FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY
Model Numbers:Service and
Parts Manual
QuickCleanUltrasonic
QC1-01QC3(R)-01QC6(R)-01
004-10061-00 Rev. AA8 (1/21/20)Language of origin: English
ii© Midmark Corporation 2018 004-10061-00 (07/11/18 )
Tabl
e of
Con
tent
sGENERAL INFORMATION
Symbols .................................................................iiiOrdering Parts .......................................................iiiSerial Number Location .......................................iiiSpecifications ....................................................... ivMaintenance ......................................................... vi
TROubLEshOOTINGView Cycle History ........................................... A-2View Temperature Sensor Readings ............. A-3Troubleshooting Chart ..................................... A-4
TEsTING & REpAIRTemperature Sensor Test ................................ B-2 Level Sensor Test ........................................ B-3Replace Level Sensor ...................................... B-4Test PSB Fuses (Power Supply & Generator PCB ) ... B-5Test Fuses (Main Power Socket) .............................. B-6Test IGBT’s ........................................................ B-7 Adjust LCD Display Contrast .......................B-8Replace LCD Display ....................................... B-9Replace Power Switch ................................... B-10Replace Power Socket ...................................B-11Ground Continuity Test .................................. B-12Isulation Resistance Test .............................. B-13 Transducer Testing .................................... B-14
Gen
eral
Info
Sect
ion
A
Sect
ion
E
Sect
ion
B
Sect
ion
CSe
ctio
n D
(*) Indicates multiple pages due to model / serial number break(s).
AccEss pROcEduREsRemove Base Plate - Table Top .................... C-2Remove Tank Assembly - Table Top ............. C-3Remove Recessed Unit from Counter Top ... C-4Remove Base Plate - Recessed ................... C-5Remove Tank Assembly - Recessed ............. C-6
WIRING dIAGRAMsTable Top Unit ................................................... D-2Recessed Unit ................................................... D-3Recessed Display Panel Assembly ............... D-4
pARTs LIsTTable Top (QC1-01, QC3-01,QC6-01) ................... E-2Recessed (QC3R-01, QC6R-01) .......................... E-3Basket .............................................................. E-12Beaker Holder Assembly ............................... E-13
iii© Midmark Corporation 2018 004-10061-00 (11/13/18)
Gen
eral
Info
rmat
ion WARNING
Indicates a potentially hazardous situation which could result in serious injury if not avoided.
Equipment Alert Indicates a potentially hazardous situation which could result in equipment damage if not avoided.
caution Indicates a potentially hazardous situation which may result in minor or moderate injury if not avoided. It may also be used to alert against unsafe practices
symbols Ordering partsThe following information is required when ordering parts: • Serialnumber&modelnumber • Partnumberfordesiredpart (Refer to Section E: Exploded Views & Parts Lists) For warranty parts orders, call Midmark’s Technical Service Department with the required information. Hours:8:00amto5:00p.m.EST(MondaythruFriday) Phone:1-800-Midmark(1-800-643-6275)
serial Number Location
NoteUsed for special instructions or additional information.
serial Number
iv© Midmark Corporation 2018 004-10061-00 (06/08/17)
SpecificationsOperating EnvironmentAmbientTemperature 41°Fto104°F(5to40°C)Maximumrelativehumidity 80%R.H.inroomtemperaturesupto88°F(31°C)decreasinglinearly
to50%R.H.at104°F(40°C)Altitudeabovesealevel Upto6,560ft(2000m)OperatingEnvironment Indooruseonly
UL 61010-1:2012CAN/CSA C22.2 No. 61010-1:2012
Safetyrequirementsforelectricalequipmentformeasurement,controlandlaboratoryuse
FCCCFR47Part15PARTB EMCComplianceEN 60601-1-2:2007 EMCCompliance
(ECmedicaldevicesemissionsGroup1ClassA)EN 61326:2013 EMCcompliance
(EC,emissionsGroup1ClassA)
Electrical Requirements
standards
NOTE: Iftheequipmentisusedinamannernotspecifiedbythemanufacturer,theprotectionprovidedbythe equipmentmaybeimpaired.
Connecttoa115VAC60Hzfullygroundedsupplyviaa3pinplug.Itcanbedangeroustooperateanultrasonicbathwithoutagrounded connection.
Thepowerplugisfittedwitha10Afuse(T10AH250V).NEVERFITAFUSEOFAHIGHERRATING.Ensurethatexcesspowercableisstoredneatly.
115ElectricalDetailsPower supply: 115V @ 60 HzUltrasonic frequency: 32 to 38kHzPollution degree: 2Installation category: II
Note: Supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.
v© Midmark Corporation 2018 004-10061-00 (06/08/17)
WorkingCapacity
Tank capacitymax.
TotalPower(W)
Ultrasonic power
(W)QC1 1.2 gal
(4.5L)1.3 gal(5L)
110 100
QC3(QC3R)
3.3 gal(12.5L)
3.4 gal (13 L)
210 200
QC6(QC6R)
6.6 gal(25L)
7.4 gal (28 L)
410 400
TankExternalDimensionsin(mm)
TankInternal Dimensionsin(mm)
BasketInternal Dimensionsin(mm)
A B C D E F G X Y Z
QC1 13.5(345)
7.8(200)
10.2(260)
10.4(265)
11.8(300)
5.9(150)
5.9(150)
10.4(265)
4.5(115)
4.3(110)
QC3 14.7(375)
13.5(345)
16.5(420)
10.4(265)
13.0(330)
11.8(300)
5.9 (150)
11.6(295)
10.4(265)
4.3(110)
QC6 21.6(550)
13.7 (350)
16.5(420)
14.3(365)
19.9(505)
11.8(300)
7.9(200)
18.3(465)
10.4(265)
6.3(160)
dimensions
capacity / power
vi© Midmark Corporation 2018 004-10061-00 (11/13/18)
Maintenance
Maintaining Your Ultrasonic
Itisimportanttokeepyourbathclean.Notonlywillcontaminatedliquidreducetheperformanceofthebath,itmayalsodamageit.Changethecleaningliquidregularly.Yourcleaningprocesswilldeterminehowoftentochangetheliquid–themoresoiledyouritems,themoreoftenyouwillneedtochangetheliquid.
