਓਸ ਦੀ ਲਾਗਤ ਜਿਵਚ ਮਦਦ ਦ ੇਇਛ ੁਕ ਹ ਤਾਂ ਤਹਾਨ ਅ ਤਮ ਤਾਰੀਖ਼ ਤ ਪਿਹਲਾਂ ਕੁ ਝ ਖਾਸ ਕਦਮ ਚ ਕਣ ਦੀ ਲੜ ਹ ਸਕਦੀ ਹ ,ਤਹੁਾਨ ਮਫ਼ਤ ਿਵਚ ਤ ਆਪਣੀ ਭਾਸ਼ਾ ਿਵ ਚ ਜਾਣਕਾਰੀ ਅਤ ਮਦਦ ਪਾਪਤ ਕਰਨ ਦਾ ਅਿਧਕਾਰ ਹ ,ਕਾਲ 800-722-1471 (TTY: 800-842-5357).
ਪ ਜਾਬੀ (Punjabi): ਇਸ ਨ ਿਟਸ ਿਵਚ ਖਾਸ ਜਾਣਕਾਰੀ ਹ. ਇਸ ਨ ਿਟਸ ਿਵਚ Premera Blue Cross ਵਲ ਤੁਹਾਡੀ
ੰ
ੰ
ੇ ੇ ੇ ੱ ੂ ੋ ੈ ੋੋ ਂ ੁ ੇ ੱ ੋ ੇ ੱੱ ੁ ੱ ੂੁ ੱ ੇ ੱ ੇ ੍ਰ ੈ
ੋ ੰ ੂ ੱ ੁ ੋ ੋ ੈ ੰ
ੋ ੈ ੋ
(Farsi): فارسی فرم بارهدر ھمم اطالعات حاوی است ممکن يهمالعا اين. ميباشد ھمم اطالعات یوحا يهمالعا اين
در ھمم ھای خيتار به باشد.پ رایبستاکنممماش زينهھ اختدپر در مککيا تان بيمهوشش حقظ
Premera Blue Cross طريق از ماش مهبيوشش يا و تقاضا ای پ. يدماين جهتو يهمالعا اين
حق شما. يدشاب داشته اجتياح صیاخ کارھای امانج برای صیمشخ ایھ خيتار به تان، انیمدر ھای کسب برای .نماييد دريافت گانيرا ورط به ودخ زبان به را کمک و اطالعات اين که داريد را اين
استم ) 5357-842-800 مارهباش ماست TTY انکاربر(800-722-1471 مارهش با اطالعات .اييدنم برقرار
้
Polskie (Polish): To ogłoszenie może zawierać ważne informacje. To ogłoszenie może
zawierać ważne informacje odnośnie Państwa wniosku lub zakresu świadczeń poprzez Premera Blue Cross. Prosimy zwrócic uwagę na kluczowe daty, które mogą być zawarte w tym ogłoszeniu aby nie przekroczyć terminów w przypadku utrzymania polisy ubezpieczeniowej lub pomocy związanej z kosztami. Macie Państwo prawo do bezpłatnej informacji we własnym języku. Zadzwońcie pod 800-722-1471 (TTY: 800-842-5357).
Português (Portuguese): Este aviso contém informações importantes. Este aviso poderá conter informações importantes a respeito de sua aplicação ou cobertura por meio do Premera Blue Cross. Poderão existir datas importantes neste aviso. Talvez seja necessário que você tome providências dentro de determinados prazos para manter sua cobertura de saúde ou ajuda de custos. Você tem o direito de obter e sta informação e ajuda em seu idioma e sem custos. Ligue para 800-722-1471 (TTY: 800-842-5357).
Română (Romanian): Prezenta notificare conține informații importante. Această notificare poate conține informații importante privind cererea sau acoperirea asigurării dumneavoastre de sănătate prin Premera Blue Cross. Pot exista date cheie în această notificare. Este posibil să fie nevoie să acționați până la anumite termene limită pentru a vă menține acoperirea asigurării de sănătate sau asistența privitoare la costuri. Aveți dreptul de a obține gratuit aceste informații și ajutor în limba dumneavoastră. Sunați la 800-722-1471 (TTY: 800-842-5357).
Pусский (Russian): Настоящее уведомление содержит важную информацию. Это уведомление может содержать важную информацию о вашем заявлении или страховом покрытии через Premera Blue Cross. В настоящем уведомлении могут быть указаны ключевые даты. Вам, возможно, потребуется принять меры к определенным предельным срокам для сохранения страхового покрытия или помощи с расходами. Вы имеете право на бесплатное получение этой информации и помощь на вашем языке. Звоните по телефону 800-722-1471 (TTY: 800-842-5357).