Beforecleaningtheequipmentalwaysswitchoffanddisconnectitfromthepowersupplyandallowittocooldowntolessthan40°C/104°F.Cleanbywipingwithadampsoapyclothandrinse.
Thebaseofthebathgeneratestheultrasonicactivitybyvibratingatveryhighspeeds.Ifanycon-taminantsareincontactwiththebath,theyactasanabrasive,causingwearonthemetalsurface.Inextremecases,thebathwilldevelopholesandstarttoleak.
Perform Periodic “Foil” TestFollow the foil test guidelines in accordance with your local regulations.
Purpose: UsedtoDeterminetheEfficiencyofanUltrasonicCleaner.• Periodictestingoftheultrasoniccleanertoassurethatitisoperatingproperly.• Testingshouldalsobeperformedwheneverdeteriorationinthecleanerperformanceissuspected.• Foiltestarerelativelyeasytoperformandwillprovideapermanentrecordforthefuturecomparative
evaluationoftheultrasoniccleanersperformance.
A-1© Midmark Corporation 2018 004-10061-00 (06/08/17)
Sect
ion
ATroubleshooting
Calling for ServiceContactyourMidmarkAuthorisedDealer.Modelandserialnumberinformationwillberequiredwhencallingforserviceandcanbefoundonthebackoftheunit.
A-2© Midmark Corporation 2018 004-10061-00 (11/13/18)
Temp at Cycle End
Cycle Time Setting
View Cycle History View Cycle History.A) To view information about the last 5 cycles, press and hold the top and bottom arrow keys while turning the power switch on.
cycle Configuration
degas On
Leap On
Level On
cycle passed
1
2
3
4
5
6
789
10
11
12
13
14
15
TAbLE 1
Abort Reason description
1 Abort by user2 Abort by Level Low4 Abort by Low sonics
8 Abort by power Failure16 Abort Water Over Temperature32 Abort heater Over Temperature64 Abort unit Over Temperature
TAbLE 2
Top Arrow Key
Bottom Arrow Key
Temp at Start of Cycle °C Temp Setting for
Start of Cycle (N/A For This Model)
Cycle Configuration (SEE TABLE 1)
Aborted Cycle Reason (SEE TABLE 2)
Last 5 Cycles (A thru E)
Cycle Power Setting
A-3© Midmark Corporation 2018 004-10061-00 (11/13/18)
View Temperature Sensors
View Temperature Sensors
A) To view the current values of the temerature sensors on the display, press and hold the bottom arrow key while turning the power switch on.
Note: Temperature sensor failures for ch1 & ch3 would be indicated by the one of the two readings.
197 Open sensor
0 shorted sensor
Top Arrow Key
Bottom Arrow Key
Channel 1 Water Temp in °C Channel 1 °F Software Mode
N/A
Channel 2 (N/A For This Model). Should Display 50-51
Channel 3 Base Plate Temp in °C
A-4© Midmark Corporation 2018 004-10061-00 (11/13/18)
Troubleshooting chart
Problem Symptom Probable Cause Check CorrectionLiquid Level Low (repeatedly) Liquid Level Low (repeatedly) LiquidisnottouchingtheLevel
Sensor.CheckthattheLevelSensorandsurroundingareaiscleanof contaminant.
Filltanktorequiredlevel.
Detergentnotadded. (watertopure)
Checkcorrectdetergentisbeingused.
Adddetergent.
Disconnection or failure of Level SensorComponent.
CheckconnectiontoControlBoardandresistanceofLevelSensor.(See Level Sensor Test)
ReconnectorreplaceLevelSensoriffaulty.(see testing & repair)
Over Temp (repeatedly) Over Temp (repeatedly) Liquidisabove60°C/140°F. (TankTemperatureSensor)
Checkfluidtemperature,ensuretemperatureisbelow60°C/140°F.
Allowfluidtocoolordrainsomefluid&replacewithcoolerfluid.
Rapid Temperature Fluctuations
Rapid Temperature Fluctuations
Disconnection or failure of TemperatureSensor.
CheckconnectiontocontrolboardandresistanceofTemperatureSensor. (See Temp Sensor Test)
ReconnectorreplaceTemperatureSensoriffaulty.(see testing & repair)
Sensor Fault / Over Temp (in Cycle)
Sensor Fault / Over Temp (in Cycle).Erratic temp readings
TemperatureSensoropencircuitordisconnected.
CheckconnectiontoControlBoardandresistanceofTemperatureSensor.(See Temp Sensor Test)
ReconnectorreplaceTemperatureSensoriffaulty.(see testing & repair)
Sensor Fault / (in Standby) Sensor Fault / (in Standby) TemperatureSensorclosecircuit.
CheckresistanceofTemperatureSensor.(See Temp Sensor Test)
ReplaceTemperatureSensoriffaulty. (see testing & repair)
Unit Over Temp Unit Over Temp Transistorshavereached100°C/212°F. (BasePlateTemperatureSensor)
Checkwithuserthefrequencyofuse(avoidconsistentusewithouttimetocool).
Allowtimetocool.
Checkthatthebathhasnotbeenoverloaded(largemass).
Removeexcessiveitemsfromtankandrestartcycle.
Checkthatthebathhasnotbeenrunwithoutsufficientfluid.
Filltanktorequiredlevel.
Checkthatthelevelsensorhasnotbeenturnedoffintheadvancedmenuandrunwithoutsufficientfluid.
sonics Low Sonics Low Sonicsfeedbackexceedinglimits.
Checkthatthetankhasnotbeenoverloaded(largemass).
Removeexcessiveitemsfromtankandrestartcycle.
UltrasonicGeneratorPCBFuseblown.
Check2xFusesonUltrasonicGeneratorPCB.(See IGBT Test)
Replacefuses.(See PCB Fuses)
IGBTsblown. Check2xIGBTsonUltrasonicGeneratorPCB.(See IGBT Test)
ReplaceBasePlateAsm.
WARNING some procedures require power to be connected with covers removed. Line voltage is present. use extreme caution to prevent electric shock.
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
A-5© Midmark Corporation 2018 004-10061-00(9/24/18)
Troubleshooting chart
Problem Symptom Probable Cause Check CorrectionNo Display No Display Nopower. Checkthatmainpowerisconnect-
ed,CheckthatmainPowerSwitchisinONposition.
Connectmainsandswitchonunitatthe rear.
Nopower. Checkmainfuses. Replacefusesatrearofbath.
RibboncablebetweenControlPCBandPowerSupplyPCBis disconnected.
Checkribboncableconnection. Reconnectribboncable.
PowerSupplyPCBfusesblown.
CheckPowerSupplyPCBfuses. ReplacePowerSupplyPCBfuses.
FaultyLCD. Checkalltheabove. ReplaceLCD.
Backlight without text on LCD Display
Backlight without text on LCD Display
Contrast not set. Checkiftextappearswhile adjustingthepotontheControlPCB.
Adjustcontrast. (LCD Display Contrast)
Pixels missing from LCD Display
Pixels missing from LCD Display
FluidenteredLCDDisplay. Eliminatethesourceoffluid. ReplaceLCDDisplay.
Touch Pad buttons not responding
Touch Pad buttons not responding
FluidenteredTouchPad. Eliminatethesourceoffluid. ReplaceTouchPad.
FaultyTouchPad. Checkalltheabove. ReplaceTouchPad.
Fluid Leaking Fluid Leaking Plumbing. Checkallplumbingjoints. Tighten,sealorreplaceaffectedjoints.
Severe cavitation erosion. Checkforexcessivepittinginthebaseofthetank,visiblefromoutside.
Replaceultrasonic.
No Power No Power Notpluggedintopoweroutlet. Checkmainpowerconnection. Connecttopower.
IECfusesblown. Over current. ReplaceIECFuses.(See Power Socket Fuses)
FaultyPowerSwitch. CheckswitchilluminatesinOnposition.
Ifnotilluminated,replacePowerswitch. (See Power Switch)
GFIistripped CheckifGFIistripped ResetGFIWeak/FaultyGFIoutlet. Checkifsterilizerworkson
differentGFIoutletContactanelectriciantoreplaceGFIoutlet
Voltageleakthrough components
Checkallelectricalcomponentsandwireconnectionsfor deterioration
Secure/Clean/Replace compromisedcomponent.
B-1© Midmark Corporation 2018 004-10061-00 (11/13/18)
Sect
ion
BTesting & Repair Temperature Sensor Test ................................ B-2
Replace Tank Temperature Sensor ............. B-3Replace Base Plate Temperature Sensor .... B-4Level Sensor Test ............................................. B-5Replace Level Sensor ...................................... B-6Test PSB Fuses (Power Supply & Generator PCB ) ... B-7Test Fuses (Main Power Socket) .............................. B-8Test IGBT’s ........................................................ B-9 Adjust LCD Display Contrast ..................... B-10Replace LCD Display ......................................B-11Replace Power Switch ................................... B-12Replace Power Socket .................................. B-13
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
B-2© Midmark Corporation 2018 004-10061-00 (11/13/18)
Temperature Sensor Test NoteIt is not necessary to remove tank from case to test Temperature Sensor. Only remove tank when tank sensor replacement is necessary.
Control PCB
Degrees°F 0 1 2 3 4 5 6 7 8 950 103.90 104.12 104.34 104.55 104.77 104.98 105.20 105.42 105.63 105.85
60 106.07 106.28 106.50 106.71 106.93 107.15 107.36 107.58 107.79 108.01
70 108.23 108.44 108.66 108.87 109.09 109.30 109.52 109.73 109.95 110.17
80 110.38 110.60 110.81 111.03 111.24 111.46 111.67 111.89 112.10 112.32
90 112.53 112.75 112.96 113.18 113.39 113.61 113.82 114.04 114.25 114.47
100 114.68 114.90 115.11 115.33 115.54 115.76 115.97 116.18 116.40 116.61
110 116.83 117.04 117.26 117.47 117.68 117.90 118.11 118.33 118.54 118.76
120 118.97 119.18 119.40 119.61 119.82 120.04 120.25 120.47 120.68 120.89
130 121.11 121.32 121.53 121.75 121.96 122.18 122.39 122.60 122.82 123.03
140 123.24 123.46 123.67 123.88 124.09 124.31 124.52 124.73 124.95 125.16
Resistance in Ω (Ohms)To Tank SensorTo Base Plate Sensor
Test Temperature Sensor.A) Turn power off, unplug and drain water from tank.B) Turn unit upside down to remove bottom plate. (see access procedures).C) Set meter to Ώ.D) Remove appropriate red & white Temp Sensor wires from the Control PCB by pushing down on the tab and pulling out the wire.E) Measure resistance, refer to the table below. Resistance should correspond with the ambient temperature. i.e 75°F should be approximately109.3Ω. F) If the reading is not representative of the table, replace the sensor. (see testing & repair).