Fa’asamoa (Samoan): Atonu ua iai i lenei fa’asilasilaga ni fa’amatalaga e sili ona taua e tatau ona e malamalama i ai. O lenei fa’asilasilaga o se fesoasoani e fa’amatala atili i ai i le tulaga o le polokalame, Premera Blue Cross, ua e tau fia maua atu i ai. Fa’amolemole, ia e iloilo fa’alelei i aso fa’apitoa olo’o iai i lenei fa’asilasilaga taua. Masalo o le’a iai ni feau e tatau ona e faia ao le’i aulia le aso ua ta’ua i lenei fa’asilasilaga ina ia e iai pea ma maua fesoasoani mai ai i le polokalame a le Malo olo’o e iai i ai. Olo’o iai iate oe le aia tatau e maua atu i lenei fa’asilasilaga ma lenei fa’matalaga i legagana e te malamalama i ai aunoa ma se togiga tupe. Vili atu i le telefoni 800-722-1471 (TTY: 800-842-5357).
Español ( ): Este Aviso contiene información importante. Es posible que este aviso contenga información importante acerca de su solicitud o cobertura a través de Premera Blue Cross. Es posible que haya fechas clave en este
tiene derecho a recibir esta información y ayuda en su idioma sin costo
aviso. Es posible que deba tomar alguna medida antes de determinadas fechas para mantener su cobertura médica o ayuda con los costos. Usted
alguno. Llame al 800-722-1471 (TTY: 800-842-5357).
Spanish
Tagalog (Tagalog): Ang Paunawa na ito ay naglalaman ng mahalagang impormasyon. Ang paunawa na ito ay maaaring naglalaman ng mahalagang impormasyon tungkol sa iyong aplikasyon o pagsakop sa pamamagitan ng Premera Blue Cross. Maaaring may mga mahalagang petsa dito sa paunawa. Maaring mangailangan ka na magsagawa ng hakbang sa ilang mga itinakdang panahon upang mapanatili ang iyong pagsakop sa kalusugan o tulong na walang gastos. May karapatan ka na makakuha ng ganitong impormasyon at tulong sa iyong wika ng walang gastos. Tumawag sa 800-722-1471 (TTY: 800-842-5357).
ไทย (Thai): ประกาศนมขอมลสาคญ ประกาศนอาจมขอมลทสาคญเกยวกบการการสมครหรอขอบเขตประกน สขภาพของคณผาน Premera Blue Cross และอาจมกาหนดการในประกาศน คณอาจจะตอง ดาเนนการภายในกาหนดระยะเวลาทแนนอนเพอจะรกษาการประกนสขภาพของคณหรอการชวยเหลอท มคาใชจาย คณมสทธทจะไดรบขอมลและความชวยเหลอนในภาษาของคณโดยไม่มคาใชจาย โทร 800-722-1471 (TTY: 800-842-5357)
้ี ี ้ ู ํ ั ้ี ี ้ ู ่ี ํ ั ่ี ั ั ื ัุ ุ ่ ี ํ ี ุ ้ํ ิ ํ ่ี ่ ่ื ั ั ุ ุ ื ่ ื ่ีี ่ ้ ่ ุ ี ิ ิ ่ี ้ ั ้ ู ่ ื ้ี ุ ี ่ ้ ่
Український (Ukrainian): Це повідомлення містить важливу інформацію. Це повідомлення може містити важливу інформацію про Ваше звернення щодо страхувального покриття через Premera Blue Cross. Зверніть увагу на ключові дати, які можуть бути вказані у цьому повідомленні. Існує імовірність того, що Вам треба буде здійснити певні кроки у конкретні кінцеві строки для того, щоб зберегти Ваше медичне страхування або отримати фінансову допомогу. У Вас є право на отримання цієї інформації та допомоги безкоштовно на Вашій рідній мові. Дзвоніть за номером телефону 800-722-1471 (TTY: 800-842-5357).
Tiếng Việt (Vietnamese): Thông báo này cung cấp thông tin quan trọng. Thông báo này có thông tin quan trọng về đơn xin tham gia hoặc hợp đồng bảo hiểm của quý vị qua chương trình Premera Blue Cross. Xin xem ngày quan trọng trong thông báo này. Quý vị có thể phải thực hiện theo thông báo đúng trong thời hạn để duy trì bảo hiểm sức khỏe hoặc được trợ giúp thêm về chi phí. Quý vị có quyền được biết thông tin này và được trợ giúp bằng ngôn ngữ của mình miễn phí. Xin gọi số 800-722-1471 (TTY: 800-842-5357).