To Tank Sensor
Base Plate Temperature SensorRecessed
Table Top
Tank Temperature Sensor
To Base Plate Sensor
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
B-3© Midmark Corporation 2018 004-10061-00 (11/13/18)
Test Fill Level Sensor.A) Turn power off, unplug and drain water from tank.B) Remove the Base Plate following the procedures shown in this manual (see access procedures). C)SetmetertoΏD) Remove yellow level sensor wire from the control PCB.E) Measure the conductivity between the level sensor and the tank (Test A) should read open. If the conductivity is not open, check sensor insulation or replace. (see level sensor replacement)F) Measuring the conductivity between the level sensor inside the tank (TestB)andtheendofthecableshouldreadaround1Ω.Checkforthe correct assembly of the Level sensor. If conductivity is open, replace sensor. (see level sensor replacement)
Fill Level Sensor Test
Test Meter Reading Required ActionTest A OL when tank
is emptyIftheconductivityisnotopen,checksensor insulationorreplace.
Test b < 1 ohm Wire Harness OK Replaceifreadings
are out of range.
Fill Level Sensor
TEST BTEST
A
NoteIt is not necessary to remove tank from case to test Level Sensor. Only remove tank when tank sensor replacement is necessary.
Control PCB
To Fill Level Sensor
To Fill Level Sensor
Recessed
Table Top
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
B-4© Midmark Corporation 2018 004-10061-00 (11/13/18)
Replace Fill Level sensor
step 1: Replace Fill Level sensor.A) Turn power off, unplug and drain water from tank.B) Remove the tank following the procedures shown in this manual (see access procedures).C) Remove yellow wire from Control PCB by pressing on tab and pulling wire out of connector.D) Remove nuts (2), plain washers (3), fill level sensor harness & Nylon washers (2) from outside of tank. E) Remove screw and nylon insulator from in side of tank. F) Reassemble in reverse order.
M3 x 12mm screw and Insulator are inside tank
Nylon Washer
Level Sensor Harness
M3 Nyloc Nut
Nylon Insulator
M3x12 Dome Head Screw
M4 Plain Washer
M3 Plain Nut
NoteFill level sensor screw and plain nut must be tight enough to form a good seal. Overtightening will lead to insulator deforming and possible leaking.
M3 Plain Washer
Control PCB
To Fill Level Sensor
To Fill Level Sensor
Recessed
Table Top
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
B-5© Midmark Corporation 2018 004-10061-00 (11/13/18)
check pcb Fuses
Power Supply PCB
FUSE 2
FUSE 1
Ultrasonic Generator PCB
FUSE 2
FUSE 1
Test Meter Reading Required ActionFuse conductivity test 0-1Ω. Fuseisgood
Open or OLFuseisbadreplacewith
samevaluefuse..
Test Fuses.A) Turn power off, unplug and drain water from tank.B) Turn unit upside down and remove bottom plate (see access procedures).C)SetmetertoΩD) Remove fuses from top of holder E) Measuringacrossthefuseshouldread0-1Ω.Iffusemeasuresopenor OL, replace the fuse.
NoteTo remove the fuses on the Power Supply, & Ultrasonic Generator PCB’s carefully pull up on the top of the holder.
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
B-6© Midmark Corporation 2018 004-10061-00 (11/13/18)
check Main power socket Fuses
step 1: Test Main power socket Fuses.A) Unplug the power cord from the main power socket at the rear of the unit. B) Grip top and bottom tabs of the fuse cover located at the bottom of the main power socket and pull straight out.C) Rotate the end of the fuse cover down to expose the fuses.D) Remove fuses, test, and replace any bad fuses with fuses of the same rating.
Tabs
Fuse Cover
Main Power Socket
Test Meter Reading Required ActionFuse conductivity test 0-1Ω. Fuseisgood
Open or OLFuseisbadreplacewith
samevaluefuse..
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
B-7© Midmark Corporation 2018 004-10061-00 (11/13/18)
Ultrasonic Generator PCB
Test IGBT’s. A) Turn power off, unplug and drain water from tank.B) Turn unit upside down and remove bottom plate (see access procedures).C) Set meter to D) Place the black meter lead on the center leg of the IGBT and the red lead on one of the side legs then the other.E) Each measurement should be between 350 - 700.
Test Meter Reading Required ActionIGbT Test between
350 - 700Readingofbetween 350-700forbothtest theIGBTisgood
A reading outsideof 350 - 700
Readingoutsideof 350-700foreithertestIGBTisbadandneeds
replaced
check IGbT’s
IGBT
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
B-8© Midmark Corporation 2018 004-10061-00 (11/13/18)
Adjusting Lcd display contrast
Maximum Contrast
Optimum Contrast
Minimum Contrast
step 1: Adjusting Lcd contrast.A) Gain access to the Control PCB using the procedures shown in this manual. (see LCD display removal)B) Locate (VR7) on the Control PCB & turn it to adjust the LCD display to the optimum contrast level shown. C) Rotating Pot clockwise will make dispaly lighter and counterclockwise will make display darker.
WARNING some procedures require power to be connected with covers removed. Line voltage is present. use extreme caution to prevent electric shock.
B-9© Midmark Corporation 2018 004-10061-00 (11/13/18)
Replace Lcd display Table Top
step 1: Remove and Replace Lcd display.A) Turn power off, unplug and drain water from tank.B) Remove the tank following the procedures shown in this manual. (see access procedures) C) Remove sensor wires going to the Control PCB by pressing in on tab & pulling on wire. D) Remove ribbon cable by pulling outward on the side tabs and removing connector.E) Remove front cover mounting torx screws(6), remove front cover.F) Remove outer mounting torx screws (4) attaching the Control PCB and LCD Display .G) Disconnect the touch pad ribbon cable connecting the Control PCB to the Touch Pad.H) Remove stand off screws holding the two circuit boards together.I) Separate the boards by pulling them apart at the connection.J) Reassemble unit with new LCD Display by reversing these directions.
Mounting Screws
Front cover Mounting screws
Ribbon CableConnection
Remove Wires
Board Connection
Touch Pad Ribbon Cable
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
B-10© Midmark Corporation 2018 004-10061-00 (11/13/18)
Replace power switch
step 1: Replace power switchA) Turn power off, unplug and drain water from tank.B) Remove Bottom Plate Assembly (see access procedures). C) Remove the brown & blue wires from the back of the switch.D) Press in on the side tabs and push the switch out of the case.E) Replace switch, install wires and re-assemble.
To Case
Main Filter Power Socket
Power Switch
Power Switch
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
B-11© Midmark Corporation 2018 004-10061-00 (11/13/18)
Replace power socket
Power Socket
To Case
Main Filter Power Socket
Power Switch
step 1: Replace power socketA) Turn power off, unplug and drain water from tank.B) Remove Bottom Plate Assembly (see access procedures). C) Remove the brown, blue & green wires from the back of the socket.D) Remove the screws and pull the socket out of the case.E) Replace socket, install wires and re-assemble.
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
B-12© Midmark Corporation 2018 004-10061-00 (11/13/18))
Ground continuity Test
step 1: Ground continuity TestA) Turn power off and unplug main power socket at the rear of the unit. B) Visually inspect cord, plugs, and etc. for damage or cracks.C) Set meter to continuity test mode as shown.D) Connect one test lead to the power plug ground pin and other lead to a conductive surface on the QuickClean™ Ultrasonic Cleaner.
Note: Refer to multimeter instructions for specific operation and testing procedures.
Continuity Test Mode
Ground Pin
Conductive Surface
Test Meter Reading Required ActionGround continuity Test Less than 1Ω.
/ Meter beeps Groundisgood
Greater than 1Ω/ Meter doesn’t
beep
Groundisbad.Verifythatcordsandwiringareundamaged,routed,connectedproperlythentestunitagain.Ifunitcontinuestofail,
contactMidmark’sTechnicalService,
NoteUse a standard multimeter to perform this test.
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
B-13© Midmark Corporation 2018 004-10061-00 (11/13/18)
Insulation Resistance Test
step 1: Insulation Resistance TestA) Turn power off and unplug main power socket at the rear of the unit. B) Connect large test lead to both the hot and neutral pin and connect the other test lead to the ground pin as shown.C) Push the power switch to the on position on the QuickClean™ Ulrasonic Cleaner.D) Set Insulation resistance meter to 500V and run insulation test.
Note: Refer to insulation resistance tester instructions for specific operation and testing procedure.
Meter Setting
Large Test Lead
Test Meter Reading Required Action
Insulation Resistance Test
Greater than 1 MΩ. Insulationisgood.
Less than 1 MΩ.
Insulationisbad.Verifythatcordsandwiringareundamaged,routed,&connectedproperlythentestunitagain.Ifunitcontinuestofail,
contactMidmark’sTechnicalService.
WARNING To prevent electrical shock, fire, or personal injury, do not touch the Quickclean™ ultrasonic cleaner during testing and make sure the test has been completed before disconnecting.
Equipment Alert Only perform Insulation Resistance Test after performing and passing the Ground Continuity Test.
Ground Pin
NoteAn Insulation Resistance Tester is required to perform this test.
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
B-14© Midmark Corporation 2018
Generator Board
004-10061-00(4/30/19)
step 1: Inspect Transducer ceramic Rings.A) Turn power off and unplug main power socket at the rear of the unit. B) Remove bottom cover. Visually inspect the Ceramic Rings on each transducer for cracks or missing pieces.C) If all transducers are not in good condition Contact Midmark’s Technical Service.
Test Meter Reading Required Action
Transducer Resistance / “Ω” Ohm
Test
Open circuit Resistanceisgood.
closed circuit Resistanceisbad.ContactMidmark’s TechnicalService.
Transducer capacitance Test
=/- 5%
Qc1 = 7nFQc3 = 14nF Qc6 = 28nF
Capacitanceinrange.Transducersaregood.
Capacitanceoutofrange.TransducersarebadContactMidmark’sTechnicalService.
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
Transducer Testing step 2: Test Transducers.A) Disconnect red and black leads from Generator Board, briefly connect them together to discharge the transducers. B) Use meter to measure resistance across the red and black wires. C) If meter is equipped, change the meter setting to measure capacitance, connect leads to the red and black wires
Ceramic Rings
Bottom of Tank
C-1© Midmark Corporation 2018 004-10061-00 (06/08/17)
Sect
ion
CAccess procedures
Remove Base Plate - Table Top .................... C-2Remove Tank Assembly - Table Top ............. C-3Remove Recessed Unit from Counter Top ... C-4Remove Base Plate - Recessed ................... C-5Remove Tank Assembly - Recessed ............. C-6
C-2© Midmark Corporation 2018 004-10061-00 (11/13/18)
step 1: Remove base plate Assembly.A) Turn power off, unplug, and drain water from tank and place the front (display side) of ultrasonic down on a soft cloth to prevent damage to the display.B) Remove the four rubber feet (each corner) and screw(s) (number of screws depends on model) from bottom of the ultrasonic.C) Separate bottom plate from ultrasonic and rotate top of plate down and away from unit. This allows you to place the bottom plate on the table with out disconnecting wires.
Remove base plate Assembly (Table Top)WARNING
To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
C-3© Midmark Corporation 2018 004-10061-00 (11/13/18)
step 1: Remove base plate & disconnect Wiring.A) Remove Bottom Plate Assembly. (see base plate removal)B) Disconnect red and black power wires at the Ultrasonic Generator PCB by loosening the screws on the connector and removing the wires. C) Disconnect yellow level sensor wire & red and white thermal sensor wires (going to side of tank) from Control PCB using a small screwdriver to push on the tabs while pulling wires from connector.
step 2: Remove Tank Assembly.A) Remove nyloc nuts, washers & terminal holding the ground wire from the tank studs in the corners.B) Depress and hold the quick fitting elbow retaining ring (shown in red) toward the elbow to release it’s grip on the metal tube coming from the bottom of the tank. C) Slowly remove the tank from the case. The metal tube extending into the elbow will come out of the elbow fitting as the tank is removed.D) Reassemble in reverse order.E) Replace any zip ties that were cut during disassembly.
Remove Tank Assembly (Table Top)
Ultrasonic Generator PCB
Power Supply PCB
Quick Fitting Elbow
Red and White Thermal Sensor Wires
Quick Fitting Elbow
M4 Nyloc Nut, M5 Washer, & Ground Terminal
Yellow Level Sensor Wire
Equipment Alert To prevent damage to the wires ensure they are routed correctly and secured with zip ties.
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
C-4© Midmark Corporation 2018 004-10061-00 (11/13/18)
step 1: Remove ultrasonic From counter Top.A) Unthread and remove remote panel cable plug from socket at bottom of unit. B) Remove power cord from power connection and hose from drain elbow.C) Locate retaining clips in each corner holding onto under side of counter top, unthread screws almost all the way & pull down on bottom of clips.D) Rotate bottom of retaining clips towards the side of the unit allowing the ultrasonic to clear the inside of the hole in the counter top.E) Push/pull ultrasonic up and out of the counter top.
Remove ultrasonic From counter Top (Recessed)
Clip Bottom
Retaining Clip
Remote Cable Plug
Socket
Power Connection
On/OffSwitch
Drain Valve
Drain Elbow
Hose Connection
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
C-5© Midmark Corporation 2018 004-10061-00 (11/13/18)
Remove base plate Assembly (Recessed)
step 1: Remove base plate Assembly.A) Place front (plug/power switch) side of unit down and remove the screws from bottom plate (number of screws depends on model). B) Separate bottom plate from ultrasonic and rotate top of plate down and away from unit. This allows you to place the bottom plate on the table without disconnecting wires.
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
C-6© Midmark Corporation 2018 004-10061-00 (11/13/18)
step 1: disconnect Wiring.A) Remove Bottom Plate Assembly (See base plate removal).B) Disconnect red and black power wires at the Ultrasonic Generator PCB by loosening the screws on the connector and removing the wires. C) Disconnect yellow level sensor wire & red and white thermal sensor wires (going to side of tank) from Control PCB using a small screwdriver to push on the tabs while pulling wires from connector.
step 2: Remove/Replace Tank Assembly.A) Remove nyloc nut and washer from the tank stud in each corner.B) Remove the terminal holding the ground wire to the tank stud located in the same corner as the power switch.C) Depress and hold the quick fitting retaining ring (shown in red) toward the fitting to release it’s grip on the metal tube coming from the bottom of the tank. D) Slowly remove the tank from the case. The metal tube extending into the elbow will come out of the elbow fitting as the tank is removed.E) Reassemble in reverse order.F) Replace any zip ties that were cut during disassembly.
Remove Tank Assembly (Recessed)
Ultrasonic Generator PCBYellow Level Sensor Wire
Power Supply PCB
Quick Fitting
Red and White Thermal Sensor Wires
Quick Fitting
Main Power Wires
To Front Panel Asm
M4 Nyloc Nut, M5 Washer, & Ground Terminal
Equipment Alert To prevent damage to the wires ensure they are routed correctly and secured with zip ties.
WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
D-1© Midmark Corporation 2018 004-10061-00 (06/08/17)
Sect
ion
DWiring diagrams Table Top Unit ................................................... D-2
Recessed Unit ................................................... D-3Recessed Display Panel Assembly ............... D-4
D-2© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00 (11/13/18)
Filter
Qc1-01,Qc3-01,Qc6-01all
Power Switch
Power Connection
Power Supply
Ultrasonic Generator
Control & Display PCB
Touchpad
Temperature Sensor Bottom Plate
Level Sensor
Temperature Sensor Tank
Wiring diagram Table Top
Fuses:10 amp slow blow
Fuses:5 amp slow blow
Fuses:10 amp slow blow
D-3© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00(9/24/18)
Wiring diagram Recessed
Qc3R-01,Qc6R-01all
Power Switch
Filter
Power Supply
Ultrasonic Generator
Temperature Sensor Tank
Temperature Sensor Bottom Plate
Level Sensor
IDCBreakoutPCB
Power Connection
Fuses:10 amp slow blow
Fuses:5 amp slow blow
Fuses:10 amp slow blow
D-4© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00 (11/13/18)
Qc3R-01,Qc6R-01all
LCD Display
To Fill Level Sensor
Brow
nG
reyYellowG
reenP
inkB
lueR
edD
ark Orange
White
Purple
Orange
Black
Base Plate
Con
trol B
oard
Touchpad
From Ultrasonic Generator PCB
IDCBreakoutPCB
Wiring diagram Recessed display
OrangeDark Orange White Black BlueRedBrown Purple
To Tank Temp SensorTo Bottom Plate Temp Sensor
GreyGreenPinkYellow
E-1© Midmark Corporation 2018 004-10061-00 (06/08/17)
Sect
ion
EExploded Views & parts Lists
XXX-XXXX-XX
XXX-XXXX-XX
XXX-XXXX-XX
Table Top ...........................................E-2Recessed...........................................E-3
E-2© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00 (11/13/18)
Table Top unit
Qc1-01,Qc3-01,Qc6-01all
Base Plate Assembly ...................E-8Aluminium plate with Power Supply and Ultrasonic Generator PCBs mounted to it. Attached to the case by screw(s) and screws holding the rubber feet.
Case ................................................E-6S/S case, power switch, main filter, power inlet and ball valve to drain. All other assem-blies are fitted to the case. The size varies depending on model
Tank Assembly ..............................E-4S/S Tank, Temperature sensor, Liquid Level Sensor, drain outlet and Transducers.
Transducers per model:QC1-01 x 2 TransducerQC3-01 x 4 TransducerQC6-01 x 8 Transducer
The tank is attached to the case by internal nuts, wired to the Base Plate Assembly and Front Panel Assembly. A series of plumbing connects to the drain.
Front panel Assembly ...............E-10ABS moulding that consists of Keypad, LCD Display, Control PCB and Label. Attached to case by internal screws.
E-3© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00 (11/13/18)
Recessed unit
Qc3R-01,Qc6R-01all
Base Plate Assembly ...................E-9Aluminium plate with Power Supply, Ultrasonic Generator PCB, And XXX PCB’s mounted to it. Attached to the case by screw(s).
Case .................................................E-7S/S case, power switch, main filter, power inlet and ball valve to drain. All other assem-blies are fitted to the case. The size varies depending on model
Tank Assembly ..............................E-5S/S Tank, Temperature sensor, Liquid Level Sensor, drain outlet and Transducers.
Transducers per model:QC3R-01 x 4 TransducerQC6R-01 x 8 Transducer
The tank is attached to the case by internal nuts, wired to the Base Plate Assembly and Front Panel Assembly. A series of plumbing connects to the drain. Fixing Clips secure the unit to the counter top from the bottom.
Front panel Assembly ...............E-11ABS moulding that consists of Keypad, LCD Display, Control PCB and Cable connecting to the base plate. Attached to cabinet by screws with covers.
E-4© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00 (07/11/18)
Qc1-01,Qc3-01,Qc6-01all
9
8
1 002-10107-0010
76
5 4
2 a) 053-10032-00 b) 053-10033-00 c) 053-10034-00
Tank Assembly Table TopItem Description Qty
1 FillLevelSensor(includes 4 thru 10) 1
2 TankLida) QC1-01b)QC3-01c) QC6-01
1
3 TemperatureSensor 1
4 LevelSensorScrew 1
5 LevelSensorInsulator 1
6 LevelSensorNylonWasher 2
7 LevelSensorPlainNut 1
8 LevelSensorHarness 1
9 LevelSensorNylocNut 1
10 LevelSensorPlainWasher 3
3
E-5© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00(3/29/19)
Qc3R-01,Qc6R-01all
2 a) 029-10151-00 b) 029-10152-00
Tank Assembly RecessedItem Description Qty
1 FillLevelSensor(includes 6 thru 12) 1
2 TankLid(includes item 3)a) QC3R-01b)QC6R-01
1
3 MidmarkLabel 1
4 RetainingClip 4
5 TemperatureSensor 1
6 LevelSensorScrew 1
7 LevelSensorInsulator 1
8 LevelSensorNylonWasher 2
9 LevelSensorPlainNut 1
10 LevelSensorHarness 1
11 LevelLockNylocNut 1
12 LevelSensorPlainWasher 3
11
10
1 002-10107-0012
9 8
76
4 016-10041-00
061-10180-00 3
5
E-6© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00(3/29/19)
Qc1-01thru 602439
case Table TopItem Description Qty
1 QuickFittingElbow 1
2 PVC1/2” 1
3 QuickFittinga) QC1-01 (QC x 3/8” thread)b)QC3-01&QC6-01(QC x 1/2” thread)
1
4 Drain Hose (1/2” ID ) by the inch 48
5 BulkheadFittinga) QC1-01 ( 3/8” )b)QC3-01&QC6-01 (1/2” )
1
6 ShutoffValvea) QC1-01 (3/8” Male/Female)b)QC3-01&QC6-01(1/2” Male/Female)
1
7 Elbowa) QC1-01 (3/8” BSPT x 12 mm barb)b)QC3-01&QC6-01 (1/2” BSPT x 12 mm barb)
1
8 MainPowerSwitch(DPST) 1
9 MainPowerFuse (20mm, 10A (T) A/S, ceramic) 2
10 MainPowerSocket(includes 9) 1
11 MainPowerFiltera) QC1-01 (3 Amp)b)QC3-01&QC6-01(5 amp)
1
5
015-10100-00 10
4 014-0609-00
015-10187-00 9
7 a) 014-10009-00 b) 014-10008-00
6 a) 014-10029-00 b) 014-10028-00
8 015-10194-00
5
3
2
1
11
Qc3-01,Qc6-01all
E-6.1© Midmark Corporation 2017 TP204 20-42-FO-00015 Rev A1 C2169
Models:serial Numbers:
Always verify model & serial number 004-10061-00(3/29/19)© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number
Qc1-01602440 to Present
case Table TopItem Description Qty
1 QuickFittingElbow 1
2 PVC1/2” 1
3 QuickFitting (QC x 1/2” thread) 1
4 Drain Hose (1/2” ID ) by the inch 48
5 BulkheadFitting (1/2” ) 1
6 ShutoffValve (1/2” Male/Female) 1
7 Elbow (1/2” BSPT x 12 mm barb) 1
8 MainPowerSwitch(DPST) 1
9 MainPowerFuse (20mm, 10A (T) A/S, ceramic) 2
10 MainPowerSocket(includes 9) 1
11 MainPowerFiltera) QC1-01 (3 Amp)b)QC3-01&QC6-01(5 amp)
1
5
015-10100-00 10
4 014-0609-00
015-10187-00 9
8 015-10194-00
5
3
2
1
11
Qc3-01,Qc6-01all
6 014-10028-00
7 014-10008-00
E-7© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00(3/29/19)
Qc3R-01,Qc6R-01all
2
014-0609-00 5
015-10187-00 7
015-10100-00 6
case RecessedItem Description Qty
1 QuickFitting (QC x 1/2” thread) 1
2 ShutoffValve(1/2” Male/Female) 1
3 Elbow (1/2” BSPT x 12 mm barb) 1
4 MainPowerFilter (5 amp) 1
5 Drain Hose (1/2” ID ) by the inch 108
6 MainPowerSocket 1
7 MainPowerFuse(10A (T) A/S,Slow Blo) 1
8 MainPowerSwitch 2
3 a) 014-10008-00
2 a) 014-10028-00
015-10194-00 8
1
4
E-8© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00 (1/21/20)
Qc1-01,Qc3-01,Qc6-01all
base plate Assembly Table TopItem Description Qty
1 PowerSupplyPCB 1
2 Screw AR
3 RibbonCable 1
4 PowerSupplyPCBFuses (10A (T) A/S,Slow Blo)
2
5 BasePlateAssembly(includes 6 thru 9)a) QC1-01b)QC3-01c) QC6-01
1
6 TemperatureSensor 2
7 UltrasonicGeneratorPCBFuses (5A (T) A/S,Slow Blo)
2
8 UltrasonicGeneratorPCB 1
9 BasePlate 1
10 BaseHardwareKit(includes 11 thru 13) 1
11 Foot 4
12 Washer AR
13 Screw AR
a) 029-10087-00 5b) 029-10089-00 c) 029-10090-00
2
4 015-10187-00
7 015-10187-01
8
2
002-10181-00 10
11
13
12
96
015-10130-00 1
015-10144-00 3To Display PCB
E-9© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00 (1/21/20)
Qc3R-01,Qc6R-01all
base plate Assembly RecessedItem Description Qty
1 IDCBreakoutPCB 1
2 PowerSupplyPCB 1
3 RibbonCable 1
4 PowerSupplyPCBFuses (20mm, 10A (T) A/S, ceramic) 2
5 Screw AR
6 BasePlateAssembly(includes 7 thru 11)a) QC3R-01b)QC6R-01
1
7 TemperatureSensor 1
8 12pinconnector 1
9 UltrasonicGeneratorPCBFuses (20mm, 5 A (T) A/S, ceramic) 2
10 UltrasonicGeneratorPCB 1
11 BasePlate 1
12 Washer AR
13 Screw AR
015-10130-00 2
015-10106-00 1
9 015-10187-01
015-10155-00 8
a) 029-10158-00 6b) 029-10161-00
015-10144-00 3
5
5
5
10
11
12
13
7
4 015-10187-00
E-10© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00(3/29/19)
1 a) 002-10177-00 b) 002-10179-00
Qc1-01,Qc3-01,Qc6-01all
8 002-10166-00
511
061-10180-00 2
Front panel Assembly Table TopItem Description Qty
1 FrontPanelAssembly(includes 2 thru 4)a)QC1-01&QC3-01b)QC6-01
1
2 MidmarkLabel 1
3 TouchPad NA
4 FrontPanel NA
5 Screw AR
6 AngledPlateLeft 1
7 AngledPlateRight 1
8 FrontPanelPCBKit(includes 9 & 10) 1
9 DisplayPCB 1
10 ControlPCB 1
11 Screw AR
3
4
11
6
5
7
910
E-11© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00(3/29/19)
Qc3R-01,Qc6R-01all
16
Front panel Assembly RecessedItem Description Qty
1 FrontPanelAssembly(includes 3 thru 14) 1
2 FrontPanelKit(includes 4 thru 7) 1
3 Screw&ScrewCoverKit(includes 4 & 5) 1
4 ScrewCover 4
5 Screw 4
6 TouchPad 1
7 Frontcover 1
8 AngledPlateLeft 1
9 AngledPlateRight 1
10 DisplayPCB 1
11 Screw AR
12 Bracket 1
13 ControlCableKit 1
14 ControlPCB 1
15 BackCover 1
16 Screw AR
13 015-10112-00
029-10141-00 1
16
15
8
11
12
12
9
7
11
11
54
6 2 002-10167-00
3 002-10182-00
10
14 002-10312-00
E-12© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00 (07/11/18)
1 a) 002-10007-00 b) 002-10008-00 c) 002-10009-00
Qc1-01,Qc3(R)-01,Qc6(R)-01all
basketItem Description Qty
1 Basket (Includes Item 2)a) QC1-01b)QC3(R)-01c) QC6(R)-01
1
2 BasketRubberFootKit(contains4feet) 1
2 002-10185-00
E-13© Midmark Corporation 2018
Models:serial Numbers:
Always verify model & serial number 004-10061-00(3/29/19)
1 a) 9A612001 b) 9A613001 c) 9A614001
Qc1-01,Qc3-01,Qc6-01all
beaker holder AssemblyItem Description Qty
1 BeakerHolderAssembly(includes 2 thru 5)a) QC1-01b)QC3(R)-01c) QC6(R)-01
1
2 Beakerholder 1
3 O-ring AR
4 Beaker AR
5 Lid AR
5 053-1778-00
4 053-1779-00
3 053-1782-00
2